Reliability of peg restrained intrinsic muscle evaluator for measurement of intrinsic hand muscle strength in adults with tetraplegia

N. Yozbatiran, M. Russell, Radha Korupolu, R. Grossman, Elizabeth G. Toups, G. Francisco
{"title":"Reliability of peg restrained intrinsic muscle evaluator for measurement of intrinsic hand muscle strength in adults with tetraplegia","authors":"N. Yozbatiran, M. Russell, Radha Korupolu, R. Grossman, Elizabeth G. Toups, G. Francisco","doi":"10.4103/jisprm.jisprm_27_20","DOIUrl":null,"url":null,"abstract":"Background: The manual muscle test (MMT) is the traditional method to assess the intrinsic hand muscle test. However, this test is qualitative and subjective. A device capable of rapidly measuring motor output along a linear scale may be of value in the evaluation of hand recovery from any neurological impairment. Objective: To demonstrate inter-rater reliability of the Peg Restrained Intrinsic Muscle Evaluator (PRIME) device for measuring intrinsic hand muscle strength in adults with impaired hand functions. Methods: 16 subjects (2 female, 14 male, mean age=47±15 years) with cervical spinal cord injury were enrolled. Intrinsic hand muscle strength of adults were performed by two-tester with use of PRIME device and manual muscle testing, within the same session. Inter-tester correlation was tested by Pearson's correlation and Intraclass correlation coefficient (ICC). Results: Intraclass correlation coefficients for hypothenar, first dorsal interosseous, abductor pollicis brevis and opponence pollicis muscles ranged from 0.7 to 0.8 with hypothenar muscle test being lowest ICC=0.7 and thumb opposition highest, ICC=0.81. Conclusion: Results from this study suggest that PRIME device demonstrates good reliability within testers for quantified measuring of intrinsic hand muscle strength. Further use of PRIME in clinic will aid in diagnostics, medical decision making and evaluation of rehabilitation progress in patients with cervical spinal cord injury.","PeriodicalId":75125,"journal":{"name":"The journal of the International Society of Physical and Rehabilitation Medicine","volume":"4 1","pages":"34 - 39"},"PeriodicalIF":0.0000,"publicationDate":"2021-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"The journal of the International Society of Physical and Rehabilitation Medicine","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.4103/jisprm.jisprm_27_20","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Background: The manual muscle test (MMT) is the traditional method to assess the intrinsic hand muscle test. However, this test is qualitative and subjective. A device capable of rapidly measuring motor output along a linear scale may be of value in the evaluation of hand recovery from any neurological impairment. Objective: To demonstrate inter-rater reliability of the Peg Restrained Intrinsic Muscle Evaluator (PRIME) device for measuring intrinsic hand muscle strength in adults with impaired hand functions. Methods: 16 subjects (2 female, 14 male, mean age=47±15 years) with cervical spinal cord injury were enrolled. Intrinsic hand muscle strength of adults were performed by two-tester with use of PRIME device and manual muscle testing, within the same session. Inter-tester correlation was tested by Pearson's correlation and Intraclass correlation coefficient (ICC). Results: Intraclass correlation coefficients for hypothenar, first dorsal interosseous, abductor pollicis brevis and opponence pollicis muscles ranged from 0.7 to 0.8 with hypothenar muscle test being lowest ICC=0.7 and thumb opposition highest, ICC=0.81. Conclusion: Results from this study suggest that PRIME device demonstrates good reliability within testers for quantified measuring of intrinsic hand muscle strength. Further use of PRIME in clinic will aid in diagnostics, medical decision making and evaluation of rehabilitation progress in patients with cervical spinal cord injury.
peg约束内肌评估器测量四肢瘫痪成人内肌力量的可靠性
背景:手工肌肉测试(MMT)是评估手部固有肌肉测试的传统方法。然而,这种测试是定性的和主观的。一种能够沿线性刻度快速测量运动输出的装置可能在评估任何神经损伤的手部恢复方面具有价值。目的:验证Peg约束内征肌肉评估器(PRIME)测量成人手功能障碍内征肌肉力量的可靠性。方法:选取16例颈脊髓损伤患者,其中女性2例,男性14例,平均年龄47±15岁。在同一时段内,采用PRIME装置和手动肌肉测试两种方法对成人的内在手部肌肉力量进行测试。采用Pearson相关和class内相关系数(Intraclass correlation coefficient, ICC)检验检验者间相关。结果:鱼际下肌、第一背骨间肌、拇短外展肌和拇对跖肌的类内相关系数在0.7 ~ 0.8之间,其中鱼际下肌的相关系数最低,ICC=0.7,拇指对跖肌的相关系数最高,ICC=0.81。结论:本研究的结果表明,PRIME装置在测试器中具有良好的可靠性,可用于手部固有肌肉力量的量化测量。进一步将PRIME应用于临床,将有助于颈脊髓损伤患者的诊断、医疗决策和康复进展评估。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
审稿时长
15 weeks
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信