Research and development of the sOLVe Tube™ dual lumen endobronchial tube: from concept to construct.

IF 2.7 Q3 ENGINEERING, BIOMEDICAL
Frontiers in medical technology Pub Date : 2023-09-15 eCollection Date: 2023-01-01 DOI:10.3389/fmedt.2023.1158154
Patricia Nwajuaku, Igor Barjaktarevic, Nir Hoftman
{"title":"Research and development of the sOLVe Tube™ dual lumen endobronchial tube: from concept to construct.","authors":"Patricia Nwajuaku,&nbsp;Igor Barjaktarevic,&nbsp;Nir Hoftman","doi":"10.3389/fmedt.2023.1158154","DOIUrl":null,"url":null,"abstract":"<p><strong>Introduction: </strong>Dual lumen endobronchial tubes (DLTs) are frequently used for lung isolation and one lung ventilation in thoracic surgery and other specialized clinical scenarios. Modern DLTs are large and rigid, and account for half of all tracheobronchial injuries. Their 70 year old design has numerous flaws which limit their safety and clinical utility. Our research team set out to design a new and improved DLT to mitigate these shortcomings, and then test the proposed device to ensure proper function.</p><p><strong>Methods: </strong>Using published airway anatomy data and computed tomography imaging from 195 thoracic surgery patients, we designed a new DLT with a single size/configuration that would fit into adult surgery patients. This single \"Universal design\" was intended to replace both left and right sided 35Fr-41Fr DLTs (8 total products), while remaining small in diameter (35Fr). Other design goals included: 1) making intubation easier and safer, 2) allowing full sized therapeutic bronchoscopes to fit into this tube, 3) making the DLT more resistant to dislodgement. After design process completion the proposed dimensions were tested against 195 patients' left and right mainstem bronchi for radiographic fit. Once production prototypes were manufactured, they were tested in large adult Yorkshire pigs and fresh human cadavers for anatomic fit and performance.</p><p><strong>Results: </strong>The proposed design passed the radiographic fit test in all 195 patients for both left and right mainstem endobronchial placement. Intubation was successful and deemed atraumatic in all pigs and cadavers, and the device appropriately fit in both the right and left mainstem bronchi. Lung isolation was successfully achieved and the device proved resistant to axial force dislodgement.</p><p><strong>Conclusion: </strong>We propose a new design for a novel DLT meant to replace 8 currently supplied adult configurations with a single, one size/configuration fits all product that allows for large bore bronchoscopy and resists axial force dislodgement.</p>","PeriodicalId":94015,"journal":{"name":"Frontiers in medical technology","volume":"5 ","pages":"1158154"},"PeriodicalIF":2.7000,"publicationDate":"2023-09-15","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10541992/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Frontiers in medical technology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.3389/fmedt.2023.1158154","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2023/1/1 0:00:00","PubModel":"eCollection","JCR":"Q3","JCRName":"ENGINEERING, BIOMEDICAL","Score":null,"Total":0}
引用次数: 0

Abstract

Introduction: Dual lumen endobronchial tubes (DLTs) are frequently used for lung isolation and one lung ventilation in thoracic surgery and other specialized clinical scenarios. Modern DLTs are large and rigid, and account for half of all tracheobronchial injuries. Their 70 year old design has numerous flaws which limit their safety and clinical utility. Our research team set out to design a new and improved DLT to mitigate these shortcomings, and then test the proposed device to ensure proper function.

Methods: Using published airway anatomy data and computed tomography imaging from 195 thoracic surgery patients, we designed a new DLT with a single size/configuration that would fit into adult surgery patients. This single "Universal design" was intended to replace both left and right sided 35Fr-41Fr DLTs (8 total products), while remaining small in diameter (35Fr). Other design goals included: 1) making intubation easier and safer, 2) allowing full sized therapeutic bronchoscopes to fit into this tube, 3) making the DLT more resistant to dislodgement. After design process completion the proposed dimensions were tested against 195 patients' left and right mainstem bronchi for radiographic fit. Once production prototypes were manufactured, they were tested in large adult Yorkshire pigs and fresh human cadavers for anatomic fit and performance.

Results: The proposed design passed the radiographic fit test in all 195 patients for both left and right mainstem endobronchial placement. Intubation was successful and deemed atraumatic in all pigs and cadavers, and the device appropriately fit in both the right and left mainstem bronchi. Lung isolation was successfully achieved and the device proved resistant to axial force dislodgement.

Conclusion: We propose a new design for a novel DLT meant to replace 8 currently supplied adult configurations with a single, one size/configuration fits all product that allows for large bore bronchoscopy and resists axial force dislodgement.

Abstract Image

Abstract Image

Abstract Image

sOLVe管的研制™ 双腔支气管内膜管:从概念到结构。
简介:在胸部手术和其他专业临床场景中,双腔支气管内导管(DLT)经常用于肺部隔离和单肺通气。现代DLT体积大且坚硬,占所有气管支气管损伤的一半。他们70年的设计有许多缺陷,限制了他们的安全性和临床实用性。我们的研究团队着手设计一种新的、改进的DLT,以减轻这些缺点,然后测试所提出的设备,以确保其正常功能。方法:利用195名胸部手术患者已发表的气道解剖数据和计算机断层扫描成像,我们设计了一种新的DLT,其尺寸/配置单一,适合成人手术患者。这种单一的“通用设计”旨在取代左侧和右侧的35Fr-41Fr DLT(共8种产品),同时保持较小的直径(35Fr)。其他设计目标包括:1)使插管更容易、更安全,2)允许全尺寸治疗性支气管镜插入该管,3)使DLT更耐移位。在设计过程完成后,对195名患者的左主干和右主干支气管进行了拟议尺寸的测试,以进行放射学拟合。一旦生产原型被制造出来,它们就在约克郡的大型成年猪和新鲜的人体尸体上进行了解剖学拟合和性能测试。结果:所提出的设计在所有195名患者中均通过了左主干和右主干支气管内放置的放射学拟合测试。插管是成功的,在所有猪和尸体上都被认为是无创伤的,该装置适合右主干和左主干支气管。成功实现了肺部隔离,证明该装置能够抵抗轴向力移位。结论:我们提出了一种新型DLT的新设计,旨在用一种单一的、一刀切的产品取代目前提供的8种成人配置,该产品允许大口径支气管镜检查并抵抗轴向力移位。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
3.70
自引率
0.00%
发文量
0
审稿时长
13 weeks
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信