David H Owen, Dongyang Wang, Xu Cong, Cameron Mowbray, Diana M Perriman, Chris J Roberts, Paul N Smith, Herwig Drobetz, David Ackland
{"title":"Biomechanical Performance of Total Wrist Arthrodesis Plates With and Without Arthrodesis of the Carpometacarpal Joint.","authors":"David H Owen, Dongyang Wang, Xu Cong, Cameron Mowbray, Diana M Perriman, Chris J Roberts, Paul N Smith, Herwig Drobetz, David Ackland","doi":"10.1177/15589447231198263","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>It is unknown whether total wrist arthrodesis (TWA) should be performed with or without arthrodesis of the carpometacarpal joint (CMCJ). The aim of this study is to compare CMCJ-spanning TWA plates using 3D printed wrist arthrodesis model with and without arthrodesis of the CMCJ.</p><p><strong>Methods: </strong>Total wrist arthrodesis plates mounted to 3D printed models were tested under a 4-N bending load at 4 Hz for 50 000 cycles, increased by 15% every 10 000 cycles until failure.</p><p><strong>Results: </strong>Plates with arthrodesis CMCJ were stiffer and failed at a significantly greater load and number of cycles than plates mounted to models without CMCJ arthrodesis. The Synthes stainless steel locking TWA plate performed better than the Trimed plate applied to the model without CMCJ arthrodesis and the Acumed plate applied to the model with CMCJ arthrodesis. Based on these findings, we recommend arthrodesis of the CMCJ in TWA.</p><p><strong>Conclusions: </strong>Incorporation of the CMCJ in TWA may protect against plate failure. If arthrodesis of the CMCJ is not performed, plate removal should be considered before breakage occurs.</p><p><strong>Level of evidence: </strong>IV.</p>","PeriodicalId":12902,"journal":{"name":"HAND","volume":" ","pages":"230-236"},"PeriodicalIF":1.8000,"publicationDate":"2025-03-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11833838/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"HAND","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1177/15589447231198263","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2023/10/7 0:00:00","PubModel":"Epub","JCR":"Q2","JCRName":"ORTHOPEDICS","Score":null,"Total":0}
引用次数: 0
Abstract
Background: It is unknown whether total wrist arthrodesis (TWA) should be performed with or without arthrodesis of the carpometacarpal joint (CMCJ). The aim of this study is to compare CMCJ-spanning TWA plates using 3D printed wrist arthrodesis model with and without arthrodesis of the CMCJ.
Methods: Total wrist arthrodesis plates mounted to 3D printed models were tested under a 4-N bending load at 4 Hz for 50 000 cycles, increased by 15% every 10 000 cycles until failure.
Results: Plates with arthrodesis CMCJ were stiffer and failed at a significantly greater load and number of cycles than plates mounted to models without CMCJ arthrodesis. The Synthes stainless steel locking TWA plate performed better than the Trimed plate applied to the model without CMCJ arthrodesis and the Acumed plate applied to the model with CMCJ arthrodesis. Based on these findings, we recommend arthrodesis of the CMCJ in TWA.
Conclusions: Incorporation of the CMCJ in TWA may protect against plate failure. If arthrodesis of the CMCJ is not performed, plate removal should be considered before breakage occurs.
期刊介绍:
HAND is the official journal of the American Association for Hand Surgery and is a peer-reviewed journal featuring articles written by clinicians worldwide presenting current research and clinical work in the field of hand surgery. It features articles related to all aspects of hand and upper extremity surgery and the post operative care and rehabilitation of the hand.