Can Clinicians Trust Objective Measures of Hip Muscle Strength From Portable Dynamometers? A Systematic Review With Meta-analysis and Evidence Gap Map of 107 Studies of Reliability and Criterion Validity Using the COSMIN Methodology.
Marina C Waiteman, Micah C Garcia, Ronaldo V Briani, Grant Norte, Neal R Glaviano, Fábio M De Azevedo, David M Bazett-Jones
{"title":"Can Clinicians Trust Objective Measures of Hip Muscle Strength From Portable Dynamometers? A Systematic Review With Meta-analysis and Evidence Gap Map of 107 Studies of Reliability and Criterion Validity Using the COSMIN Methodology.","authors":"Marina C Waiteman, Micah C Garcia, Ronaldo V Briani, Grant Norte, Neal R Glaviano, Fábio M De Azevedo, David M Bazett-Jones","doi":"10.2519/jospt.2023.12045","DOIUrl":null,"url":null,"abstract":"<p><p><b>OBJECTIVE:</b> To summarize the evidence on reliability and criterion validity of hip muscle strength testing using portable dynamometers. <b>DESIGN:</b> Systematic review with meta-analysis. <b>LITERATURE SEARCH:</b> Five databases were searched from inception to March 2023. <b>STUDY SELECTION CRITERIA:</b> We included studies investigating reliability or criterion validity of hip flexor, extensor, abductor, adductor, or internal/external rotator strength testing with portable dynamometers in injury-free individuals or those with pelvic/lower limb musculoskeletal disorders. <b>DATA SYNTHESIS:</b> We performed meta-analyses for each muscle group, position, and method of fixation. We rated pooled results as sufficient (>75% of studies with correlations ≥0.70), insufficient (>75% of studies with correlations <0.70), or inconsistent (sufficient/insufficient results). We assessed the quality of evidence, created evidence gap maps, and made clinical recommendations. <b>RESULTS:</b> We included a total of 107 studies (reliability 103, validity 14). The intrarater and interrater reliability for hip muscle strength testing across different positions and methods of fixation was sufficient (intraclass correlation coefficient = 0.78-0.96) with low- to high-quality evidence. Criterion validity was less investigated and mostly inconsistent (very low-to moderate-quality evidence) with a wide range of correlations (r = 0.40-0.93). <b>CONCLUSION:</b> Hip muscle strength testing using portable dynamometers is reliable. The use of portable dynamometers as clinical surrogates for measuring strength using an isokinetic dynamometer requires further investigation. Clinicians testing hip muscle strength with portable dynamometers should use external fixation seated for hip flexors, prone or supine for hip extensors, side-lying or supine for abductors and adductors, and prone and seated for internal and external rotators. <i>J Orthop Sports Phys Ther 2023;53(11):655-672. Epub 3 October 2023. doi:10.2519/jospt.2023.12045</i>.</p>","PeriodicalId":50099,"journal":{"name":"Journal of Orthopaedic & Sports Physical Therapy","volume":" ","pages":"655-672"},"PeriodicalIF":6.0000,"publicationDate":"2023-11-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Orthopaedic & Sports Physical Therapy","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.2519/jospt.2023.12045","RegionNum":1,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"ORTHOPEDICS","Score":null,"Total":0}
引用次数: 0
Abstract
OBJECTIVE: To summarize the evidence on reliability and criterion validity of hip muscle strength testing using portable dynamometers. DESIGN: Systematic review with meta-analysis. LITERATURE SEARCH: Five databases were searched from inception to March 2023. STUDY SELECTION CRITERIA: We included studies investigating reliability or criterion validity of hip flexor, extensor, abductor, adductor, or internal/external rotator strength testing with portable dynamometers in injury-free individuals or those with pelvic/lower limb musculoskeletal disorders. DATA SYNTHESIS: We performed meta-analyses for each muscle group, position, and method of fixation. We rated pooled results as sufficient (>75% of studies with correlations ≥0.70), insufficient (>75% of studies with correlations <0.70), or inconsistent (sufficient/insufficient results). We assessed the quality of evidence, created evidence gap maps, and made clinical recommendations. RESULTS: We included a total of 107 studies (reliability 103, validity 14). The intrarater and interrater reliability for hip muscle strength testing across different positions and methods of fixation was sufficient (intraclass correlation coefficient = 0.78-0.96) with low- to high-quality evidence. Criterion validity was less investigated and mostly inconsistent (very low-to moderate-quality evidence) with a wide range of correlations (r = 0.40-0.93). CONCLUSION: Hip muscle strength testing using portable dynamometers is reliable. The use of portable dynamometers as clinical surrogates for measuring strength using an isokinetic dynamometer requires further investigation. Clinicians testing hip muscle strength with portable dynamometers should use external fixation seated for hip flexors, prone or supine for hip extensors, side-lying or supine for abductors and adductors, and prone and seated for internal and external rotators. J Orthop Sports Phys Ther 2023;53(11):655-672. Epub 3 October 2023. doi:10.2519/jospt.2023.12045.
期刊介绍:
The Journal of Orthopaedic & Sports Physical Therapy® (JOSPT®) publishes scientifically rigorous, clinically relevant content for physical therapists and others in the health care community to advance musculoskeletal and sports-related practice globally. To this end, JOSPT features the latest evidence-based research and clinical cases in musculoskeletal health, injury, and rehabilitation, including physical therapy, orthopaedics, sports medicine, and biomechanics.
With an impact factor of 3.090, JOSPT is among the highest ranked physical therapy journals in Clarivate Analytics''s Journal Citation Reports, Science Edition (2017). JOSPT stands eighth of 65 journals in the category of rehabilitation, twelfth of 77 journals in orthopedics, and fourteenth of 81 journals in sport sciences. JOSPT''s 5-year impact factor is 4.061.