Loss of cardiac synchronization due to automatic sense testing.

Pacing and clinical electrophysiology : PACE Pub Date : 2022-09-01 Epub Date: 2022-08-16 DOI:10.1111/pace.14574
Hans Römers, Max Liebregts, Zeliha Koyak, Jippe C Balt
{"title":"Loss of cardiac synchronization due to automatic sense testing.","authors":"Hans Römers,&nbsp;Max Liebregts,&nbsp;Zeliha Koyak,&nbsp;Jippe C Balt","doi":"10.1111/pace.14574","DOIUrl":null,"url":null,"abstract":"<p><p>In the latest generation Biotronik cardiac resynchronization devices, cardiac resynchronization therapy (CRT) may be inadvertently interrupted due to automatic sense testing. This issue can easily be recognized during device interrogation by the \"CRT interrupt warning.\"To avoid CRT interruption, both understanding of the algorithm and correct device programming are critical. The automatic sense testing algorithm has no built-in protection to avoid CRT interrupted pacing. When implanting this generation of devices programming needs to be adequate to avoid the occurrence of this phenomenon.</p>","PeriodicalId":520740,"journal":{"name":"Pacing and clinical electrophysiology : PACE","volume":" ","pages":"1085-1089"},"PeriodicalIF":0.0000,"publicationDate":"2022-09-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Pacing and clinical electrophysiology : PACE","FirstCategoryId":"5","ListUrlMain":"https://doi.org/10.1111/pace.14574","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2022/8/16 0:00:00","PubModel":"Epub","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

In the latest generation Biotronik cardiac resynchronization devices, cardiac resynchronization therapy (CRT) may be inadvertently interrupted due to automatic sense testing. This issue can easily be recognized during device interrogation by the "CRT interrupt warning."To avoid CRT interruption, both understanding of the algorithm and correct device programming are critical. The automatic sense testing algorithm has no built-in protection to avoid CRT interrupted pacing. When implanting this generation of devices programming needs to be adequate to avoid the occurrence of this phenomenon.

由于自动感觉测试导致的心脏同步性丧失。
在最新一代Biotronik心脏再同步化装置中,由于自动感觉测试,心脏再同步化治疗(CRT)可能会在无意中中断。这个问题可以很容易地在设备询问时通过“CRT中断警告”识别出来。为了避免CRT中断,理解算法和正确的器件编程都是至关重要的。自动感测算法没有内置保护,避免CRT中断起搏。在植入这一代设备时,需要充分的编程来避免这种现象的发生。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信