IMPACT OF PENETRATION AND IMAGE ANALYSIS IN OPTICAL COHERENCE TOMOGRAPHY ON THE MEASUREMENT OF CHOROIDAL VASCULARITY PARAMETERS.

Sisi Chen, Gu Zheng, Xiangle Yu, Yanfeng Jiang, Zhiyang Lin, Guangqing Lin, Wen Chen, Meixiao Shen, Fan Lu
{"title":"IMPACT OF PENETRATION AND IMAGE ANALYSIS IN OPTICAL COHERENCE TOMOGRAPHY ON THE MEASUREMENT OF CHOROIDAL VASCULARITY PARAMETERS.","authors":"Sisi Chen,&nbsp;Gu Zheng,&nbsp;Xiangle Yu,&nbsp;Yanfeng Jiang,&nbsp;Zhiyang Lin,&nbsp;Guangqing Lin,&nbsp;Wen Chen,&nbsp;Meixiao Shen,&nbsp;Fan Lu","doi":"10.1097/IAE.0000000000003547","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>To investigate the impact of penetration and image analysis in different optical coherence tomography (OCT) instruments on the measurement of choroidal vascularity parameters.</p><p><strong>Methods: </strong>Twenty-three healthy volunteers were imaged using two swept-source OCTs and one spectral-domain OCT. A fully automatic segmentation method based on ResNet-UNet and Niblack local threshold binarization was performed to quantify the relevant choroidal vascular parameters, including choroidal vascularity index, total choroidal volume, and luminal volume. The intraclass correlation coefficient (ICC) and coefficient of repeatability (COR) were used to analyze the repeatability and consistency of automatic and manual segmentation, respectively.</p><p><strong>Results: </strong>Both swept-source OCT devices showed good consistency of luminal volume and total choroidal volume measurements (all ICC value >0.98 with COR% < 8.53%) based on manual segmentation, whereas the consistency of the spectral-domain OCT was lower (ICC value <0.60 with COR% > 40%), which was greatly improved after using the automatic algorithm (ICC value >0.99 with COR% < 4%). The repeatability of choroidal vascularity index obtained from different OCT images using manual or automatic segmentation showed good agreement (all ICC values >0.85), whereas the choroidal vascularity index measurement from the spectral-domain OCT was larger than the other two swept-source OCT devices (ICC value <0.65).</p><p><strong>Conclusion: </strong>For healthy youngsters, the penetration of OCT plays a role in the measurement precision for choroidal vascularity parameters, and automatic segmentation can improve the ability of choroidal boundary identification with deficient penetration, suggesting these factors need to be considered in clinical work.</p>","PeriodicalId":377573,"journal":{"name":"Retina (Philadelphia, Pa.)","volume":" ","pages":"1965-1974"},"PeriodicalIF":0.0000,"publicationDate":"2022-10-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"5","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Retina (Philadelphia, Pa.)","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1097/IAE.0000000000003547","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 5

Abstract

Purpose: To investigate the impact of penetration and image analysis in different optical coherence tomography (OCT) instruments on the measurement of choroidal vascularity parameters.

Methods: Twenty-three healthy volunteers were imaged using two swept-source OCTs and one spectral-domain OCT. A fully automatic segmentation method based on ResNet-UNet and Niblack local threshold binarization was performed to quantify the relevant choroidal vascular parameters, including choroidal vascularity index, total choroidal volume, and luminal volume. The intraclass correlation coefficient (ICC) and coefficient of repeatability (COR) were used to analyze the repeatability and consistency of automatic and manual segmentation, respectively.

Results: Both swept-source OCT devices showed good consistency of luminal volume and total choroidal volume measurements (all ICC value >0.98 with COR% < 8.53%) based on manual segmentation, whereas the consistency of the spectral-domain OCT was lower (ICC value <0.60 with COR% > 40%), which was greatly improved after using the automatic algorithm (ICC value >0.99 with COR% < 4%). The repeatability of choroidal vascularity index obtained from different OCT images using manual or automatic segmentation showed good agreement (all ICC values >0.85), whereas the choroidal vascularity index measurement from the spectral-domain OCT was larger than the other two swept-source OCT devices (ICC value <0.65).

Conclusion: For healthy youngsters, the penetration of OCT plays a role in the measurement precision for choroidal vascularity parameters, and automatic segmentation can improve the ability of choroidal boundary identification with deficient penetration, suggesting these factors need to be considered in clinical work.

光学相干断层扫描中穿透和图像分析对脉络膜血管参数测量的影响。
目的:探讨不同光学相干断层扫描(OCT)仪器的透入和图像分析对脉络膜血管参数测量的影响。方法:对23名健康志愿者进行扫描源oct和谱域oct成像,采用基于ResNet-UNet和Niblack局部阈值二值化的全自动分割方法,量化脉络膜血管指数、脉络膜总容积和管腔容积等脉络膜血管相关参数。采用类内相关系数(ICC)和重复性系数(COR)分别对自动分割和手工分割的重复性和一致性进行分析。结果:两种扫描源OCT设备基于人工分割的管腔容积和脉络膜总容积测量结果一致性较好(ICC值均>0.98,COR% < 8.53%),而光谱域OCT的一致性较低(ICC值40%),采用自动算法后(ICC值>0.99,COR% < 4%),一致性大大提高。人工和自动分割不同OCT图像所得脉络膜血管指数的重复性均较好(ICC值均>0.85),而光谱域OCT所得脉络膜血管指数测量值大于其他两种扫描源OCT设备(ICC值)。对于健康青少年,OCT的穿透性对脉络膜血管参数的测量精度有一定的影响,自动分割可以提高缺乏穿透性的脉络膜边界识别能力,这些因素在临床工作中需要考虑。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信