Day-to-day repeatability of the Pulse Time Index of Norm.

IF 1.3 Q4 ENGINEERING, BIOMEDICAL
Medical Devices-Evidence and Research Pub Date : 2014-02-27 eCollection Date: 2014-01-01 DOI:10.2147/MDER.S58507
Igor N Posokhov, Aleksandra O Konradi, Eugeny V Shlyakhto, Oleg V Mamontov, Artemy V Orlov, Anatoly N Rogoza
{"title":"Day-to-day repeatability of the Pulse Time Index of Norm.","authors":"Igor N Posokhov,&nbsp;Aleksandra O Konradi,&nbsp;Eugeny V Shlyakhto,&nbsp;Oleg V Mamontov,&nbsp;Artemy V Orlov,&nbsp;Anatoly N Rogoza","doi":"10.2147/MDER.S58507","DOIUrl":null,"url":null,"abstract":"<p><p>The pulse wave velocity (PWV) threshold for hypertensive target organ damage is presently set at 10 meters per second. New 24-hour monitors (eg, BPLab® and Vasotens®) provide several PWV measurements over a period of 24-72 hours. A new parameter, ie, the Pulse Time Index of Norm (PTIN), can be calculated from these data. The PTIN is defined as the percentage of a 24-hour period during which the PWV does not exceed 10 meters per second. The aim of the present study was to test the new PTIN for clinical feasibility using day-to-day repeatability analysis. Oscillometrically generated waveform files (n=85), which were previously used for research studies, were reanalyzed using the new 2013 version software of the Vasotens technology program, which enables calculation of PTIN. The intraclass correlation coefficient was 0.98 and Cronbach's alpha was 0.97, indicating that the PTIN has excellent day-to-day repeatability and internal consistency. The present results show adequate repeatability, and PTIN assessment using the Vasotens technology appears to be feasible. </p>","PeriodicalId":47140,"journal":{"name":"Medical Devices-Evidence and Research","volume":" ","pages":"29-33"},"PeriodicalIF":1.3000,"publicationDate":"2014-02-27","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.2147/MDER.S58507","citationCount":"11","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Medical Devices-Evidence and Research","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.2147/MDER.S58507","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2014/1/1 0:00:00","PubModel":"eCollection","JCR":"Q4","JCRName":"ENGINEERING, BIOMEDICAL","Score":null,"Total":0}
引用次数: 11

Abstract

The pulse wave velocity (PWV) threshold for hypertensive target organ damage is presently set at 10 meters per second. New 24-hour monitors (eg, BPLab® and Vasotens®) provide several PWV measurements over a period of 24-72 hours. A new parameter, ie, the Pulse Time Index of Norm (PTIN), can be calculated from these data. The PTIN is defined as the percentage of a 24-hour period during which the PWV does not exceed 10 meters per second. The aim of the present study was to test the new PTIN for clinical feasibility using day-to-day repeatability analysis. Oscillometrically generated waveform files (n=85), which were previously used for research studies, were reanalyzed using the new 2013 version software of the Vasotens technology program, which enables calculation of PTIN. The intraclass correlation coefficient was 0.98 and Cronbach's alpha was 0.97, indicating that the PTIN has excellent day-to-day repeatability and internal consistency. The present results show adequate repeatability, and PTIN assessment using the Vasotens technology appears to be feasible.

Abstract Image

Abstract Image

标准脉冲时间指数的日常重复性。
高血压靶器官损伤的脉搏波速度(PWV)阈值目前设定为每秒10米。新的24小时监测仪(例如,BPLab®和Vasotens®)在24-72小时内提供几种PWV测量。从这些数据中可以计算出一个新的参数,即脉冲时间Norm指数(PTIN)。PTIN定义为24小时内PWV不超过10m / s的百分比。本研究的目的是通过日常重复性分析来测试新的PTIN的临床可行性。振荡生成的波形文件(n=85)以前用于研究,使用新的2013版Vasotens技术程序软件重新分析,该软件可以计算PTIN。类内相关系数为0.98,Cronbach’s alpha为0.97,表明PTIN具有良好的日常重复性和内部一致性。目前的结果显示出足够的可重复性,使用Vasotens技术评估PTIN似乎是可行的。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Medical Devices-Evidence and Research
Medical Devices-Evidence and Research ENGINEERING, BIOMEDICAL-
CiteScore
2.80
自引率
0.00%
发文量
41
审稿时长
16 weeks
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信