{"title":"Biomechanical influence of thread form on stress distribution over short implants (≤6 mm) using finite element analysis.","authors":"Fan Liu, Zhi-Hong Mao, Wujie Peng, Sheng Wen","doi":"10.1515/bmt-2020-0215","DOIUrl":null,"url":null,"abstract":"<p><p>The macro and micro design is essential to the biomechanical performance of a short implant. In this study, the implant thread parameters of short implants used in edentulous maxillae will be discussed. The aim of the study is to analyse biomechanical distinctions in different thread parameters over short implants by applying the vertical or oblique load of 130 N on dental prosthesis. A 6*5 mm implant will be used in posterior maxillae arch, where the molar region locates. The CAD model has been assembled by three parts, a crown, an implant system and a jaw. By applying the vertical or oblique load to the crown, the Von-Mises stresses of cortical bone and trabecular bone will be evaluated in pairs along the lines v1-v2 & a1-a2. The results showed that the reverse buttress thread would induce more stresses in cancellous bone whereas the buttress did the opposite. The trapezoidal thread (V-thread) is more favourable than the reverse buttress thread in accordance to the FEA result. The rectangle threads will induce more uneven stresses in cancellous bones.</p>","PeriodicalId":8900,"journal":{"name":"Biomedical Engineering / Biomedizinische Technik","volume":"67 1","pages":"53-60"},"PeriodicalIF":1.3000,"publicationDate":"2022-01-20","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"1","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Biomedical Engineering / Biomedizinische Technik","FirstCategoryId":"5","ListUrlMain":"https://doi.org/10.1515/bmt-2020-0215","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2022/2/23 0:00:00","PubModel":"Print","JCR":"Q4","JCRName":"ENGINEERING, BIOMEDICAL","Score":null,"Total":0}
引用次数: 1
Abstract
The macro and micro design is essential to the biomechanical performance of a short implant. In this study, the implant thread parameters of short implants used in edentulous maxillae will be discussed. The aim of the study is to analyse biomechanical distinctions in different thread parameters over short implants by applying the vertical or oblique load of 130 N on dental prosthesis. A 6*5 mm implant will be used in posterior maxillae arch, where the molar region locates. The CAD model has been assembled by three parts, a crown, an implant system and a jaw. By applying the vertical or oblique load to the crown, the Von-Mises stresses of cortical bone and trabecular bone will be evaluated in pairs along the lines v1-v2 & a1-a2. The results showed that the reverse buttress thread would induce more stresses in cancellous bone whereas the buttress did the opposite. The trapezoidal thread (V-thread) is more favourable than the reverse buttress thread in accordance to the FEA result. The rectangle threads will induce more uneven stresses in cancellous bones.
期刊介绍:
Biomedical Engineering / Biomedizinische Technik (BMT) is a high-quality forum for the exchange of knowledge in the fields of biomedical engineering, medical information technology and biotechnology/bioengineering. As an established journal with a tradition of more than 60 years, BMT addresses engineers, natural scientists, and clinicians working in research, industry, or clinical practice.