Retinal image quality with multifocal, EDoF, and accommodative intraocular lenses as studied by pyramidal aberrometry.

Jorge L Alio, Francesco D'Oria, Francesca Toto, Joan Balgos, Antonio Palazon, Francesco Versaci, Jorge L Alio Del Barrio
{"title":"Retinal image quality with multifocal, EDoF, and accommodative intraocular lenses as studied by pyramidal aberrometry.","authors":"Jorge L Alio,&nbsp;Francesco D'Oria,&nbsp;Francesca Toto,&nbsp;Joan Balgos,&nbsp;Antonio Palazon,&nbsp;Francesco Versaci,&nbsp;Jorge L Alio Del Barrio","doi":"10.1186/s40662-021-00258-y","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>To study and compare the clinical optical image quality following implantation with different premium IOLs by analysing the point spread function (PSF) Strehl ratio using a pyramidal wavefront sensor (PWS)-based aberrometer.</p><p><strong>Methods: </strong>This study included 194 eyes implanted with: (a) 19 AcrySof SA60AT (control group); (b) 19 Miniwell; (c) 24 LENTIS Mplus LS-313 MF30; d) 33 LENTIS Mplus LS-313 MF15; (e) 17 AkkoLens Lumina; (f) 31 AT LISA Tri 839MP; (g) 20 Precizon Presbyopic; (h) 20 AcrySof IQ PanOptix; (i) 11 Tecnis Eyhance. Main outcome measures were PSF Strehl ratio, PSF Strehl ratio excluding second-order aberrations (PSFw2), total root mean square (RMS), low-order aberration (LOA) and high-order aberration (HOA) RMS measured by PWS aberrometer.</p><p><strong>Results: </strong>AT LISA Tri had the highest PSFw2 Strehl ratio at both 3.0- and 4.0-mm pupil size (0.52 ± 0.14 and 0.31 ± 0.10; P < 0.05), followed by SA60AT (0.41 ± 0.11 and 0.28 ± 0.07) and PanOptix (0.4 ± 0.07 and 0.26 ± 0.04). AT LISA Tri was found to provide a significantly better retinal image quality than PanOptix at both 3.0 mm (P < 0.0001) and 4.0 mm (P = 0.004). Mplus MF15 was found to be significantly better than Mplus MF30 at both 3.0 mm (P < 0.0001) and 4.0 mm (P = 0.002). Total RMS, LOA RMS, HOA RMS, PSF Strehl ratio and PSFw2 varied significantly between the studied groups (P < 0.001).</p><p><strong>Conclusions: </strong>Far distance clinical image quality parameters measured by PWS aberrometer differed significantly according to the technology of the implanted lens. AT LISA Tri, SA60AT and PanOptix showed the highest values of far distance retinal image quality, while the lowest PSFw2 Strehl ratios were displayed by Miniwell, Mplus MF30 and Precizon Presbyopic.</p>","PeriodicalId":520624,"journal":{"name":"Eye and vision (London, England)","volume":" ","pages":"37"},"PeriodicalIF":0.0000,"publicationDate":"2021-10-06","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8496005/pdf/","citationCount":"19","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Eye and vision (London, England)","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1186/s40662-021-00258-y","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 19

Abstract

Background: To study and compare the clinical optical image quality following implantation with different premium IOLs by analysing the point spread function (PSF) Strehl ratio using a pyramidal wavefront sensor (PWS)-based aberrometer.

Methods: This study included 194 eyes implanted with: (a) 19 AcrySof SA60AT (control group); (b) 19 Miniwell; (c) 24 LENTIS Mplus LS-313 MF30; d) 33 LENTIS Mplus LS-313 MF15; (e) 17 AkkoLens Lumina; (f) 31 AT LISA Tri 839MP; (g) 20 Precizon Presbyopic; (h) 20 AcrySof IQ PanOptix; (i) 11 Tecnis Eyhance. Main outcome measures were PSF Strehl ratio, PSF Strehl ratio excluding second-order aberrations (PSFw2), total root mean square (RMS), low-order aberration (LOA) and high-order aberration (HOA) RMS measured by PWS aberrometer.

Results: AT LISA Tri had the highest PSFw2 Strehl ratio at both 3.0- and 4.0-mm pupil size (0.52 ± 0.14 and 0.31 ± 0.10; P < 0.05), followed by SA60AT (0.41 ± 0.11 and 0.28 ± 0.07) and PanOptix (0.4 ± 0.07 and 0.26 ± 0.04). AT LISA Tri was found to provide a significantly better retinal image quality than PanOptix at both 3.0 mm (P < 0.0001) and 4.0 mm (P = 0.004). Mplus MF15 was found to be significantly better than Mplus MF30 at both 3.0 mm (P < 0.0001) and 4.0 mm (P = 0.002). Total RMS, LOA RMS, HOA RMS, PSF Strehl ratio and PSFw2 varied significantly between the studied groups (P < 0.001).

Conclusions: Far distance clinical image quality parameters measured by PWS aberrometer differed significantly according to the technology of the implanted lens. AT LISA Tri, SA60AT and PanOptix showed the highest values of far distance retinal image quality, while the lowest PSFw2 Strehl ratios were displayed by Miniwell, Mplus MF30 and Precizon Presbyopic.

Abstract Image

用锥体像差法研究多焦、EDoF和可调节人工晶状体的视网膜图像质量。
背景:采用基于锥体波前传感器(PWS)的像差仪分析点扩展函数(PSF) Strehl比,研究并比较不同优质iol植入术后的临床光学像质量。方法:194只眼植入术:(a) 19 acryysof SA60AT(对照组);(b) 19个迷你井;(c) 24 LENTIS Mplus LS-313 MF30;d) 33个LENTIS Mplus LS-313 MF15;(e) 17 AkkoLens Lumina;(f) 31 AT LISA Tri 839MP;(g)精密老花眼;(h) 20 acryysof IQ PanOptix;(i) 11 Tecnis Eyhance。主要观察指标为PSF Strehl比、PSF Strehl比排除二阶像差(PSFw2)、总均方根(RMS)、低阶像差(LOA)、高阶像差(HOA) RMS。结果:AT LISA Tri在3.0和4.0 mm瞳孔时PSFw2 Strehl比值最高(0.52±0.14和0.31±0.10);P结论:PWS像差仪测量的远端临床像质量参数因人工晶状体技术的不同而有显著差异。AT LISA Tri、SA60AT和PanOptix的远端视网膜图像质量最高,而Miniwell、Mplus MF30和preczon Presbyopic的PSFw2 Strehl比最低。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信