D K Steckler, R N Goldberg, Y B Tewari, T J Buckley
{"title":"High Precision Microcalorimetry: Apparatus, Procedures, and Biochemical Applications.","authors":"D K Steckler, R N Goldberg, Y B Tewari, T J Buckley","doi":"10.6028/jres.091.018","DOIUrl":null,"url":null,"abstract":"<p><p>Apparatus and procedures used for high-precision microcalorimetric measurements are described. The calorimeter is of the heat-conduction type and utilizes semi-conductor thermoelectric modules. The bicompartmental reaction vessel is made of high-density polyethylene and holds about 0.5 mL of solution in each compartment. Imprecision of heat measurement is 0.2 percent when measuring 300 mJ of heat produced by a rapid chemical reaction. Three microcalorimeters are operated simultaneously using a microcomputer and a data acquisition system. Thermochemical and kinetic applications are described. The acquisition of data from an isoperibol solution calorimeter is also described.</p>","PeriodicalId":93321,"journal":{"name":"Journal of research of the National Bureau of Standards (1977)","volume":"91 3","pages":"113-121"},"PeriodicalIF":0.0000,"publicationDate":"1986-05-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6658430/pdf/jres-91-113.pdf","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of research of the National Bureau of Standards (1977)","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.6028/jres.091.018","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0
Abstract
Apparatus and procedures used for high-precision microcalorimetric measurements are described. The calorimeter is of the heat-conduction type and utilizes semi-conductor thermoelectric modules. The bicompartmental reaction vessel is made of high-density polyethylene and holds about 0.5 mL of solution in each compartment. Imprecision of heat measurement is 0.2 percent when measuring 300 mJ of heat produced by a rapid chemical reaction. Three microcalorimeters are operated simultaneously using a microcomputer and a data acquisition system. Thermochemical and kinetic applications are described. The acquisition of data from an isoperibol solution calorimeter is also described.