{"title":"Potential Induced Radioactivity in Materials Processed with X-ray Energy Above 5 MeV.","authors":"Hervé Michel,&nbsp;Thomas Kroc,&nbsp;Brian J McEvoy,&nbsp;Deepak Patil,&nbsp;Pierre Reppert,&nbsp;Mark A Smith","doi":"10.2345/0899-8205-55.s3.17","DOIUrl":null,"url":null,"abstract":"<p><p>Section 5.1.2 of ANSI/AAMI/ISO 11137-1 states that \"the potential for induced radioactivity in product shall be assessed.\" This article describes how compliance with this requirement may be achieved using qualified test methods. Materials of consideration are conceptually discussed, and results of testing conducted on products processed with a 7.5-MeV X-ray irradiation process are provided. As X-ray becomes more widely used in healthcare sterilization, having standard assessment protocols for activation coupled with a shared database of material test results will benefit manufacturers seeking to utilize this innovative technology.</p>","PeriodicalId":35656,"journal":{"name":"Biomedical Instrumentation and Technology","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"2021-03-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8655703/pdf/i0899-8205-55-s3-17.pdf","citationCount":"2","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Biomedical Instrumentation and Technology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.2345/0899-8205-55.s3.17","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"Medicine","Score":null,"Total":0}
引用次数: 2

Abstract

Section 5.1.2 of ANSI/AAMI/ISO 11137-1 states that "the potential for induced radioactivity in product shall be assessed." This article describes how compliance with this requirement may be achieved using qualified test methods. Materials of consideration are conceptually discussed, and results of testing conducted on products processed with a 7.5-MeV X-ray irradiation process are provided. As X-ray becomes more widely used in healthcare sterilization, having standard assessment protocols for activation coupled with a shared database of material test results will benefit manufacturers seeking to utilize this innovative technology.

用5兆电子伏特以上的x射线处理材料的电位诱导放射性。
ANSI/AAMI/ISO 11137-1第5.1.2节规定“应评估产品中诱发放射性的可能性”。本文描述了如何使用合格的测试方法来满足这一要求。对考虑的材料进行了概念性讨论,并提供了对7.5 mev x射线辐照工艺处理的产品进行的测试结果。随着x射线在医疗保健灭菌中的应用越来越广泛,拥有标准的激活评估方案以及共享的材料测试结果数据库将有利于寻求利用这一创新技术的制造商。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Biomedical Instrumentation and Technology
Biomedical Instrumentation and Technology Computer Science-Computer Networks and Communications
CiteScore
1.10
自引率
0.00%
发文量
16
期刊介绍: AAMI publishes Biomedical Instrumentation & Technology (BI&T) a bi-monthly peer-reviewed journal dedicated to the developers, managers, and users of medical instrumentation and technology.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信