Adverse device-device interaction between pacemaker and subcutaneous implantable cardiac defibrillator.

IF 1.3
Pacing and clinical electrophysiology : PACE Pub Date : 2021-11-01 Epub Date: 2021-07-31 DOI:10.1111/pace.14325
Nicholas Abbott, Aron Bender, Charles Henrikson, Jared Miller, Babak Nazer, Seshadri Balaji
{"title":"Adverse device-device interaction between pacemaker and subcutaneous implantable cardiac defibrillator.","authors":"Nicholas Abbott,&nbsp;Aron Bender,&nbsp;Charles Henrikson,&nbsp;Jared Miller,&nbsp;Babak Nazer,&nbsp;Seshadri Balaji","doi":"10.1111/pace.14325","DOIUrl":null,"url":null,"abstract":"<p><p>A 26-year-old patient with prior surgery for Ebstein's anomaly and a pacemaker (placed for post-surgical heart block and poor underlying rhythm) underwent SICD was placement. During defibrillation testing, device-device interaction led to undersensing of ventricular fibrillation with failure to shock. Increasing the pacemaker sensitivity resolved the problem but post shock pacing was unable to capture the heart after both shocks. The patient underwent removal of both the pacemaker and the SICD and placement of a transvenous ICD. Complex device-device interactions can occur in patients who are pacemaker dependent and undergo placement of a SICD.</p>","PeriodicalId":520740,"journal":{"name":"Pacing and clinical electrophysiology : PACE","volume":" ","pages":"1944-1948"},"PeriodicalIF":1.3000,"publicationDate":"2021-11-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.1111/pace.14325","citationCount":"3","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Pacing and clinical electrophysiology : PACE","FirstCategoryId":"5","ListUrlMain":"https://doi.org/10.1111/pace.14325","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2021/7/31 0:00:00","PubModel":"Epub","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 3

Abstract

A 26-year-old patient with prior surgery for Ebstein's anomaly and a pacemaker (placed for post-surgical heart block and poor underlying rhythm) underwent SICD was placement. During defibrillation testing, device-device interaction led to undersensing of ventricular fibrillation with failure to shock. Increasing the pacemaker sensitivity resolved the problem but post shock pacing was unable to capture the heart after both shocks. The patient underwent removal of both the pacemaker and the SICD and placement of a transvenous ICD. Complex device-device interactions can occur in patients who are pacemaker dependent and undergo placement of a SICD.

起搏器与皮下植入式心脏除颤器之间的不良器械相互作用。
一名26岁的患者因Ebstein异常手术和起搏器(因术后心脏传导阻滞和潜在心律不佳而放置)接受了SICD的放置。在除颤试验中,设备与设备之间的相互作用导致对心室颤动的感知不足,而不能引起电击。提高起搏器的灵敏度解决了这个问题,但在两次冲击后,休克后起搏无法捕捉心脏。患者接受了起搏器和SICD的移除,并放置了经静脉ICD。在依赖起搏器并放置SICD的患者中可能发生复杂的设备-设备相互作用。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信