{"title":"Does a combined screw and dowel construct improve tibial fixation during anterior cruciate ligament reconstruction?","authors":"Pierre Laumonerie, Meagan E Tibbo, Gregoire Laumond, Dominique Barbier, Pauline Assemat, Pascal Swider, Franck Accadbled","doi":"10.1007/s00590-021-03049-2","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>The aims of the present study were to compare the biomechanical properties of tibial fixation in hamstring-graft ACL reconstruction using interference screw and a novel combination interference screw and dowel construct.</p><p><strong>Material and methods: </strong>We compared the fixation of 30 (2- and 4-stranded gracilis and semitendinosis tendons) in 15 fresh-frozen porcine tibiae with a biocomposite resorbable interference screw (Group 1) and a screw and dowel construct (Group 2). Each graft was subjected to load-to-failure testing (50 mm/min) to determine maximum load, displacement at failure and pullout strength.</p><p><strong>Results: </strong>There were no significant differences between the biomechanical properties of the constructs. Multivariate analysis demonstrated that combination constructs (β = 140.20, p = 0.043), screw diameter (β = 185, p = 0.006) and 4-strand grafts (β = 51, p = 0.050) were associated with a significant increase in load at failure. Larger screw diameter was associated with increased construct stiffness (β = 20.15, p = 0.020).</p><p><strong>Conclusion: </strong>The screw and dowel construct led to significantly increased fixation properties compared to interference screws alone in a porcine model. Increased screw diameter and utilization of 4-strand ACL grafts also led to improvement in load-to-failure of the construct. However, this is an in vitro study and additional investigations are needed to determine whether the results are reproducible in vivo.</p><p><strong>Level of evidence: </strong>Level V; Biomechanical study.</p>","PeriodicalId":520619,"journal":{"name":"European journal of orthopaedic surgery & traumatology : orthopedie traumatologie","volume":" ","pages":"759-765"},"PeriodicalIF":1.5000,"publicationDate":"2022-05-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"European journal of orthopaedic surgery & traumatology : orthopedie traumatologie","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1007/s00590-021-03049-2","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2021/6/14 0:00:00","PubModel":"Epub","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0
Abstract
Purpose: The aims of the present study were to compare the biomechanical properties of tibial fixation in hamstring-graft ACL reconstruction using interference screw and a novel combination interference screw and dowel construct.
Material and methods: We compared the fixation of 30 (2- and 4-stranded gracilis and semitendinosis tendons) in 15 fresh-frozen porcine tibiae with a biocomposite resorbable interference screw (Group 1) and a screw and dowel construct (Group 2). Each graft was subjected to load-to-failure testing (50 mm/min) to determine maximum load, displacement at failure and pullout strength.
Results: There were no significant differences between the biomechanical properties of the constructs. Multivariate analysis demonstrated that combination constructs (β = 140.20, p = 0.043), screw diameter (β = 185, p = 0.006) and 4-strand grafts (β = 51, p = 0.050) were associated with a significant increase in load at failure. Larger screw diameter was associated with increased construct stiffness (β = 20.15, p = 0.020).
Conclusion: The screw and dowel construct led to significantly increased fixation properties compared to interference screws alone in a porcine model. Increased screw diameter and utilization of 4-strand ACL grafts also led to improvement in load-to-failure of the construct. However, this is an in vitro study and additional investigations are needed to determine whether the results are reproducible in vivo.