Understanding the Radial Force of Stroke Thrombectomy Devices to Minimize Vessel Wall Injury: Mechanical Bench Testing of the Radial Force Generated by a Novel Braided Thrombectomy Assist Device Compared to Laser-Cut Stent Retrievers in Simulated MCA Vessel Diameters.

Q1 Medicine
Interventional Neurology Pub Date : 2020-01-01 Epub Date: 2019-08-05 DOI:10.1159/000501080
Jeffrey M Katz, Abdullah M Hakoun, Amir R Dehdashti, Alex B Chebl, Vikram Janardhan, Vallabh Janardhan
{"title":"Understanding the Radial Force of Stroke Thrombectomy Devices to Minimize Vessel Wall Injury: Mechanical Bench Testing of the Radial Force Generated by a Novel Braided Thrombectomy Assist Device Compared to Laser-Cut Stent Retrievers in Simulated MCA Vessel Diameters.","authors":"Jeffrey M Katz,&nbsp;Abdullah M Hakoun,&nbsp;Amir R Dehdashti,&nbsp;Alex B Chebl,&nbsp;Vikram Janardhan,&nbsp;Vallabh Janardhan","doi":"10.1159/000501080","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Recent reports have raised various concerns about the risk of vessel wall injury while withdrawing current laser-cut stent retrievers during active strut apposition to the vessel walls. The development of braided thrombectomy assist devices in conjunction with aspiration systems may be gentler on the fragile brain vessels and more optimized with regard to the radial force (RF) for vessel diameters of proximal (M1) and distal (M2) large vessel occlusions (LVOs).</p><p><strong>Methods: </strong>Mechanical bench testing of the RF was performed using a radial compression station mounted on a tensile testing machine. The total RF in newtons (N) generated in vessels with diameters ranging from 2.25 to 3 mm as seen in proximal LVOs (∼M1), and in vessel diameters ranging from 1.5 to 2.24 mm as seen in distal LVOs (∼M2), was measured. The outer diameter of each stent was recorded, and an RF ≤1 N was grouped as \"low,\" while an RF >1 N was grouped as \"high\" for this analysis.</p><p><strong>Results: </strong>The total RFs of all laser-cut stent retrievers were all higher in the simulated M2 vessels (>1 N) than in the M1 vessels (<1 N), whereas the total RFs of the braided thrombectomy assist devices were uniformly low in both the simulated M1 and the simulated M2 vessels.</p><p><strong>Conclusions: </strong>Novel braided thrombectomy assist devices in conjunction with aspiration systems have lower RFs than existing laser-cut stent retrievers in M1 and M2 vessel diameters. Further in vivo studies are needed to delineate the impact of lowering the RF on vessel wall integrity.</p>","PeriodicalId":46280,"journal":{"name":"Interventional Neurology","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"2020-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.1159/000501080","citationCount":"16","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Interventional Neurology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1159/000501080","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2019/8/5 0:00:00","PubModel":"Epub","JCR":"Q1","JCRName":"Medicine","Score":null,"Total":0}
引用次数: 16

Abstract

Background: Recent reports have raised various concerns about the risk of vessel wall injury while withdrawing current laser-cut stent retrievers during active strut apposition to the vessel walls. The development of braided thrombectomy assist devices in conjunction with aspiration systems may be gentler on the fragile brain vessels and more optimized with regard to the radial force (RF) for vessel diameters of proximal (M1) and distal (M2) large vessel occlusions (LVOs).

Methods: Mechanical bench testing of the RF was performed using a radial compression station mounted on a tensile testing machine. The total RF in newtons (N) generated in vessels with diameters ranging from 2.25 to 3 mm as seen in proximal LVOs (∼M1), and in vessel diameters ranging from 1.5 to 2.24 mm as seen in distal LVOs (∼M2), was measured. The outer diameter of each stent was recorded, and an RF ≤1 N was grouped as "low," while an RF >1 N was grouped as "high" for this analysis.

Results: The total RFs of all laser-cut stent retrievers were all higher in the simulated M2 vessels (>1 N) than in the M1 vessels (<1 N), whereas the total RFs of the braided thrombectomy assist devices were uniformly low in both the simulated M1 and the simulated M2 vessels.

Conclusions: Novel braided thrombectomy assist devices in conjunction with aspiration systems have lower RFs than existing laser-cut stent retrievers in M1 and M2 vessel diameters. Further in vivo studies are needed to delineate the impact of lowering the RF on vessel wall integrity.

了解中风取栓装置的径向力以最大限度地减少血管壁损伤:在模拟MCA血管直径中,与激光切割支架回收器相比,一种新型编织取栓辅助装置产生的径向力的机械台测试。
背景:最近的报道提出了各种关于血管壁损伤风险的担忧,当在主动支架与血管壁相对置时,取出当前的激光切割支架回收器。结合抽吸系统的编织取栓辅助装置的发展可能对脆弱的脑血管更温和,并且对于近端(M1)和远端(M2)大血管闭塞(LVOs)的血管直径的径向力(RF)更优化。方法:采用安装在拉力试验机上的径向压缩站对射频进行机械台架试验。测量了近端LVOs (~ M1)和远端LVOs (~ M2)中直径为1.5 ~ 2.24 mm的血管中产生的总RF (N)。记录每个支架的外径,将RF≤1 N的归为“低”,将RF >1 N的归为“高”。结果:在模拟的M2血管(>1 N)中,所有激光切割支架取物器的总rf均高于M1血管(结论:在M1和M2血管直径中,新型编织取栓辅助装置与抽吸系统的rf低于现有激光切割支架取物器。需要进一步的体内研究来描述降低射频对血管壁完整性的影响。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Interventional Neurology
Interventional Neurology CLINICAL NEUROLOGY-
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信