Prospective Comparison of Transient Elastography Using Two Different Devices: Performance of FibroScan and FibroTouch.

IF 2.6 Q2 GASTROENTEROLOGY & HEPATOLOGY
Hepatic Medicine : Evidence and Research Pub Date : 2020-03-27 eCollection Date: 2020-01-01 DOI:10.2147/HMER.S245455
Joao Tiago Serra, Johannes Mueller, Haidong Teng, Omar Elshaarawy, Sebastian Mueller
{"title":"Prospective Comparison of Transient Elastography Using Two Different Devices: Performance of FibroScan and FibroTouch.","authors":"Joao Tiago Serra,&nbsp;Johannes Mueller,&nbsp;Haidong Teng,&nbsp;Omar Elshaarawy,&nbsp;Sebastian Mueller","doi":"10.2147/HMER.S245455","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>Transient elastography (TE) using FibroScan (FS) has been established to non-invasively assess liver fibrosis and steatosis. The aim of this study was to compare the recently introduced FibroTouch (FT) device with the established FS with respect to liver stiffness and CAP.</p><p><strong>Patients and methods: </strong>Thirty-nine patients with and without liver disease were included. All patients were measured three times with FS (FibroScan 530 compact, Echosens, France) and FT (FibroTouch-FT100, Wuxi Hisky Med, China). For FS, M and XL probe were used according to the manufacturer's specifications. For steatosis, CAP and the comparable FT equivalent UAP (ultrasound attenuation parameter) was determined. Finally, FT and FS were explored in liver tissue-mimicking phantoms.</p><p><strong>Results: </strong>LS between FS and FT correlated well with r=0.91. Root-mean-square (RMS) of the coefficient of variation for LS was better in FS (11.1% vs 27.4%). Bland-Altman analysis showed a 3.1 kPa mean overestimation of LS by FT. In addition, UAP strongly and linearly depended on the BMI following UAP=3.02 × BMI+186. In phantoms, a similar relation was found with UAP (phantom)= 3.78 × BMI + 146 suggesting that UAP is directly calculated from entered BMI instead of assessing shear-wave attenuation. Consequently, RMS-CV was lower for FT (6.0% vs 9.7%). However, if using different BMI, CV-RMS for FT increased to 12.7%. LS of a patient with manifest liver cirrhosis and ascites was 38.8 kPa using the FS-XL probe but almost normal with FT (7.2 kPa).</p><p><strong>Conclusion: </strong>Although LS by FT shows good correlation with LS-FS, it has larger variation, continuously overestimates LS and completely fails in ascites. Moreover, FT-UAP seems to be a misleading parameter for steatosis assessment because it is at least in part calculated from mandatory entered patient data. In conclusion, novel LS cut-off values need to be defined for LS-FT and usage of UAP is not recommended.</p>","PeriodicalId":12917,"journal":{"name":"Hepatic Medicine : Evidence and Research","volume":"12 ","pages":"41-48"},"PeriodicalIF":2.6000,"publicationDate":"2020-03-27","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.2147/HMER.S245455","citationCount":"7","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Hepatic Medicine : Evidence and Research","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.2147/HMER.S245455","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2020/1/1 0:00:00","PubModel":"eCollection","JCR":"Q2","JCRName":"GASTROENTEROLOGY & HEPATOLOGY","Score":null,"Total":0}
引用次数: 7

Abstract

Purpose: Transient elastography (TE) using FibroScan (FS) has been established to non-invasively assess liver fibrosis and steatosis. The aim of this study was to compare the recently introduced FibroTouch (FT) device with the established FS with respect to liver stiffness and CAP.

Patients and methods: Thirty-nine patients with and without liver disease were included. All patients were measured three times with FS (FibroScan 530 compact, Echosens, France) and FT (FibroTouch-FT100, Wuxi Hisky Med, China). For FS, M and XL probe were used according to the manufacturer's specifications. For steatosis, CAP and the comparable FT equivalent UAP (ultrasound attenuation parameter) was determined. Finally, FT and FS were explored in liver tissue-mimicking phantoms.

Results: LS between FS and FT correlated well with r=0.91. Root-mean-square (RMS) of the coefficient of variation for LS was better in FS (11.1% vs 27.4%). Bland-Altman analysis showed a 3.1 kPa mean overestimation of LS by FT. In addition, UAP strongly and linearly depended on the BMI following UAP=3.02 × BMI+186. In phantoms, a similar relation was found with UAP (phantom)= 3.78 × BMI + 146 suggesting that UAP is directly calculated from entered BMI instead of assessing shear-wave attenuation. Consequently, RMS-CV was lower for FT (6.0% vs 9.7%). However, if using different BMI, CV-RMS for FT increased to 12.7%. LS of a patient with manifest liver cirrhosis and ascites was 38.8 kPa using the FS-XL probe but almost normal with FT (7.2 kPa).

Conclusion: Although LS by FT shows good correlation with LS-FS, it has larger variation, continuously overestimates LS and completely fails in ascites. Moreover, FT-UAP seems to be a misleading parameter for steatosis assessment because it is at least in part calculated from mandatory entered patient data. In conclusion, novel LS cut-off values need to be defined for LS-FT and usage of UAP is not recommended.

Abstract Image

Abstract Image

Abstract Image

使用两种不同设备的瞬态弹性成像的前瞻性比较:纤维扫描和纤维触摸的性能。
目的:使用纤维扫描(FS)的瞬时弹性成像(TE)已被建立用于无创评估肝纤维化和脂肪变性。本研究的目的是比较最近推出的FibroTouch (FT)设备与已建立的FS在肝脏硬度和cap方面的差异。患者和方法:包括39名患有和非肝病的患者。所有患者用FS (FibroScan 530 compact, Echosens,法国)和FT (FibroTouch-FT100, Wuxi Hisky Med,中国)测量三次。对于FS,根据制造商的规格使用M和XL探头。对于脂肪变性,测定CAP和可比较的FT等效UAP(超声衰减参数)。最后,在肝组织模拟模型中探讨FT和FS。结果:FS与FT之间的LS相关性良好,r=0.91。LS的变异系数均方根(RMS)优于FS (11.1% vs 27.4%)。Bland-Altman分析显示,FT对LS的平均高估为3.1 kPa。此外,UAP=3.02 × BMI+186后,UAP与BMI呈强烈线性关系。在幻像中,UAP(幻像)= 3.78 × BMI + 146也发现了类似的关系,这表明UAP是直接从输入的BMI计算的,而不是评估剪切波衰减。因此,FT的RMS-CV较低(6.0% vs 9.7%)。然而,如果使用不同的BMI, FT的CV-RMS增加到12.7%。有明显肝硬化和腹水的患者,FS-XL探头的LS为38.8 kPa, FT几乎正常(7.2 kPa)。结论:LS by FT虽然与LS- fs有较好的相关性,但变异较大,持续高估LS,在腹水中完全失效。此外,FT-UAP似乎是脂肪变性评估的一个误导性参数,因为它至少部分是根据强制输入的患者数据计算的。总之,需要为LS- ft定义新的LS截止值,不建议使用UAP。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Hepatic Medicine : Evidence and Research
Hepatic Medicine : Evidence and Research GASTROENTEROLOGY & HEPATOLOGY-
自引率
0.00%
发文量
15
审稿时长
16 weeks
期刊介绍: Hepatic Medicine: Evidence and Research is an international, peer-reviewed, open access, online journal. Publishing original research, reports, editorials, reviews and commentaries on all aspects of adult and pediatric hepatology in the clinic and laboratory including the following topics: Pathology, pathophysiology of hepatic disease Investigation and treatment of hepatic disease Pharmacology of drugs used for the treatment of hepatic disease Although the main focus of the journal is to publish research and clinical results in humans; preclinical, animal and in vitro studies will be published where they will shed light on disease processes and potential new therapies. Issues of patient safety and quality of care will also be considered. As of 1st April 2019, Hepatic Medicine: Evidence and Research will no longer consider meta-analyses for publication.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信