Theresa V Feddersen, Pejman Rowshanfarzad, Tamara N Abel, Martin A Ebert
{"title":"Commissioning and performance characteristics of a pre-clinical image-guided radiotherapy system.","authors":"Theresa V Feddersen, Pejman Rowshanfarzad, Tamara N Abel, Martin A Ebert","doi":"10.1007/s13246-019-00755-4","DOIUrl":null,"url":null,"abstract":"<p><p>Characteristics of a small-animal radiotherapy device, the X-RAD SmART, are described following commissioning of the device for pre-clinical radiotherapy research. Performance characteristics were assessed using published standards and compared with previous results published for similar systems. Operational radiation safety was established. Device X-ray beam quality and output dose-rate were found to be consistent with those reported for similar devices. Output steadily declined over 18 months though remained within tolerance levels. There is considerable variation in output factor across the international installations for the smallest field size (varying by more than 30% for 2.5 mm diameter fields). Measured depth dose and profile data was mostly consistent with that published, with some differences in penumbrae and generally reduced flatness. Target localisation is achieved with an imaging panel and with automatic corrections for panel flex and device mechanical instability, targeting within 0.2 mm is achievable. The small-animal image-guided radiotherapy platform has been implemented and assessed and found to perform as specified. The combination of kV energy and high spatial precision make it suitable for replicating clinical dose distributions at the small-animal scale, though dosimetric uncertainties for the narrowest fields need to be acknowledged.</p>","PeriodicalId":55430,"journal":{"name":"Australasian Physical & Engineering Sciences in Medicine","volume":"42 2","pages":"541-551"},"PeriodicalIF":0.0000,"publicationDate":"2019-06-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6557883/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Australasian Physical & Engineering Sciences in Medicine","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1007/s13246-019-00755-4","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2019/4/15 0:00:00","PubModel":"Epub","JCR":"Q3","JCRName":"Biochemistry, Genetics and Molecular Biology","Score":null,"Total":0}
引用次数: 0
Abstract
Characteristics of a small-animal radiotherapy device, the X-RAD SmART, are described following commissioning of the device for pre-clinical radiotherapy research. Performance characteristics were assessed using published standards and compared with previous results published for similar systems. Operational radiation safety was established. Device X-ray beam quality and output dose-rate were found to be consistent with those reported for similar devices. Output steadily declined over 18 months though remained within tolerance levels. There is considerable variation in output factor across the international installations for the smallest field size (varying by more than 30% for 2.5 mm diameter fields). Measured depth dose and profile data was mostly consistent with that published, with some differences in penumbrae and generally reduced flatness. Target localisation is achieved with an imaging panel and with automatic corrections for panel flex and device mechanical instability, targeting within 0.2 mm is achievable. The small-animal image-guided radiotherapy platform has been implemented and assessed and found to perform as specified. The combination of kV energy and high spatial precision make it suitable for replicating clinical dose distributions at the small-animal scale, though dosimetric uncertainties for the narrowest fields need to be acknowledged.
期刊介绍:
Australasian Physical & Engineering Sciences in Medicine (APESM) is a multidisciplinary forum for information and research on the application of physics and engineering to medicine and human physiology. APESM covers a broad range of topics that include but is not limited to:
- Medical physics in radiotherapy
- Medical physics in diagnostic radiology
- Medical physics in nuclear medicine
- Mathematical modelling applied to medicine and human biology
- Clinical biomedical engineering
- Feature extraction, classification of EEG, ECG, EMG, EOG, and other biomedical signals;
- Medical imaging - contributions to new and improved methods;
- Modelling of physiological systems
- Image processing to extract information from images, e.g. fMRI, CT, etc.;
- Biomechanics, especially with applications to orthopaedics.
- Nanotechnology in medicine
APESM offers original reviews, scientific papers, scientific notes, technical papers, educational notes, book reviews and letters to the editor.
APESM is the journal of the Australasian College of Physical Scientists and Engineers in Medicine, and also the official journal of the College of Biomedical Engineers, Engineers Australia and the Asia-Oceania Federation of Organizations for Medical Physics.