Comparative Evaluation of Retentive Properties of Two Compatible Ball Attachments in Mandibular Implant-Retained Overdentures: An In Vitro Study.

Maryam Memarian, Simindokht Zarrati, Sedigheh Karimi, Mehran Bahrami
{"title":"Comparative Evaluation of Retentive Properties of Two Compatible Ball Attachments in Mandibular Implant-Retained Overdentures: An In Vitro Study.","authors":"Maryam Memarian,&nbsp;Simindokht Zarrati,&nbsp;Sedigheh Karimi,&nbsp;Mehran Bahrami","doi":"","DOIUrl":null,"url":null,"abstract":"<p><strong>Objectives: </strong>The retentive properties of implant-retained overdentures (IRO) may be influenced by the type of attachments. The aim of this research was to compare the retention of two dental implant systems with compatible ball attachments, namely Straumann® system (SS) and Rhein83 SRL system (RS) after fatigue testing.</p><p><strong>Materials and methods: </strong>Two laboratory models consisting of two parallel Straumann® fixtures at a distance of 22 mm were prepared. Five pairs of each systems' ball attachments were examined (n=5). The samples were soaked in artificial saliva. The retention strength values (RSV) were recorded before the fatigue test and after 1100, 2200, 3300, 4400, and 5500 insertion and removal cycles at a speed of 51 mm/minute with a 50-N load cell in a universal testing machine. The data were analyzed by repeated measures analysis of variance (ANOVA) followed by independent sample t-test with Bonferroni corrections.</p><p><strong>Results: </strong>There was a decrease in the RSV in both systems after 5500 cycles of insertion and removal. There was a significant statistical difference between the RSV of the normal Sphero Block of the RS (17.52±0.68 N) and that of the Spare Lamella retention inserts of the SS (19.72±0.74 N, P=0.001).</p><p><strong>Conclusions: </strong>Although the RSVs of the RS and SS were almost similar before the fatigue test, as the number of insertion and removal cycles increased, the RSV decreased more significantly in the RS compared to the SS.</p>","PeriodicalId":30286,"journal":{"name":"Journal of Dentistry of Tehran University of Medical Sciences","volume":"15 2","pages":"106-115"},"PeriodicalIF":0.0000,"publicationDate":"2018-03-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6026313/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Dentistry of Tehran University of Medical Sciences","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Objectives: The retentive properties of implant-retained overdentures (IRO) may be influenced by the type of attachments. The aim of this research was to compare the retention of two dental implant systems with compatible ball attachments, namely Straumann® system (SS) and Rhein83 SRL system (RS) after fatigue testing.

Materials and methods: Two laboratory models consisting of two parallel Straumann® fixtures at a distance of 22 mm were prepared. Five pairs of each systems' ball attachments were examined (n=5). The samples were soaked in artificial saliva. The retention strength values (RSV) were recorded before the fatigue test and after 1100, 2200, 3300, 4400, and 5500 insertion and removal cycles at a speed of 51 mm/minute with a 50-N load cell in a universal testing machine. The data were analyzed by repeated measures analysis of variance (ANOVA) followed by independent sample t-test with Bonferroni corrections.

Results: There was a decrease in the RSV in both systems after 5500 cycles of insertion and removal. There was a significant statistical difference between the RSV of the normal Sphero Block of the RS (17.52±0.68 N) and that of the Spare Lamella retention inserts of the SS (19.72±0.74 N, P=0.001).

Conclusions: Although the RSVs of the RS and SS were almost similar before the fatigue test, as the number of insertion and removal cycles increased, the RSV decreased more significantly in the RS compared to the SS.

Abstract Image

Abstract Image

Abstract Image

两种相容球状附着体在下颌种植覆盖义齿固位性能的体外比较研究。
目的:种植固位覆盖义齿的固位性能可能受到附着体类型的影响。本研究的目的是比较两种具有兼容球附件的种植体系统,即Straumann®系统(SS)和Rhein83 SRL系统(RS)在疲劳测试后的固位。材料和方法:制备了两个实验室模型,由两个平行的Straumann®夹具组成,距离为22 mm。检查了每个系统的五对球附件(n=5)。样品浸泡在人工唾液中。在通用试验机中,采用50-N称重传感器,以51 mm/min的速度,记录疲劳试验前和1100、2200、3300、4400和5500次插入和取出循环后的保持强度值(RSV)。资料分析采用重复测量方差分析(ANOVA),然后进行Bonferroni校正的独立样本t检验。结果:两种系统的RSV在5500次插入和取出循环后均有所下降。RS的正常Sphero Block的RSV(17.52±0.68 N)与SS的Spare Lamella retention inserts的RSV(19.72±0.74 N, P=0.001)差异有统计学意义。结论:虽然在疲劳试验前RS和SS的RSV几乎相似,但随着插入和取出周期次数的增加,RS的RSV比SS的RSV下降更明显。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
审稿时长
24 weeks
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信