Tushita Patel, Kelly Klanian, Zongyi Gong, Mark B Williams
{"title":"Detective Quantum Efficiency of a CsI-CMOS X-ray Detector for Breast Tomosynthesis Operating in High Dynamic Range and High Sensitivity Modes.","authors":"Tushita Patel, Kelly Klanian, Zongyi Gong, Mark B Williams","doi":"10.1007/978-3-642-31271-7_11","DOIUrl":null,"url":null,"abstract":"<p><p>The spatial frequency dependent detective quantum efficiency (DQE) of a CsI-CMOS x-ray detector was measured in two operating modes: a high dynamic range (HDR) mode and a high sensitivity (HS) mode. DQE calculations were performed using the IEC-62220-1-2 Standard. For detector entrance air kerma values between ~7 µGy and 60 µGy the DQE is similar in either HDR mode or HS mode, with a value of ~0.7 at low frequency and ~ 0.15 - 0.20 at the Nyquist frequency f<sub>N</sub> = 6.7 mm<sup>-1</sup>. In HDR mode the DQE remains virtually constant for operation with K<sub>a</sub> values between ~7 µGy and 119 µGy but decreases for K<sub>a</sub> levels below ~ 7 µGy. In HS mode the DQE is approximately constant over the full range of entrance air kerma tested between 1.7 µGy and 60 µGy but kerma values above ~75 µGy produce hard saturation. Quantum limited operation in HS mode for entrance kerma as small as 1.7 µGy makes it possible to use a large number of low dose views to improve angular sampling and decrease acquisition time.</p>","PeriodicalId":92005,"journal":{"name":"Breast imaging : 11th International Workshop, IWDM 2012, Philadelphia, PA, USA, July 8-11, 2012 : proceedings. International Workshop on Breast Imaging (11th : 2012 : Philadelphia, Pa.)","volume":"7361 ","pages":"80-87"},"PeriodicalIF":0.0000,"publicationDate":"2012-07-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.1007/978-3-642-31271-7_11","citationCount":"11","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Breast imaging : 11th International Workshop, IWDM 2012, Philadelphia, PA, USA, July 8-11, 2012 : proceedings. International Workshop on Breast Imaging (11th : 2012 : Philadelphia, Pa.)","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1007/978-3-642-31271-7_11","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 11
Abstract
The spatial frequency dependent detective quantum efficiency (DQE) of a CsI-CMOS x-ray detector was measured in two operating modes: a high dynamic range (HDR) mode and a high sensitivity (HS) mode. DQE calculations were performed using the IEC-62220-1-2 Standard. For detector entrance air kerma values between ~7 µGy and 60 µGy the DQE is similar in either HDR mode or HS mode, with a value of ~0.7 at low frequency and ~ 0.15 - 0.20 at the Nyquist frequency fN = 6.7 mm-1. In HDR mode the DQE remains virtually constant for operation with Ka values between ~7 µGy and 119 µGy but decreases for Ka levels below ~ 7 µGy. In HS mode the DQE is approximately constant over the full range of entrance air kerma tested between 1.7 µGy and 60 µGy but kerma values above ~75 µGy produce hard saturation. Quantum limited operation in HS mode for entrance kerma as small as 1.7 µGy makes it possible to use a large number of low dose views to improve angular sampling and decrease acquisition time.