Fixed-flexion knee radiography using a new positioning device produced highly repeatable measurements of joint space width: ELSA-Brasil Musculoskeletal Study (ELSA-Brasil MSK)

Rosa Weiss Telles, Luciana Costa-Silva, Luciana A.C. Machado, Rodrigo Citton Padilha dos Reis, Sandhi Maria Barreto
{"title":"Fixed-flexion knee radiography using a new positioning device produced highly repeatable measurements of joint space width: ELSA-Brasil Musculoskeletal Study (ELSA-Brasil MSK)","authors":"Rosa Weiss Telles,&nbsp;Luciana Costa-Silva,&nbsp;Luciana A.C. Machado,&nbsp;Rodrigo Citton Padilha dos Reis,&nbsp;Sandhi Maria Barreto","doi":"10.1016/j.rbre.2016.11.010","DOIUrl":null,"url":null,"abstract":"<div><h3>Objective</h3><p>To describe the performance of a non-fluoroscopic fixed-flexion PA radiographic protocol with a new positioning device, developed for the assessment of knee osteoarthritis (OA) in Brazilian Longitudinal Study of Adult Health Musculoskeletal Study (ELSA-Brasil MSK).</p></div><div><h3>Material and methods</h3><p>A test–retest design including 19 adults (38 knee images) was conducted. Feasibility of the radiographic protocol was assessed by image quality parameters and presence of radioanatomic alignment according to intermargin distance (IMD) values. Repeatability was assessed for IMD and joint space width (JSW) measured at three different locations.</p></div><div><h3>Results</h3><p>Approximately 90% of knee images presented excellent quality. Frequencies of nearly perfect radioanatomic alignment (IMD ≤1<!--> <!-->mm) ranged from 29% to 50%, and satisfactory alignment was found in up to 71% and 76% of the images (IMD ≤1.5<!--> <!-->mm and ≤1.7<!--> <!-->mm, respectively). Repeatability analyses yielded the following results: IMD [SD of mean difference<!--> <!-->=<!--> <!-->1.08; coefficient of variation (%CV)<!--> <!-->=<!--> <!-->54.68%; intraclass correlation coefficient (ICC) (95%CI)<!--> <!-->=<!--> <!-->0.59 (0.34–0.77)]; JSW [SD of mean difference<!--> <!-->=<!--> <!-->0.34–0.61; %CV<!--> <!-->=<!--> <!-->4.48%–9.80%; ICC (95%CI)<!--> <!-->=<!--> <!-->0.74 (0.55–0.85)–0.94 (0.87–0.97)]. Adequately reproducible measurements of IMD and JSW were found in 68% and 87% of the images, respectively.</p></div><div><h3>Conclusions</h3><p>Despite the difficulty in achieving consistent radioanatomic alignment between subsequent radiographs in terms of IMD, the protocol produced highly repeatable JSW measurements when these were taken at midpoint and 10<!--> <!-->mm from the medial extremity of the medial tibial plateau. Therefore, measurements of JSW at these locations can be considered adequate for the assessment of knee OA in ELSA-Brasil MSK.</p></div>","PeriodicalId":101096,"journal":{"name":"Revista Brasileira de Reumatologia (English Edition)","volume":"57 2","pages":"Pages 154-161"},"PeriodicalIF":0.0000,"publicationDate":"2017-03-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.1016/j.rbre.2016.11.010","citationCount":"1","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Revista Brasileira de Reumatologia (English Edition)","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S2255502116301079","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 1

Abstract

Objective

To describe the performance of a non-fluoroscopic fixed-flexion PA radiographic protocol with a new positioning device, developed for the assessment of knee osteoarthritis (OA) in Brazilian Longitudinal Study of Adult Health Musculoskeletal Study (ELSA-Brasil MSK).

Material and methods

A test–retest design including 19 adults (38 knee images) was conducted. Feasibility of the radiographic protocol was assessed by image quality parameters and presence of radioanatomic alignment according to intermargin distance (IMD) values. Repeatability was assessed for IMD and joint space width (JSW) measured at three different locations.

Results

Approximately 90% of knee images presented excellent quality. Frequencies of nearly perfect radioanatomic alignment (IMD ≤1 mm) ranged from 29% to 50%, and satisfactory alignment was found in up to 71% and 76% of the images (IMD ≤1.5 mm and ≤1.7 mm, respectively). Repeatability analyses yielded the following results: IMD [SD of mean difference = 1.08; coefficient of variation (%CV) = 54.68%; intraclass correlation coefficient (ICC) (95%CI) = 0.59 (0.34–0.77)]; JSW [SD of mean difference = 0.34–0.61; %CV = 4.48%–9.80%; ICC (95%CI) = 0.74 (0.55–0.85)–0.94 (0.87–0.97)]. Adequately reproducible measurements of IMD and JSW were found in 68% and 87% of the images, respectively.

Conclusions

Despite the difficulty in achieving consistent radioanatomic alignment between subsequent radiographs in terms of IMD, the protocol produced highly repeatable JSW measurements when these were taken at midpoint and 10 mm from the medial extremity of the medial tibial plateau. Therefore, measurements of JSW at these locations can be considered adequate for the assessment of knee OA in ELSA-Brasil MSK.

固定屈曲膝关节x线摄影使用一种新的定位装置产生高度可重复的关节间隙宽度测量:ELSA-Brasil肌肉骨骼研究(ELSA-Brasil MSK)
目的描述巴西成人健康肌肉骨骼纵向研究(ELSA-Brasil MSK)中用于评估膝关节骨性关节炎(OA)的一种新型定位装置的非透视固定屈曲PA x线摄影方案的性能。材料与方法对19例成人(38张膝关节图像)进行重测设计。通过图像质量参数和根据边缘距离(IMD)值是否存在放射解剖对准来评估放射照相方案的可行性。评估了在三个不同位置测量的IMD和关节间隙宽度(JSW)的重复性。结果约90%的膝关节图像质量良好。接近完美的放射解剖学对准(IMD≤1 mm)的频率为29%至50%,满意的对准在高达71%和76%的图像中发现(IMD分别≤1.5 mm和≤1.7 mm)。重复性分析结果如下:IMD[均数差的标准差= 1.08;变异系数(%CV) = 54.68%;类内相关系数(ICC) (95%CI) = 0.59 (0.34 ~ 0.77);JSW[平均差值SD = 0.34-0.61;% cv = 4.48%-9.80%;国际刑事法庭(95% ci) = 0.74(0.55 - -0.85) -0.94(0.87 - -0.97)]。在68%和87%的图像中分别发现了足够可重复的IMD和JSW测量。结论:尽管在IMD的后续x线片之间难以实现一致的放射解剖对齐,但该方案在距胫骨内侧平台内侧端中点和10mm处进行了高度可重复的JSW测量。因此,在这些位置测量JSW可以被认为是足够的评估ELSA-Brasil MSK膝关节OA。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信