Noninvasive Measurements of Pressure for Detecting Compartment Syndromes.

Samantha H Lee, Michael Padilla, John E Lynch, Alan R Hargens
{"title":"Noninvasive Measurements of Pressure for Detecting Compartment Syndromes.","authors":"Samantha H Lee,&nbsp;Michael Padilla,&nbsp;John E Lynch,&nbsp;Alan R Hargens","doi":"","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>We tested a noninvasive ultrasound, Pulse Phase Locked Loop (PPLL) technique for estimating Intramuscular Pressure (IMP) in a model Acute Compartment Syndrome (ACS); and compared it to a Near-Infrared Spectroscopy (NIRS) method.</p><p><strong>Questions: </strong>We wanted to validate our model compartment syndrome, and to compare and validate the PPLL and NIRS methods of detecting compartment syndrome.</p><p><strong>Methods: </strong>To simulate the tamponade of an ACS, external-pressure levels from 10 to 70 mm Hg were applied to the legs of 15 healthy adult subjects to raise their IMP. Receiver Operator Characteristic (ROC) curves were used to determine the sensitivity and specificity of diagnosing elevated IMP by the two noninvasive techniques.</p><p><strong>Results: </strong>NIRS data varied significantly with compression (p=0.003) with large subject-to-subject variability (p<0.001). PPLL data also varied significantly with compression (p=0.004), but subject-to-subject variation was not significant (p=0.47), suggesting that individual variation does not affect the diagnostic accuracy of the PPLL technique. Sensitivity and specificity for diagnosing normal IMP by the PPLL (<30 mm Hg from a slit catheter reading) from elevated IMP (>30 mm Hg) were 0.75 and 0.75, respectively, and the area under the curve (AUC) was 0.78. For the NIRS, the sensitivity and specificity were 0.65 and 0.65, respectively, and the AUC was 0.68.</p><p><strong>Conclusions: </strong>Both NIRS and PPLL recordings are able to differentiate a simulated ACS up to 70 mm Hg. However, the PPLL technique is a slightly better diagnostic predictor than NIRS with less subject-to-subject variability and slightly better sensitivity and specificity.</p><p><strong>Level of evidence: </strong>Level II, Diagnostic test.</p>","PeriodicalId":90612,"journal":{"name":"Journal of orthopedics & rheumatology","volume":"1 1","pages":"5"},"PeriodicalIF":0.0000,"publicationDate":"2013-12-21","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4197938/pdf/nihms593633.pdf","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of orthopedics & rheumatology","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Background: We tested a noninvasive ultrasound, Pulse Phase Locked Loop (PPLL) technique for estimating Intramuscular Pressure (IMP) in a model Acute Compartment Syndrome (ACS); and compared it to a Near-Infrared Spectroscopy (NIRS) method.

Questions: We wanted to validate our model compartment syndrome, and to compare and validate the PPLL and NIRS methods of detecting compartment syndrome.

Methods: To simulate the tamponade of an ACS, external-pressure levels from 10 to 70 mm Hg were applied to the legs of 15 healthy adult subjects to raise their IMP. Receiver Operator Characteristic (ROC) curves were used to determine the sensitivity and specificity of diagnosing elevated IMP by the two noninvasive techniques.

Results: NIRS data varied significantly with compression (p=0.003) with large subject-to-subject variability (p<0.001). PPLL data also varied significantly with compression (p=0.004), but subject-to-subject variation was not significant (p=0.47), suggesting that individual variation does not affect the diagnostic accuracy of the PPLL technique. Sensitivity and specificity for diagnosing normal IMP by the PPLL (<30 mm Hg from a slit catheter reading) from elevated IMP (>30 mm Hg) were 0.75 and 0.75, respectively, and the area under the curve (AUC) was 0.78. For the NIRS, the sensitivity and specificity were 0.65 and 0.65, respectively, and the AUC was 0.68.

Conclusions: Both NIRS and PPLL recordings are able to differentiate a simulated ACS up to 70 mm Hg. However, the PPLL technique is a slightly better diagnostic predictor than NIRS with less subject-to-subject variability and slightly better sensitivity and specificity.

Level of evidence: Level II, Diagnostic test.

Abstract Image

Abstract Image

Abstract Image

无创压力测量检测筋膜室综合征。
背景:我们测试了一种无创超声,脉冲锁相环(PPLL)技术来估计模型急性筋膜间室综合征(ACS)的肌内压力(IMP);并将其与近红外光谱(NIRS)方法进行了比较。问题:我们想验证我们的模型室综合征,并比较和验证PPLL和NIRS检测室综合征的方法。方法:模拟ACS的填塞,对15名健康成人受试者施加10 ~ 70 mm Hg的外压水平以提高IMP,采用受试者操作者特征曲线(Receiver Operator Characteristic, ROC)来确定两种无创技术诊断IMP升高的敏感性和特异性。结果:NIRS数据随压缩变化显著(p=0.003),受试者间差异较大(p30 mm Hg)分别为0.75和0.75,曲线下面积(AUC)为0.78。近红外光谱的敏感性和特异性分别为0.65和0.65,AUC为0.68。结论:NIRS和PPLL记录都能够区分高达70 mm Hg的模拟ACS。然而,PPLL技术比NIRS诊断预测指标稍好,受试者之间的可变性较小,敏感性和特异性稍好。证据等级:II级,诊断测试。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书