Ikuo Yamaguchi, Tomoe Aoyama, Masaru Yamamoto, Keiko Kinosita, Yumi Ito
{"title":"[Evaluation of the sensitivity of a densitometry system, in judging the result of influenza virus antigen-detection kit using immunochromatography].","authors":"Ikuo Yamaguchi, Tomoe Aoyama, Masaru Yamamoto, Keiko Kinosita, Yumi Ito","doi":"","DOIUrl":null,"url":null,"abstract":"<p><p>Immunochromatography viral antigen-detection kits have become popular in clinical settings in Japan. Influenza virus detection kit is one of them. It is sometimes used in early phase of the disease, combined with the early treatment with anti-influenza drugs. Most of them are invented to visually read the test line on their kits. However, we should be careful about their reliability of them. Sometimes human errors occur at the visual tests, and they have different sensitivities among the kits from different companies. In this report, we evaluated the sensitivity of BD Veritor System Flu with its reader by comparing with conventional visual tests. A total of 84 people including laboratory technologists were asked to visually read test line and their answers were compared with results of BD Veritor System Reader. This study showed that the lower the concentration of standard sample was applied, the greater the error ratio of visual test became, indicating the stable sensitivity of Veritor System. Moreover, the sensitivity was compared with three other major products approved in Japan, using four influenza viruses: type A of H1N1 seasonal 2009, H1N1 pandemic 2009, H3N2 seasonal 2012 and type B of seasonal 2012. It was indicated that Veritor System had the highest limit of detection from the kits.</p>","PeriodicalId":74740,"journal":{"name":"Rinsho Biseibutsu Jinsoku Shindan Kenkyukai shi = JARMAM : Journal of the Association for Rapid Method and Automation in Microbiology","volume":"24 2","pages":"69-70"},"PeriodicalIF":0.0000,"publicationDate":"2014-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Rinsho Biseibutsu Jinsoku Shindan Kenkyukai shi = JARMAM : Journal of the Association for Rapid Method and Automation in Microbiology","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0
Abstract
Immunochromatography viral antigen-detection kits have become popular in clinical settings in Japan. Influenza virus detection kit is one of them. It is sometimes used in early phase of the disease, combined with the early treatment with anti-influenza drugs. Most of them are invented to visually read the test line on their kits. However, we should be careful about their reliability of them. Sometimes human errors occur at the visual tests, and they have different sensitivities among the kits from different companies. In this report, we evaluated the sensitivity of BD Veritor System Flu with its reader by comparing with conventional visual tests. A total of 84 people including laboratory technologists were asked to visually read test line and their answers were compared with results of BD Veritor System Reader. This study showed that the lower the concentration of standard sample was applied, the greater the error ratio of visual test became, indicating the stable sensitivity of Veritor System. Moreover, the sensitivity was compared with three other major products approved in Japan, using four influenza viruses: type A of H1N1 seasonal 2009, H1N1 pandemic 2009, H3N2 seasonal 2012 and type B of seasonal 2012. It was indicated that Veritor System had the highest limit of detection from the kits.
免疫层析病毒抗原检测试剂盒已成为流行在日本的临床设置。流感病毒检测试剂盒就是其中之一。它有时在疾病的早期阶段使用,与抗流感药物的早期治疗相结合。它们中的大多数都是为了直观地读取试剂盒上的测试线而发明的。然而,我们应该小心他们的可靠性。有时在视觉测试中会出现人为错误,并且不同公司的试剂盒具有不同的灵敏度。在本报告中,我们通过比较传统的视觉测试来评估BD Veritor System Flu及其阅读器的敏感性。包括实验室技术人员在内的84人被要求视觉阅读测试线,并将他们的答案与BD Veritor系统阅读器的结果进行比较。本研究表明,标准样品的浓度越低,视觉检测的误差率越大,说明Veritor系统灵敏度稳定。此外,与日本批准的其他三种主要产品进行敏感性比较,使用四种流感病毒:2009年H1N1季节性A型,2009年H1N1大流行,2012年H3N2季节性和2012年季节性B型。结果表明,Veritor系统的检出限最高。