Experience in using thermal disinfection to remove viable bacteria and endotoxins in centraly distributed reverse osmosis water.

Ahmed Alayoud, Mohamed Amine Hamzi, Abdelaziz Razkaoui, Mohammed Benyahia
{"title":"Experience in using thermal disinfection to remove viable bacteria and endotoxins in centraly distributed reverse osmosis water.","authors":"Ahmed Alayoud,&nbsp;Mohamed Amine Hamzi,&nbsp;Abdelaziz Razkaoui,&nbsp;Mohammed Benyahia","doi":"","DOIUrl":null,"url":null,"abstract":"<p><strong>Introduction: </strong>The water used for dilution of hemodialysis concentrates has to meet official quality recommendations regarding microbiology and chemical parameters. To avoid chemical use and to simplify treatments, hot water has been used to control microbial contamination of water distribution systems. In this study we evaluated the efficacy of heat disinfection in maintaining the quality of dialysis water generated by reverse osmosis (RO).</p><p><strong>Methods: </strong>During the first part of the study, we consecutively used (1) continuous water circulation, (2) daily heat disinfection and (3) a combination of daily heat disinfection and weekly chemical disinfection while checking bacterial count and endotoxin level every 4-5 weeks. During the second part of the study, we continued using daily heat disinfection while checking bacterial count and endotoxin level on weekly basis.</p><p><strong>Results: </strong>The endotoxin levels at all sampling points of the water treatment system were lower than 0.005/ ml throughout the study. The application of heat disinfection alone reduced bacterial levels but an escape phenomenon occurred. After an interval of 21 days, an exponential increase of bacterial count was noted and cultures from the RO unit revealed growth of Pseudomonas fluorescence. The addition of chemical disinfection was successful in eliminating micro-organisms. Throughout this study, micro-organisms and endotoxins were not detectable in dialysate fluid and substitution fluid in dialysis monitors.</p><p><strong>Conclusion: </strong>The isolation of a thermo-sensitive organism from the RO unit after a period of relying on thermal disinfection suggests the existence of dead space in the RO unit that is not adequately exposed to heat but is accessible to chemical disinfection. .</p>","PeriodicalId":89079,"journal":{"name":"Arab journal of nephrology and transplantation","volume":"7 1","pages":"27-31"},"PeriodicalIF":0.0000,"publicationDate":"2014-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Arab journal of nephrology and transplantation","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Introduction: The water used for dilution of hemodialysis concentrates has to meet official quality recommendations regarding microbiology and chemical parameters. To avoid chemical use and to simplify treatments, hot water has been used to control microbial contamination of water distribution systems. In this study we evaluated the efficacy of heat disinfection in maintaining the quality of dialysis water generated by reverse osmosis (RO).

Methods: During the first part of the study, we consecutively used (1) continuous water circulation, (2) daily heat disinfection and (3) a combination of daily heat disinfection and weekly chemical disinfection while checking bacterial count and endotoxin level every 4-5 weeks. During the second part of the study, we continued using daily heat disinfection while checking bacterial count and endotoxin level on weekly basis.

Results: The endotoxin levels at all sampling points of the water treatment system were lower than 0.005/ ml throughout the study. The application of heat disinfection alone reduced bacterial levels but an escape phenomenon occurred. After an interval of 21 days, an exponential increase of bacterial count was noted and cultures from the RO unit revealed growth of Pseudomonas fluorescence. The addition of chemical disinfection was successful in eliminating micro-organisms. Throughout this study, micro-organisms and endotoxins were not detectable in dialysate fluid and substitution fluid in dialysis monitors.

Conclusion: The isolation of a thermo-sensitive organism from the RO unit after a period of relying on thermal disinfection suggests the existence of dead space in the RO unit that is not adequately exposed to heat but is accessible to chemical disinfection. .

热消毒去除集中分布的反渗透水中活菌和内毒素的经验。
介绍:用于稀释血液透析浓缩液的水必须符合官方关于微生物和化学参数的质量建议。为了避免化学品的使用和简化处理,热水被用来控制水分配系统的微生物污染。在这项研究中,我们评估了热消毒对维持反渗透(RO)产生的透析水质量的功效。方法:在第一部分研究中,我们连续采用(1)连续水循环,(2)每日热消毒,(3)每日热消毒与每周化学消毒相结合,每4-5周检查细菌计数和内毒素水平。在研究的第二部分,我们继续使用每日热消毒,每周检查细菌计数和内毒素水平。结果:在整个研究过程中,水处理系统各采样点的内毒素水平均低于0.005/ ml。单独热消毒的应用降低了细菌水平,但发生了逃逸现象。间隔21天后,观察到细菌数量呈指数增长,RO单元培养显示荧光假单胞菌生长。化学消毒的加入成功地消除了微生物。在整个研究中,透析监测器中透析液和替代液中未检测到微生物和内毒素。结论:在依靠热消毒一段时间后,从反渗透装置中分离出一种热敏生物,表明反渗透装置中存在未充分暴露于热但可进行化学消毒的死区。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书