Ronald R Krueger, Harvey Uy, Jared McDonald, Keith Edwards
{"title":"Ultrashort-pulse lasers treating the crystalline lens: will they cause vision-threatening cataract? (An American Ophthalmological Society thesis).","authors":"Ronald R Krueger, Harvey Uy, Jared McDonald, Keith Edwards","doi":"","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>To demonstrate that ultrashort-pulse laser treatment in the crystalline lens does not form a focal, progressive, or vision-threatening cataract.</p><p><strong>Methods: </strong>An Nd:vanadate picosecond laser (10 ps) with prototype delivery system was used. Primates: 11 rhesus monkey eyes were prospectively treated at the University of Wisconsin (energy 25-45 μJ/pulse and 2.0-11.3M pulses per lens). Analysis of lens clarity and fundus imaging was assessed postoperatively for up to 4½ years (5 eyes). Humans: 80 presbyopic patients were prospectively treated in one eye at the Asian Eye Institute in the Philippines (energy 10 μJ/pulse and 0.45-1.45M pulses per lens). Analysis of lens clarity, best-corrected visual acuity, and subjective symptoms was performed at 1 month, prior to elective lens extraction.</p><p><strong>Results: </strong>Bubbles were immediately seen, with resolution within the first 24 to 48 hours. Afterwards, the laser pattern could be seen with faint, noncoalescing, pinpoint micro-opacities in both primate and human eyes. In primates, long-term follow-up at 4½ years showed no focal or progressive cataract, except in 2 eyes with preexisting cataract. In humans, <25% of patients with central sparing (0.75 and 1.0 mm radius) lost 2 or more lines of best spectacle-corrected visual acuity at 1 month, and >70% reported acceptable or better distance vision and no or mild symptoms. Meanwhile, >70% without sparing (0 and 0.5 mm radius) lost 2 or more lines, and most reported poor or severe vision and symptoms.</p><p><strong>Conclusions: </strong>Focal, progressive, and vision-threatening cataracts can be avoided by lowering the laser energy, avoiding prior cataract, and sparing the center of the lens.</p>","PeriodicalId":23166,"journal":{"name":"Transactions of the American Ophthalmological Society","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"2012-12-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3671365/pdf/1545-6110_v110_p130.pdf","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Transactions of the American Ophthalmological Society","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0
Abstract
Purpose: To demonstrate that ultrashort-pulse laser treatment in the crystalline lens does not form a focal, progressive, or vision-threatening cataract.
Methods: An Nd:vanadate picosecond laser (10 ps) with prototype delivery system was used. Primates: 11 rhesus monkey eyes were prospectively treated at the University of Wisconsin (energy 25-45 μJ/pulse and 2.0-11.3M pulses per lens). Analysis of lens clarity and fundus imaging was assessed postoperatively for up to 4½ years (5 eyes). Humans: 80 presbyopic patients were prospectively treated in one eye at the Asian Eye Institute in the Philippines (energy 10 μJ/pulse and 0.45-1.45M pulses per lens). Analysis of lens clarity, best-corrected visual acuity, and subjective symptoms was performed at 1 month, prior to elective lens extraction.
Results: Bubbles were immediately seen, with resolution within the first 24 to 48 hours. Afterwards, the laser pattern could be seen with faint, noncoalescing, pinpoint micro-opacities in both primate and human eyes. In primates, long-term follow-up at 4½ years showed no focal or progressive cataract, except in 2 eyes with preexisting cataract. In humans, <25% of patients with central sparing (0.75 and 1.0 mm radius) lost 2 or more lines of best spectacle-corrected visual acuity at 1 month, and >70% reported acceptable or better distance vision and no or mild symptoms. Meanwhile, >70% without sparing (0 and 0.5 mm radius) lost 2 or more lines, and most reported poor or severe vision and symptoms.
Conclusions: Focal, progressive, and vision-threatening cataracts can be avoided by lowering the laser energy, avoiding prior cataract, and sparing the center of the lens.