Quality assurance of TomoDirect treatment plans using I'mRT MatriXX.

Cw Kong, Sk Yu, Ky Cheung, H Geng, Yw Ho, Ww Lam, Wk Wong
{"title":"Quality assurance of TomoDirect treatment plans using I'mRT MatriXX.","authors":"Cw Kong,&nbsp;Sk Yu,&nbsp;Ky Cheung,&nbsp;H Geng,&nbsp;Yw Ho,&nbsp;Ww Lam,&nbsp;Wk Wong","doi":"10.2349/biij.8.2.e14","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>To evaluate the performance of 2D-array I'mRT MatriXX for dose verification of TomoDirect treatment plans.</p><p><strong>Methods: </strong>In this study, a 2D-array ion chamber device - the I'mRT MatriXX and Multicube Phantom from IBA - was used for dose verification of different TomoDirect plans. Pre-treatment megavoltage computed tomography (MVCT) was performed on the phantom setup for position correction. After the irradiation of treatment plans on the I'mRT MatriXX and Multicube Phantom, the measured doses of coronal planes were compared with those from the planning calculations for verification. The results were evaluated by comparing the absolute dose difference in the high dose region as well as the gamma analysis of the 2D-dose distributions on the coronal plane. The comparison was then repeated with the measured dose corrected for angular dependence of the MatriXX.</p><p><strong>Results: </strong>When angular dependence is taken into account, the passing rate of gamma analysis is over 90% for all measurements using the MatriXX. If there is no angular dependence correction, the passing rate of gamma analysis worsens for treatment plans with dose contribution from the rear. The passing rate can be as low as 53.55% in extreme cases, i.e. where all doses in the treatment plan are delivered from the rear.</p><p><strong>Conclusion: </strong>It is important to correct the measured dose for angular dependence when verifying TomoDirect treatment plans using the MatriXX. If left uncorrected, a large dose discrepancy may be introduced to the verification results.</p>","PeriodicalId":89331,"journal":{"name":"Biomedical imaging and intervention journal","volume":"8 2","pages":"e14"},"PeriodicalIF":0.0000,"publicationDate":"2012-04-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.2349/biij.8.2.e14","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Biomedical imaging and intervention journal","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.2349/biij.8.2.e14","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose: To evaluate the performance of 2D-array I'mRT MatriXX for dose verification of TomoDirect treatment plans.

Methods: In this study, a 2D-array ion chamber device - the I'mRT MatriXX and Multicube Phantom from IBA - was used for dose verification of different TomoDirect plans. Pre-treatment megavoltage computed tomography (MVCT) was performed on the phantom setup for position correction. After the irradiation of treatment plans on the I'mRT MatriXX and Multicube Phantom, the measured doses of coronal planes were compared with those from the planning calculations for verification. The results were evaluated by comparing the absolute dose difference in the high dose region as well as the gamma analysis of the 2D-dose distributions on the coronal plane. The comparison was then repeated with the measured dose corrected for angular dependence of the MatriXX.

Results: When angular dependence is taken into account, the passing rate of gamma analysis is over 90% for all measurements using the MatriXX. If there is no angular dependence correction, the passing rate of gamma analysis worsens for treatment plans with dose contribution from the rear. The passing rate can be as low as 53.55% in extreme cases, i.e. where all doses in the treatment plan are delivered from the rear.

Conclusion: It is important to correct the measured dose for angular dependence when verifying TomoDirect treatment plans using the MatriXX. If left uncorrected, a large dose discrepancy may be introduced to the verification results.

Abstract Image

Abstract Image

Abstract Image

使用I'mRT MatriXX的TomoDirect治疗方案的质量保证。
目的:评价二维阵列I'mRT MatriXX用于TomoDirect治疗方案剂量验证的性能。方法:在本研究中,使用IBA的I'mRT MatriXX和multiccube Phantom作为2d阵列离子室装置,对不同的TomoDirect计划进行剂量验证。预处理巨压计算机断层扫描(MVCT)对幻体装置进行位置校正。治疗方案在I'mRT MatriXX和multiccube Phantom上照射后,将冠状面测量剂量与计划计算剂量进行比较验证。通过比较高剂量区的绝对剂量差和冠状面二维剂量分布的伽玛分析来评价结果。然后重复比较,测量剂量校正了MatriXX的角度依赖性。结果:当考虑角度依赖性时,使用MatriXX的所有测量的伽马分析通过率超过90%。如果没有角度依赖性校正,对于剂量来自后方的治疗方案,伽马分析的通过率会下降。在极端情况下,即治疗方案中的所有剂量均从后方给药,通过率可低至53.55%。结论:在使用MatriXX验证TomoDirect治疗方案时,重要的是校正测量剂量的角度依赖性。如果不加以纠正,可能会在验证结果中引入较大的剂量差异。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信