Evaluation of imaging performance of major image guidance systems.

Mf Chan, J Yang, Y Song, C Burman, P Chan, S Li
{"title":"Evaluation of imaging performance of major image guidance systems.","authors":"Mf Chan,&nbsp;J Yang,&nbsp;Y Song,&nbsp;C Burman,&nbsp;P Chan,&nbsp;S Li","doi":"10.2349/biij.7.2.e11","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>The imaging characteristics of two popular kV cone-beam CT (CBCT) and two MVCT systems utilised in image-guided radiation therapy (IGRT) were evaluated.</p><p><strong>Materials and methods: </strong>The study was performed on Varian Clinac iX, Elekta Synergy S, Siemens Oncor, and Tomotherapy. A CT phantom (Catphan-504, Phantom Laboratory, Salem, NY) was scanned for measurements of image quality including image noise, uniformity, density accuracy, spatial resolution, contrast linearity, and contrast resolution. The measurement results were analysed using in-house image analysis software. Reproducibility, position correction, and geometric accuracy were also evaluated with markers in a smaller alignment phantom. The performance evaluation compared volumetric image properties from these four systems with those from a conventional diagnostic CT (CCT).</p><p><strong>Results: </strong>It was shown that the linearity of the two kV CBCT was fairly consistent with CCT. The Elekta CBCT with half-circle 27-cm FOV had higher CT numbers than the other three systems. The image noises of the Elekta kV CBCT, Siemens MV CBCT, and Tomotherapy fan-beam CT (FBCT) are about 2-4 times higher than that of the Varian CBCT. The spatial resolutions of two kV CBCTs and two MV CBCTs were 8-11 lp/cm and 3-5 lp/cm, respectively.</p><p><strong>Conclusion: </strong>Elekta CBCT provided a faster image reconstruction and low dose per scan for half-circle scanning. Varian CBCT had relatively lower image noise. Tomotherapy FBCT had the best uniformity.</p>","PeriodicalId":89331,"journal":{"name":"Biomedical imaging and intervention journal","volume":"7 2","pages":"e11"},"PeriodicalIF":0.0000,"publicationDate":"2011-04-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.2349/biij.7.2.e11","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Biomedical imaging and intervention journal","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.2349/biij.7.2.e11","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose: The imaging characteristics of two popular kV cone-beam CT (CBCT) and two MVCT systems utilised in image-guided radiation therapy (IGRT) were evaluated.

Materials and methods: The study was performed on Varian Clinac iX, Elekta Synergy S, Siemens Oncor, and Tomotherapy. A CT phantom (Catphan-504, Phantom Laboratory, Salem, NY) was scanned for measurements of image quality including image noise, uniformity, density accuracy, spatial resolution, contrast linearity, and contrast resolution. The measurement results were analysed using in-house image analysis software. Reproducibility, position correction, and geometric accuracy were also evaluated with markers in a smaller alignment phantom. The performance evaluation compared volumetric image properties from these four systems with those from a conventional diagnostic CT (CCT).

Results: It was shown that the linearity of the two kV CBCT was fairly consistent with CCT. The Elekta CBCT with half-circle 27-cm FOV had higher CT numbers than the other three systems. The image noises of the Elekta kV CBCT, Siemens MV CBCT, and Tomotherapy fan-beam CT (FBCT) are about 2-4 times higher than that of the Varian CBCT. The spatial resolutions of two kV CBCTs and two MV CBCTs were 8-11 lp/cm and 3-5 lp/cm, respectively.

Conclusion: Elekta CBCT provided a faster image reconstruction and low dose per scan for half-circle scanning. Varian CBCT had relatively lower image noise. Tomotherapy FBCT had the best uniformity.

Abstract Image

Abstract Image

Abstract Image

主要图像制导系统成像性能评价。
目的:评价两种常用的kV锥束CT (CBCT)和两种MVCT系统在图像引导放射治疗(IGRT)中的成像特性。材料和方法:研究采用Varian Clinac iX、Elekta Synergy S、Siemens Oncor和Tomotherapy。扫描CT模型(Catphan-504, phantom Laboratory, Salem, NY),测量图像质量,包括图像噪声、均匀性、密度精度、空间分辨率、对比度线性度和对比度分辨率。测量结果使用内部图像分析软件进行分析。再现性,位置校正,几何精度也评价标记在一个较小的对准幻影。性能评估将这四种系统的体积图像特性与传统诊断CT (CCT)的体积图像特性进行比较。结果:两kV CBCT的线性度与CCT相当一致。半圆27cm FOV的Elekta CBCT的CT数高于其他三种系统。Elekta kV CBCT、Siemens MV CBCT和Tomotherapy扇束CT (FBCT)的图像噪声约为Varian CBCT的2-4倍。2 kV cbct和2 MV cbct的空间分辨率分别为8 ~ 11 lp/cm和3 ~ 5 lp/cm。结论:Elekta CBCT具有图像重建速度快、单次扫描剂量低的优点。瓦里安CBCT具有较低的图像噪声。断层治疗FBCT均匀性最好。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信