Monitoring Changes in Fluid Status With a Wireless Multisensor Monitor: Results From the Fluid Removal During Adherent Renal Monitoring (FARM) Study

Inder S. Anand MD, Alexander D. Doan MD, King W. Ma MD, Julie A. Toth RN, Karolyn J. Geyen RN, Sara Otterness APRN-BC, CNN, Niranjan Chakravarthy PhD, Rodolphe P. Katra PhD, Imad Libbus PhD
{"title":"Monitoring Changes in Fluid Status With a Wireless Multisensor Monitor: Results From the Fluid Removal During Adherent Renal Monitoring (FARM) Study","authors":"Inder S. Anand MD,&nbsp;Alexander D. Doan MD,&nbsp;King W. Ma MD,&nbsp;Julie A. Toth RN,&nbsp;Karolyn J. Geyen RN,&nbsp;Sara Otterness APRN-BC, CNN,&nbsp;Niranjan Chakravarthy PhD,&nbsp;Rodolphe P. Katra PhD,&nbsp;Imad Libbus PhD","doi":"10.1111/j.1751-7133.2011.00271.x","DOIUrl":null,"url":null,"abstract":"<p>Body fluid assessment is important for managing chronic kidney disease (CKD) and heart failure (HF). However, accurate detection of fluid retention remains elusive. The Fluid Removal During Adherent Renal Monitoring (FARM) study is a prospective, nonrandomized trial examining the performance of a wireless, noninvasive, multisensor fluid monitoring system, applied to the chest, to determine its performance and reliability during hemodialysis. Patients undergoing regular hemodialysis (n=25) were monitored continuously for 2 consecutive dialysis sessions and the interdialysis period. Physiologic variables, including tissue impedance, were recorded. The volume of fluid removed and weight change during dialysis were measured. An average of 3.4±1.2 L of fluid was removed during dialysis, which was associated with an increase in bioimpedance of 11.3±7.2 Ω. Change in bioimpedance was highly correlated with the amount of fluid removed but less so with weight loss. Normalized bioimpedance change (21.0%±12.1% increase from baseline, P&lt;001) was larger than the normalized weight change (3.6%±1.1%, <i>P</i>&lt;.01), suggesting a higher sensitivity and dynamic range than weight change for detecting fluid removal. The fluid monitoring system accurately tracked fluid and weight loss in patients during hemodialysis, supporting its use as a tool for the management of patient fluid status in disease states. <i>Congest Heart Fail.</i> 2012;18:32–36. ©2011 Wiley Periodicals, Inc.</p>","PeriodicalId":83668,"journal":{"name":"Congestive heart failure (Greenwich, Conn.)","volume":"18 1","pages":"32-36"},"PeriodicalIF":0.0000,"publicationDate":"2011-11-21","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.1111/j.1751-7133.2011.00271.x","citationCount":"15","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Congestive heart failure (Greenwich, Conn.)","FirstCategoryId":"1085","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1111/j.1751-7133.2011.00271.x","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 15

Abstract

Body fluid assessment is important for managing chronic kidney disease (CKD) and heart failure (HF). However, accurate detection of fluid retention remains elusive. The Fluid Removal During Adherent Renal Monitoring (FARM) study is a prospective, nonrandomized trial examining the performance of a wireless, noninvasive, multisensor fluid monitoring system, applied to the chest, to determine its performance and reliability during hemodialysis. Patients undergoing regular hemodialysis (n=25) were monitored continuously for 2 consecutive dialysis sessions and the interdialysis period. Physiologic variables, including tissue impedance, were recorded. The volume of fluid removed and weight change during dialysis were measured. An average of 3.4±1.2 L of fluid was removed during dialysis, which was associated with an increase in bioimpedance of 11.3±7.2 Ω. Change in bioimpedance was highly correlated with the amount of fluid removed but less so with weight loss. Normalized bioimpedance change (21.0%±12.1% increase from baseline, P<001) was larger than the normalized weight change (3.6%±1.1%, P<.01), suggesting a higher sensitivity and dynamic range than weight change for detecting fluid removal. The fluid monitoring system accurately tracked fluid and weight loss in patients during hemodialysis, supporting its use as a tool for the management of patient fluid status in disease states. Congest Heart Fail. 2012;18:32–36. ©2011 Wiley Periodicals, Inc.

Abstract Image

用无线多传感器监测器监测液体状态的变化:从附着肾监测(FARM)研究中液体去除的结果
体液评估对于治疗慢性肾脏疾病(CKD)和心力衰竭(HF)非常重要。然而,准确检测液体潴留仍然是难以捉摸的。FARM研究是一项前瞻性、非随机试验,研究应用于胸部的无线、无创、多传感器液体监测系统的性能,以确定其在血液透析期间的性能和可靠性。定期进行血液透析的患者(n=25)连续监测2个连续透析期和透析间期。记录生理变量,包括组织阻抗。测定透析过程中排液量和体重变化。透析期间平均排出3.4±1.2 L液体,与生物阻抗增加11.3±7.2 Ω相关。生物阻抗的变化与排出的液体量高度相关,但与体重减轻的相关性较小。归一化的生物阻抗变化(比基线增加21.0%±12.1%,P<001)大于归一化的体重变化(3.6%±1.1%,P< 0.01),表明在检测液体清除方面,比体重变化具有更高的灵敏度和动态范围。液体监测系统准确地跟踪血液透析期间患者的液体和体重减少,支持其作为疾病状态下患者液体状态管理工具的使用。心力衰竭。2012;18:32-36。©2011 Wiley期刊公司
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书