Reliability of frames of reference used for tibial component rotation in total knee arthroplasty.

Q Medicine
Computer Aided Surgery Pub Date : 2011-01-01 Epub Date: 2011-02-04 DOI:10.3109/10929088.2011.552252
Stephen R Page, Angela H Deakin, Anthony P Payne, Frederic Picard
{"title":"Reliability of frames of reference used for tibial component rotation in total knee arthroplasty.","authors":"Stephen R Page,&nbsp;Angela H Deakin,&nbsp;Anthony P Payne,&nbsp;Frederic Picard","doi":"10.3109/10929088.2011.552252","DOIUrl":null,"url":null,"abstract":"<p><p>This study evaluated seven different frames of reference used for tibial component rotation in total knee arthroplasty (TKA) to determine which ones showed good reliability between bone specimens. An optoelectronic system based around a computer-assisted surgical navigation system was used to measure and locate 34 individual anatomical landmarks on 40 tibias. Each particular frame of reference was reconstructed from a group of data points taken from the surface of each bone. The transverse axis was used as the baseline to which the other axes were compared, and the differences in angular rotation between the other six reference frames and the transverse axis were calculated. There was high variability in the tibial rotational alignment associated with all frames of reference. Of the references widely used in current TKA procedures, the tibial tuberosity axis and the anterior condylar axis had lower standard deviations (6.1° and 7.3°, respectively) than the transmalleolar axis and the posterior condylar axis (9.3° for both). In conclusion, we found high variability in the frames of reference used for tibial rotation alignment. However, the anterior condylar axis and transverse axis may warrant further tests with the use of navigation. Combining different frames of reference such as the tibial tuberosity axis, anterior condylar axis and transverse axis may reduce the range of errors found in all of these measurements.</p>","PeriodicalId":50644,"journal":{"name":"Computer Aided Surgery","volume":"16 2","pages":"86-92"},"PeriodicalIF":0.0000,"publicationDate":"2011-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.3109/10929088.2011.552252","citationCount":"19","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Computer Aided Surgery","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.3109/10929088.2011.552252","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2011/2/4 0:00:00","PubModel":"Epub","JCR":"Q","JCRName":"Medicine","Score":null,"Total":0}
引用次数: 19

Abstract

This study evaluated seven different frames of reference used for tibial component rotation in total knee arthroplasty (TKA) to determine which ones showed good reliability between bone specimens. An optoelectronic system based around a computer-assisted surgical navigation system was used to measure and locate 34 individual anatomical landmarks on 40 tibias. Each particular frame of reference was reconstructed from a group of data points taken from the surface of each bone. The transverse axis was used as the baseline to which the other axes were compared, and the differences in angular rotation between the other six reference frames and the transverse axis were calculated. There was high variability in the tibial rotational alignment associated with all frames of reference. Of the references widely used in current TKA procedures, the tibial tuberosity axis and the anterior condylar axis had lower standard deviations (6.1° and 7.3°, respectively) than the transmalleolar axis and the posterior condylar axis (9.3° for both). In conclusion, we found high variability in the frames of reference used for tibial rotation alignment. However, the anterior condylar axis and transverse axis may warrant further tests with the use of navigation. Combining different frames of reference such as the tibial tuberosity axis, anterior condylar axis and transverse axis may reduce the range of errors found in all of these measurements.

全膝关节置换术中胫骨构件旋转参考框架的可靠性。
本研究评估了全膝关节置换术(TKA)中用于胫骨构件旋转的7种不同参考框架,以确定在骨标本之间哪一种具有良好的可靠性。基于计算机辅助手术导航系统的光电系统用于测量和定位40条胫骨上的34个单独的解剖标志。每个特定的参照系都是从每根骨头的表面提取的一组数据点重建的。以横轴为基准,对其他轴进行比较,计算其他6个参照系与横轴的转角差值。在所有参照系中,胫骨旋转对中存在高度可变性。在目前TKA手术中广泛使用的参考文献中,胫骨结节轴和前髁轴的标准差(分别为6.1°和7.3°)低于踝外轴和后髁轴(均为9.3°)。总之,我们发现用于胫骨旋转对齐的参考框架具有高度可变性。然而,前髁轴和横轴可能需要使用导航进行进一步的检查。结合不同的参考框架,如胫骨结节轴、前髁轴和横轴,可以减少所有这些测量中的误差范围。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Computer Aided Surgery
Computer Aided Surgery 医学-外科
CiteScore
0.75
自引率
0.00%
发文量
0
审稿时长
>12 weeks
期刊介绍: The scope of Computer Aided Surgery encompasses all fields within surgery, as well as biomedical imaging and instrumentation, and digital technology employed as an adjunct to imaging in diagnosis, therapeutics, and surgery. Topics featured include frameless as well as conventional stereotaxic procedures, surgery guided by ultrasound, image guided focal irradiation, robotic surgery, and other therapeutic interventions that are performed with the use of digital imaging technology.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信