Otoacoustic emissions growth rate threshold: distortion product in neonates.

Patricia Pinheiro de Almeida, Seisse Gabriela Gandolfi Sanches, Renata Mota Mamede Carvallo
{"title":"Otoacoustic emissions growth rate threshold: distortion product in neonates.","authors":"Patricia Pinheiro de Almeida,&nbsp;Seisse Gabriela Gandolfi Sanches,&nbsp;Renata Mota Mamede Carvallo","doi":"10.1590/s0104-56872010000400008","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Distortion product otoacoustic emission measures (DPOAE) verifies, among other characteristics, the emergence and growth rate of the DPOAE response according to sound stimulus intensity (growth curve).</p><p><strong>Aim: </strong>To estimate the DPOAE threshold in neonates according to the DPOAE growth curve at 2 kHz and 4 kHz, with stimulus presentation ranging between 35 and 70 dB SPL.</p><p><strong>Method: </strong>Participants were 51 neonates, with 24 to 84 hours of life and with no indication of risk for hearing impairment. The DPOAE were registered in the growth curve function at 2 kHz and at 4 kHz. Neonates were evaluated during the period of hospital stay, after birth. Three possible thresholds were considered (LIM 1, LIM 2 and LIM 3) based on the presence of response at 3 dB SPL in the signal to noise ratio.</p><p><strong>Results: </strong>The average thresholds intensities ranged from 47.55 to 49.85 dB at 2 kHz and from 55.52 to 59.94 dB at 4 kHz. The mean amplitude response ranged from 6.67 to 8.27 dB at 2 kHz and from 6.99 to 11.35 dB at 4 kHz. There was significant difference between the three thresholds for the two studied frequencies.</p><p><strong>Conclusion: </strong>The procedure was feasible for the neonatal population revealing mean thresholds of up to 60 dB for both frequencies. Even though participants presented elevated thresholds, robust amplitude responses were observed.</p>","PeriodicalId":74581,"journal":{"name":"Pro-fono : revista de atualizacao cientifica","volume":"22 4","pages":"409-14"},"PeriodicalIF":0.0000,"publicationDate":"2010-10-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.1590/s0104-56872010000400008","citationCount":"2","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Pro-fono : revista de atualizacao cientifica","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1590/s0104-56872010000400008","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 2

Abstract

Background: Distortion product otoacoustic emission measures (DPOAE) verifies, among other characteristics, the emergence and growth rate of the DPOAE response according to sound stimulus intensity (growth curve).

Aim: To estimate the DPOAE threshold in neonates according to the DPOAE growth curve at 2 kHz and 4 kHz, with stimulus presentation ranging between 35 and 70 dB SPL.

Method: Participants were 51 neonates, with 24 to 84 hours of life and with no indication of risk for hearing impairment. The DPOAE were registered in the growth curve function at 2 kHz and at 4 kHz. Neonates were evaluated during the period of hospital stay, after birth. Three possible thresholds were considered (LIM 1, LIM 2 and LIM 3) based on the presence of response at 3 dB SPL in the signal to noise ratio.

Results: The average thresholds intensities ranged from 47.55 to 49.85 dB at 2 kHz and from 55.52 to 59.94 dB at 4 kHz. The mean amplitude response ranged from 6.67 to 8.27 dB at 2 kHz and from 6.99 to 11.35 dB at 4 kHz. There was significant difference between the three thresholds for the two studied frequencies.

Conclusion: The procedure was feasible for the neonatal population revealing mean thresholds of up to 60 dB for both frequencies. Even though participants presented elevated thresholds, robust amplitude responses were observed.

耳声发射生长速率阈值:新生儿畸变产物。
背景:畸变产物耳声发射测量(DPOAE)根据声刺激强度(生长曲线)验证了DPOAE响应的出现和增长速度等特征。目的:根据2 kHz和4 kHz时的DPOAE生长曲线,在35 ~ 70 dB SPL的刺激表现下,估计新生儿的DPOAE阈值。方法:参与者为51名新生儿,寿命24至84小时,无听力障碍风险迹象。在2 kHz和4 kHz的生长曲线函数中记录了DPOAE。在新生儿出生后住院期间对其进行评估。基于信噪比中存在3db SPL的响应,考虑了三个可能的阈值(LIM 1, LIM 2和LIM 3)。结果:2 kHz时平均阈值强度为47.55 ~ 49.85 dB, 4 kHz时为55.52 ~ 59.94 dB。平均振幅响应在2 kHz时为6.67 ~ 8.27 dB,在4 kHz时为6.99 ~ 11.35 dB。两个研究频率的三个阈值之间存在显著差异。结论:该方法对新生儿群体是可行的,两个频率的平均阈值高达60 dB。尽管参与者表现出较高的阈值,但观察到稳健的幅度反应。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信