Cone beam computed tomography radiation dose and image quality assessments.

Swedish dental journal. Supplement Pub Date : 2010-01-01
Sara Lofthag-Hansen
{"title":"Cone beam computed tomography radiation dose and image quality assessments.","authors":"Sara Lofthag-Hansen","doi":"","DOIUrl":null,"url":null,"abstract":"<p><p>Diagnostic radiology has undergone profound changes in the last 30 years. New technologies are available to the dental field, cone beam computed tomography (CBCT) as one of the most important. CBCT is a catch-all term for a technology comprising a variety of machines differing in many respects: patient positioning, volume size (FOV), radiation quality, image capturing and reconstruction, image resolution and radiation dose. When new technology is introduced one must make sure that diagnostic accuracy is better or at least as good as the one it can be expected to replace. The CBCT brand tested was two versions of Accuitomo (Morita, Japan): 3D Accuitomo with an image intensifier as detector, FOV 3 cm x 4 cm and 3D Accuitomo FPD with a flat panel detector, FOVs 4 cm x 4 cm and 6 cm x 6 cm. The 3D Accuitomo was compared with intra-oral radiography for endodontic diagnosis in 35 patients with 46 teeth analyzed, of which 41 were endodontically treated. Three observers assessed the images by consensus. The result showed that CBCT imaging was superior with a higher number of teeth diagnosed with periapical lesions (42 vs 32 teeth). When evaluating 3D Accuitomo examinations in the posterior mandible in 30 patients, visibility of marginal bone crest and mandibular canal, important anatomic structures for implant planning, was high with good observer agreement among seven observers. Radiographic techniques have to be evaluated concerning radiation dose, which requires well-defined and easy-to-use methods. Two methods: CT dose index (CTDI), prevailing method for CT units, and dose-area product (DAP) were evaluated for calculating effective dose (E) for both units. An asymmetric dose distribution was revealed when a clinical situation was simulated. Hence, the CTDI method was not applicable for these units with small FOVs. Based on DAP values from 90 patient examinations effective dose was estimated for three diagnostic tasks: implant planning in posterior mandible and examinations of impacted lower third molars and retained upper cuspids. It varied between 11-77 microSv. Radiation dose should be evaluated together with image quality. Images of a skull phantom were obtained with both units varying tube voltage, tube current, degree of rotation and FOVs. Seven observers assessed subjective image quality using a six-point rating scale for two diagnostic tasks: periapical diagnosis and implant planning in the posterior part of the jaws. Intra-observer agreement was good and inter-observer agreement moderate. Periapical diagnosis was found to, regardless of jaw, require higher exposure parameters compared to implant planning. Implant planning in the lower jaw required higher exposure parameters compared to upper jaw. Substantial dose reduction could be made without loss of diagnostic information by using a rotation of 180 degrees, in particular implant planning in upper jaw. CBCT with small FOVs was found to be well-suited for periapical diagnosis and implant planning. The CTDI method is not applicable estimating effective dose for these units. Based on DAP values effective dose varied between 11-77 microSv (ICRP 60, 1991) in a retrospectively selected patient material. Adaptation of exposure parameters to diagnostic task can give substantial dose reduction.</p>","PeriodicalId":76572,"journal":{"name":"Swedish dental journal. Supplement","volume":" 209","pages":"4-55"},"PeriodicalIF":0.0000,"publicationDate":"2010-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Swedish dental journal. Supplement","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Diagnostic radiology has undergone profound changes in the last 30 years. New technologies are available to the dental field, cone beam computed tomography (CBCT) as one of the most important. CBCT is a catch-all term for a technology comprising a variety of machines differing in many respects: patient positioning, volume size (FOV), radiation quality, image capturing and reconstruction, image resolution and radiation dose. When new technology is introduced one must make sure that diagnostic accuracy is better or at least as good as the one it can be expected to replace. The CBCT brand tested was two versions of Accuitomo (Morita, Japan): 3D Accuitomo with an image intensifier as detector, FOV 3 cm x 4 cm and 3D Accuitomo FPD with a flat panel detector, FOVs 4 cm x 4 cm and 6 cm x 6 cm. The 3D Accuitomo was compared with intra-oral radiography for endodontic diagnosis in 35 patients with 46 teeth analyzed, of which 41 were endodontically treated. Three observers assessed the images by consensus. The result showed that CBCT imaging was superior with a higher number of teeth diagnosed with periapical lesions (42 vs 32 teeth). When evaluating 3D Accuitomo examinations in the posterior mandible in 30 patients, visibility of marginal bone crest and mandibular canal, important anatomic structures for implant planning, was high with good observer agreement among seven observers. Radiographic techniques have to be evaluated concerning radiation dose, which requires well-defined and easy-to-use methods. Two methods: CT dose index (CTDI), prevailing method for CT units, and dose-area product (DAP) were evaluated for calculating effective dose (E) for both units. An asymmetric dose distribution was revealed when a clinical situation was simulated. Hence, the CTDI method was not applicable for these units with small FOVs. Based on DAP values from 90 patient examinations effective dose was estimated for three diagnostic tasks: implant planning in posterior mandible and examinations of impacted lower third molars and retained upper cuspids. It varied between 11-77 microSv. Radiation dose should be evaluated together with image quality. Images of a skull phantom were obtained with both units varying tube voltage, tube current, degree of rotation and FOVs. Seven observers assessed subjective image quality using a six-point rating scale for two diagnostic tasks: periapical diagnosis and implant planning in the posterior part of the jaws. Intra-observer agreement was good and inter-observer agreement moderate. Periapical diagnosis was found to, regardless of jaw, require higher exposure parameters compared to implant planning. Implant planning in the lower jaw required higher exposure parameters compared to upper jaw. Substantial dose reduction could be made without loss of diagnostic information by using a rotation of 180 degrees, in particular implant planning in upper jaw. CBCT with small FOVs was found to be well-suited for periapical diagnosis and implant planning. The CTDI method is not applicable estimating effective dose for these units. Based on DAP values effective dose varied between 11-77 microSv (ICRP 60, 1991) in a retrospectively selected patient material. Adaptation of exposure parameters to diagnostic task can give substantial dose reduction.

锥束计算机断层扫描辐射剂量和图像质量评估。
诊断放射学在过去的30年里发生了深刻的变化。新技术不断涌现到牙科领域,锥束计算机断层扫描(CBCT)是其中最重要的技术之一。CBCT是一个包罗万象的术语,指的是一种技术,包括在许多方面不同的各种机器:患者定位、体积大小(FOV)、辐射质量、图像捕获和重建、图像分辨率和辐射剂量。当引进新技术时,人们必须确保诊断的准确性更好,或者至少与它有望取代的技术一样好。CBCT品牌测试的是两个版本的Accuitomo(森田,日本):3D Accuitomo用图像增强器作为探测器,视场为3厘米× 4厘米;3D Accuitomo FPD用平板探测器,视场为4厘米× 4厘米和6厘米× 6厘米。对35例患者的46颗牙齿进行分析,其中41颗进行了牙髓治疗,并将3D Accuitomo与口腔内x线摄影进行了比较。三位观察员一致评价了这些图像。结果显示CBCT成像优势,诊断为根尖周病变的牙齿数量较多(42对32)。在对30例患者的后下颌进行3D Accuitomo检查时,边缘骨嵴和下颌管作为种植体规划的重要解剖结构的可见性很高,7名观察者的观察一致性很好。射线照相技术必须进行辐射剂量评价,这就需要明确的、易于使用的方法。采用CT剂量指数(CTDI)和剂量面积积(DAP)两种方法计算两种单元的有效剂量(E)。在模拟临床情况时,发现剂量分布不对称。因此,CTDI方法不适用于这些fov较小的单元。根据90例患者检查的DAP值,估计三个诊断任务的有效剂量:后下颌种植计划和检查阻生的下第三磨牙和保留的上尖牙。它在11-77微希之间变化。辐射剂量应与图像质量一起评估。两个单元分别改变管电压、管电流、旋转度和视场,获得颅骨幻象图像。7名观察员对两项诊断任务的主观图像质量进行了评估,这两项诊断任务是:根尖周围诊断和下颌后部种植体规划。观察员内部一致意见良好,观察员间一致意见一般。与种植体计划相比,无论颌骨如何,根尖周围诊断需要更高的暴露参数。与上颌相比,下颌种植体计划需要更高的暴露参数。通过使用180度旋转,特别是上颌种植体计划,可以在不丢失诊断信息的情况下大量减少剂量。发现小视场的CBCT非常适合于根尖周诊断和种植计划。CTDI法不适用于这些单位的有效剂量估算。在回顾性选择的患者材料中,根据DAP值,有效剂量在11-77微西沃特之间变化(ICRP 60, 1991)。根据诊断任务调整照射参数可大幅度降低剂量。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书