Estimation of adaptive computed tomography dose index based on body weight in pediatric patients.

Radiation medicine Pub Date : 2008-02-01 Epub Date: 2008-02-27 DOI:10.1007/s11604-007-0199-2
Osamu Miyazaki, Tetsuya Horiuchi, Hidekazu Masaki, Shunsuke Nosaka, Mikiko Miyasaka, Yoshiyuki Tsutsumi, Yoshiyuki Okada, Masayuki Kitamura
{"title":"Estimation of adaptive computed tomography dose index based on body weight in pediatric patients.","authors":"Osamu Miyazaki,&nbsp;Tetsuya Horiuchi,&nbsp;Hidekazu Masaki,&nbsp;Shunsuke Nosaka,&nbsp;Mikiko Miyasaka,&nbsp;Yoshiyuki Tsutsumi,&nbsp;Yoshiyuki Okada,&nbsp;Masayuki Kitamura","doi":"10.1007/s11604-007-0199-2","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>The aim of this study was to create an adaptive computed tomography dose index (by volume) (CTDI vol) for pediatric patients that would be fitted to a patient's particular body weight and to compare the adaptive CTDI vol with the CTDI vol displayed on the screen of the CT console.</p><p><strong>Materials and methods: </strong>CT images of 60 children whose body weights were known were available for calculating the total amount of modified CT numbers as an attenuation area. The attenuation area values of four differentsized acryl phantoms were also calculated. The dose measurements of all four phantoms were carried out. We combined the results of the abdominal CT and phantom experiments. The weight-based complementary ratio for adaptive CTDI vol was calculated, and the result was applied to an example of pediatric abdominal CT protocol.</p><p><strong>Results: </strong>The adaptive CTDI vol was always larger than the displayed CTDI vol with both small and large fields of view (FOV). The adaptive CTDI vol was 2.2 times larger than the displayed CTDI vol in the maximum value with the large FOV and 1.2 times larger with the small FOV.</p><p><strong>Conclusion: </strong>We radiologists must be the child's advocate and protect children from the deleterious effects of any of our technologies, including a lower indication of CTDI vol on the screen of a CT console.</p>","PeriodicalId":49640,"journal":{"name":"Radiation medicine","volume":"26 2","pages":"98-103"},"PeriodicalIF":0.0000,"publicationDate":"2008-02-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.1007/s11604-007-0199-2","citationCount":"8","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Radiation medicine","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1007/s11604-007-0199-2","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2008/2/27 0:00:00","PubModel":"Epub","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 8

Abstract

Purpose: The aim of this study was to create an adaptive computed tomography dose index (by volume) (CTDI vol) for pediatric patients that would be fitted to a patient's particular body weight and to compare the adaptive CTDI vol with the CTDI vol displayed on the screen of the CT console.

Materials and methods: CT images of 60 children whose body weights were known were available for calculating the total amount of modified CT numbers as an attenuation area. The attenuation area values of four differentsized acryl phantoms were also calculated. The dose measurements of all four phantoms were carried out. We combined the results of the abdominal CT and phantom experiments. The weight-based complementary ratio for adaptive CTDI vol was calculated, and the result was applied to an example of pediatric abdominal CT protocol.

Results: The adaptive CTDI vol was always larger than the displayed CTDI vol with both small and large fields of view (FOV). The adaptive CTDI vol was 2.2 times larger than the displayed CTDI vol in the maximum value with the large FOV and 1.2 times larger with the small FOV.

Conclusion: We radiologists must be the child's advocate and protect children from the deleterious effects of any of our technologies, including a lower indication of CTDI vol on the screen of a CT console.

基于儿童体重的自适应计算机断层扫描剂量指数估算。
目的:本研究的目的是为儿童患者创建适合患者特定体重的自适应计算机断层扫描剂量指数(按体积)(CTDI vol),并将自适应CTDI vol与CT控制台屏幕上显示的CTDI vol进行比较。材料与方法:选取60例已知体重儿童的CT图像,计算修改后CT数的总量作为衰减区域。计算了四种不同尺寸的丙烯模的衰减面积值。对所有四种幻影进行了剂量测量。我们结合了腹部CT和幻像实验的结果。计算了自适应CTDI vol的基于体重的互补比,并将结果应用于儿科腹部CT方案实例。结果:无论视场大小,自适应CTDI vol均大于显示CTDI vol。在大视场条件下,自适应CTDI vol是显示CTDI vol的2.2倍,在小视场条件下是显示CTDI vol的1.2倍。结论:我们放射科医生必须成为儿童的倡导者,保护儿童免受我们任何技术的有害影响,包括CT控制台屏幕上CTDI vol的较低指示。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信