Determination of nebulizer droplet size distribution: a method based on impactor refrigeration.

Elna Berg, Jan Olof Svensson, Lars Asking
{"title":"Determination of nebulizer droplet size distribution: a method based on impactor refrigeration.","authors":"Elna Berg,&nbsp;Jan Olof Svensson,&nbsp;Lars Asking","doi":"10.1089/jam.2007.0556","DOIUrl":null,"url":null,"abstract":"<p><p>Size distributions of droplets generated by nebulizers are difficult to determine because of evaporation after aerosolization. We describe a method whereby a Next Generation Pharmaceutical Impactor (NGI; MSP Corporation, Shoreview, MN) is refrigerated at 5 degrees C before connecting it to the nebulizer in order to ensure an environment inside the NGI at close to 100% relative humidity (RH). This, in turn, reduces droplet evaporation between the nebulizer and impaction. The method development was performed with a Pari LC Plus jet nebulizer operated at 2.0 bar, with the NGI set at a flow rate of 15 L/min and with salbutamol 5.0 mg/mL as the test solution. The droplet size distributions were expressed in terms of mass median aerodynamic diameter (MMAD) and geometric standard deviation (GSD). Variation in test conditions showed that the NGI should be cooled for at least 90 min, that nebulization should be started within 5 min after removal from the refrigerator, and that coating of collecting cups to prevent \"bouncing\" is not necessary. Variation of ambient temperature and humidity had no relevant effect on results. MMAD and GSD results showed that refrigeration of the NGI resulted in droplet size distributions that are likely to reflect those originally delivered at the mouthpiece by the nebulizer. The method was shown to be robust, accurate with recovery of test solutions exceeding 99%, reproducible, and to be suitable for use with a wide range of commercially available nebulizers.</p>","PeriodicalId":14878,"journal":{"name":"Journal of aerosol medicine : the official journal of the International Society for Aerosols in Medicine","volume":"20 2","pages":"97-104"},"PeriodicalIF":0.0000,"publicationDate":"2007-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.1089/jam.2007.0556","citationCount":"56","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of aerosol medicine : the official journal of the International Society for Aerosols in Medicine","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1089/jam.2007.0556","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 56

Abstract

Size distributions of droplets generated by nebulizers are difficult to determine because of evaporation after aerosolization. We describe a method whereby a Next Generation Pharmaceutical Impactor (NGI; MSP Corporation, Shoreview, MN) is refrigerated at 5 degrees C before connecting it to the nebulizer in order to ensure an environment inside the NGI at close to 100% relative humidity (RH). This, in turn, reduces droplet evaporation between the nebulizer and impaction. The method development was performed with a Pari LC Plus jet nebulizer operated at 2.0 bar, with the NGI set at a flow rate of 15 L/min and with salbutamol 5.0 mg/mL as the test solution. The droplet size distributions were expressed in terms of mass median aerodynamic diameter (MMAD) and geometric standard deviation (GSD). Variation in test conditions showed that the NGI should be cooled for at least 90 min, that nebulization should be started within 5 min after removal from the refrigerator, and that coating of collecting cups to prevent "bouncing" is not necessary. Variation of ambient temperature and humidity had no relevant effect on results. MMAD and GSD results showed that refrigeration of the NGI resulted in droplet size distributions that are likely to reflect those originally delivered at the mouthpiece by the nebulizer. The method was shown to be robust, accurate with recovery of test solutions exceeding 99%, reproducible, and to be suitable for use with a wide range of commercially available nebulizers.

雾化器液滴大小分布的测定:一种基于冲击器制冷的方法。
雾化器产生的液滴的大小分布很难确定,因为雾化后会蒸发。我们描述了一种方法,即下一代药物冲击器(NGI;MSP Corporation, Shoreview, MN)在连接到雾化器之前在5℃下冷藏,以确保NGI内部的环境接近100%的相对湿度(RH)。这反过来又减少了雾化器和撞击之间的液滴蒸发。方法开发使用Pari LC Plus喷射雾化器,工作在2.0 bar, NGI设置为15 L/min,沙丁胺醇5.0 mg/mL为测试溶液。液滴尺寸分布以质量中值气动直径(MMAD)和几何标准差(GSD)表示。试验条件的变化表明,NGI应至少冷却90分钟,从冰箱取出后应在5分钟内开始雾化,并且不需要涂覆收集杯以防止“反弹”。环境温度和湿度的变化对结果没有相关影响。MMAD和GSD结果表明,NGI的制冷导致液滴大小分布很可能反映了最初由雾化器在口部输送的液滴大小分布。该方法被证明是可靠的,准确的,测试溶液的回收率超过99%,可重复,并且适用于广泛的市售雾化器。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信