Poor Interface Design and Lack of Usability Testing Facilitate Medical Error

Rollin J. Fairbanks M.D., M.S., N.R.E.M.T.P. (Assistant Professor), Stanley Caplan M.S., C.H.F.P. (President)
{"title":"Poor Interface Design and Lack of Usability Testing Facilitate Medical Error","authors":"Rollin J. Fairbanks M.D., M.S., N.R.E.M.T.P. (Assistant Professor),&nbsp;Stanley Caplan M.S., C.H.F.P. (President)","doi":"10.1016/S1549-3741(04)30068-7","DOIUrl":null,"url":null,"abstract":"<div><h3>Background</h3><p>A fictional scenario based on a compilation of several real events describes seven medical errors that at first appear to be caused by the paramedics and nurses involved.</p></div><div><h3>Human Factors Engineering (HFE) Analysis</h3><p>An emergency medical services paramedic attempted to use a debrillator on a 67-year-old man with ventricular tachycardia. Yet nothing happened. The defibrillator displayed an indication that it was in synchronized mode but provided no feedback to tell the user that it was not prepared to shock because of low QRS amplitude.</p></div><div><h3>Usability Testing</h3><p>A hands-on approach to discovering the difficulties and potential for error that people encounter when trying to use a product, usability testing can help to create medical devices and systems that are not only more “user friendly” and efficient—but safer.</p></div><div><h3>Recommendations</h3><p>Recommendations are presented to enable health care leaders to apply human factors considerations in their product evaluation and purchasing decisions. Medical device manufacturers should involve human factors engineers in the design process from the outset and should perform usability testing. Health care organizations should expect an optimized and tested user interface in the medical devices they purchase.</p></div><div><h3>Summary</h3><p>Many adverse events in medicine are the result of poor interface design rather than human error. The HFE concepts of usability and standardization are critical to patient safety.</p></div>","PeriodicalId":84970,"journal":{"name":"Joint Commission journal on quality and safety","volume":"30 10","pages":"Pages 579-584"},"PeriodicalIF":0.0000,"publicationDate":"2004-10-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.1016/S1549-3741(04)30068-7","citationCount":"83","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Joint Commission journal on quality and safety","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S1549374104300687","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2016/11/17 0:00:00","PubModel":"Epub","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 83

Abstract

Background

A fictional scenario based on a compilation of several real events describes seven medical errors that at first appear to be caused by the paramedics and nurses involved.

Human Factors Engineering (HFE) Analysis

An emergency medical services paramedic attempted to use a debrillator on a 67-year-old man with ventricular tachycardia. Yet nothing happened. The defibrillator displayed an indication that it was in synchronized mode but provided no feedback to tell the user that it was not prepared to shock because of low QRS amplitude.

Usability Testing

A hands-on approach to discovering the difficulties and potential for error that people encounter when trying to use a product, usability testing can help to create medical devices and systems that are not only more “user friendly” and efficient—but safer.

Recommendations

Recommendations are presented to enable health care leaders to apply human factors considerations in their product evaluation and purchasing decisions. Medical device manufacturers should involve human factors engineers in the design process from the outset and should perform usability testing. Health care organizations should expect an optimized and tested user interface in the medical devices they purchase.

Summary

Many adverse events in medicine are the result of poor interface design rather than human error. The HFE concepts of usability and standardization are critical to patient safety.

糟糕的界面设计和缺乏可用性测试会导致医疗错误
一个虚构的场景是根据几个真实事件汇编而成的,描述了七个医疗错误,乍一看,这些错误似乎是由相关的护理人员和护士造成的。人为因素工程(HFE)分析一名紧急医疗服务护理人员试图对一名患有室性心动过速的67岁男子使用除颤器。然而什么也没发生。除颤器显示它处于同步模式,但没有提供反馈告诉用户它不准备电击,因为QRS振幅低。可用性测试是一种亲自动手的方法,可以发现人们在尝试使用产品时遇到的困难和潜在的错误,可用性测试可以帮助创建更加“用户友好”和高效的医疗设备和系统,而且更安全。建议提出的建议是为了使医疗保健领导者能够在产品评估和购买决策中考虑人为因素。医疗设备制造商应该从一开始就让人为因素工程师参与设计过程,并进行可用性测试。医疗保健组织应该期望在他们购买的医疗设备中有一个经过优化和测试的用户界面。医学中的许多不良事件是界面设计不良的结果,而不是人为错误。HFE的可用性和标准化概念对患者安全至关重要。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书