Evaluation of a personal monitor employing an electrochemical sensor for assessing exposure to hydrogen peroxide in the workplace.

Jeongim Park, Marcia R Plese, Mark A Puskar
{"title":"Evaluation of a personal monitor employing an electrochemical sensor for assessing exposure to hydrogen peroxide in the workplace.","authors":"Jeongim Park,&nbsp;Marcia R Plese,&nbsp;Mark A Puskar","doi":"10.1080/15428110308984828","DOIUrl":null,"url":null,"abstract":"<p><p>A commercially available direct-reading instrument designed for personal monitoring of vapor phase hydrogen peroxide (VHP) was evaluated in the laboratory and the workplace. Monitoring VHP has gained increasing importance in the pharmaceutical industry because sterilization using VHP has proven to be a good alternative to previously used sterilizing methods. The current Occupational Safety and Health Administration impinger method for VHP measurements, based on bubbling air through an acid solution with subsequent laboratory analysis, is not practical for monitoring personal exposures. By employing an electrochemical sensor, the instrument evaluated provides real-time exposure data with auxiliary functions such as displaying concentrations in parts per million, data logging, and alarms. A double-dilution technique using a syringe pump was used to generate dynamic test atmospheres ranging from 0.2 to 10 ppm in an exposure chamber. Time-weighted average concentration data from the direct-reading instrument was compared with concentration data from the impingers. The overall accuracy was less than the +/-25%, National Institute for Occupational Safety and Health criterion. No significant differences in accuracy were observed at three humidity levels (i.e., 15, 50, and 80%). The instrument was similarly evaluated in a workplace under typical conditions. The results agreed within +/-0.2 ppm. Selected performance characteristics of the instrument also were investigated, including reproducibility, response and recovery times, calibration frequency, and suitability of the calibration adapter. Results of the investigation suggest that the instrument provides a means for simple and accurate monitoring of personal exposures to VHP in workplace environments.</p>","PeriodicalId":83618,"journal":{"name":"AIHA journal : a journal for the science of occupational and environmental health and safety","volume":"64 3","pages":"360-7"},"PeriodicalIF":0.0000,"publicationDate":"2003-05-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.1080/15428110308984828","citationCount":"5","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"AIHA journal : a journal for the science of occupational and environmental health and safety","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1080/15428110308984828","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 5

Abstract

A commercially available direct-reading instrument designed for personal monitoring of vapor phase hydrogen peroxide (VHP) was evaluated in the laboratory and the workplace. Monitoring VHP has gained increasing importance in the pharmaceutical industry because sterilization using VHP has proven to be a good alternative to previously used sterilizing methods. The current Occupational Safety and Health Administration impinger method for VHP measurements, based on bubbling air through an acid solution with subsequent laboratory analysis, is not practical for monitoring personal exposures. By employing an electrochemical sensor, the instrument evaluated provides real-time exposure data with auxiliary functions such as displaying concentrations in parts per million, data logging, and alarms. A double-dilution technique using a syringe pump was used to generate dynamic test atmospheres ranging from 0.2 to 10 ppm in an exposure chamber. Time-weighted average concentration data from the direct-reading instrument was compared with concentration data from the impingers. The overall accuracy was less than the +/-25%, National Institute for Occupational Safety and Health criterion. No significant differences in accuracy were observed at three humidity levels (i.e., 15, 50, and 80%). The instrument was similarly evaluated in a workplace under typical conditions. The results agreed within +/-0.2 ppm. Selected performance characteristics of the instrument also were investigated, including reproducibility, response and recovery times, calibration frequency, and suitability of the calibration adapter. Results of the investigation suggest that the instrument provides a means for simple and accurate monitoring of personal exposures to VHP in workplace environments.

使用电化学传感器评估工作场所过氧化氢暴露的个人监测器的评价。
在实验室和工作场所对一种市售的用于个人监测气相过氧化氢(VHP)的直读仪器进行了评估。监测VHP在制药工业中越来越重要,因为使用VHP灭菌已被证明是以前使用的灭菌方法的一个很好的替代方法。目前美国职业安全与健康管理局(Occupational Safety and Health Administration)用于VHP测量的撞击法是基于通过酸溶液的鼓泡空气,然后进行实验室分析,对于监测个人暴露并不实用。通过采用电化学传感器,该仪器可提供实时暴露数据,并具有辅助功能,如显示百万分之一的浓度、数据记录和警报。使用注射泵的双重稀释技术在暴露室中产生0.2至10ppm的动态测试气氛。将直读仪测得的时间加权平均浓度数据与撞击器测得的浓度数据进行比较。总体准确度低于+/-25%,国家职业安全与健康研究所的标准。在三种湿度水平(即15%、50%和80%)下,准确度没有显著差异。该仪器在典型条件下的工作场所进行了类似的评估。结果在+/-0.2 ppm范围内一致。还研究了仪器的选定性能特征,包括再现性、响应和恢复时间、校准频率和校准适配器的适用性。调查结果表明,该仪器提供了一种简单而准确地监测工作场所环境中个人接触VHP的手段。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信