{"title":"[The studies on microtensile bond strength measurements of two dentin adhesives].","authors":"B Xie, S H Dickens, G E Schumacher, A A Giuseppetti","doi":"","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>The purpose of this study was to investigate the feasibility of a modified microtensile method used to test bond strengths of two current one-bottle dentin bond systems (Prime & Bond NT, PBNT; Prime one Mirage, P-One) with a parallel match design.</p><p><strong>Methods: </strong>15 extracted, caries-free human molars were cut to expose occlusal dentin. A 5 mm deep slot was prepared in each crown to divide the crown into nearly equal halves for accepting treatment of the two dentin bond systems, respectively. After 24 h storage in distilled water at 37 degrees C, the bonded teeth were subjected to two treatments: 5 teeth were tested without further treatment and 10 teeth were thermocycled (2400 cycles, between 5 degrees C and 55 degrees C) prior to bond strength testing. Hour-glass shaped specimens with a distance of approximately 1.0 mm at the narrowest portion were cut from each tooth and tested in tensile mode.</p><p><strong>Results: </strong>Bond strengths (mean MPa) were: for PBNT: 42 & 31, and for P-One 64 & 38 without and with thermocycling, respectively. Two-way ANOVA revealed a significant difference in bond strengths(P < 0.001) between the two systems and when thermocycled. However, a pairwise multiple comparison (Tukey test) showed that after thermocycling the difference between the two systems was not significant (P > 0.05). Regression analysis showed that a correlation existed between the two systems' tensile bond strength values grouped by tooth (correlation coefficient r = 0.575, P < 0.05).</p><p><strong>Conclusion: </strong>The modified microtensile method with a parallel match design is feasible and suitable for evaluating two different bonding systems or dentin treatments.</p>","PeriodicalId":13173,"journal":{"name":"Hua xi yi ke da xue xue bao = Journal of West China University of Medical Sciences = Huaxi yike daxue xuebao","volume":"32 4","pages":"519-23"},"PeriodicalIF":0.0000,"publicationDate":"2001-12-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Hua xi yi ke da xue xue bao = Journal of West China University of Medical Sciences = Huaxi yike daxue xuebao","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0
Abstract
Objective: The purpose of this study was to investigate the feasibility of a modified microtensile method used to test bond strengths of two current one-bottle dentin bond systems (Prime & Bond NT, PBNT; Prime one Mirage, P-One) with a parallel match design.
Methods: 15 extracted, caries-free human molars were cut to expose occlusal dentin. A 5 mm deep slot was prepared in each crown to divide the crown into nearly equal halves for accepting treatment of the two dentin bond systems, respectively. After 24 h storage in distilled water at 37 degrees C, the bonded teeth were subjected to two treatments: 5 teeth were tested without further treatment and 10 teeth were thermocycled (2400 cycles, between 5 degrees C and 55 degrees C) prior to bond strength testing. Hour-glass shaped specimens with a distance of approximately 1.0 mm at the narrowest portion were cut from each tooth and tested in tensile mode.
Results: Bond strengths (mean MPa) were: for PBNT: 42 & 31, and for P-One 64 & 38 without and with thermocycling, respectively. Two-way ANOVA revealed a significant difference in bond strengths(P < 0.001) between the two systems and when thermocycled. However, a pairwise multiple comparison (Tukey test) showed that after thermocycling the difference between the two systems was not significant (P > 0.05). Regression analysis showed that a correlation existed between the two systems' tensile bond strength values grouped by tooth (correlation coefficient r = 0.575, P < 0.05).
Conclusion: The modified microtensile method with a parallel match design is feasible and suitable for evaluating two different bonding systems or dentin treatments.