{"title":"Metallic surface modification.","authors":"I De Scheerder, E Verbeken, J Van Humbeeck","doi":"","DOIUrl":null,"url":null,"abstract":"<p><p>The potential beneficial effect of metal surface treatment using electrochemical polishing on stent thrombogenicity and neointimal hyperplasia was evaluated in a rat A-V model and a porcine coronary model. Thrombogenicity of polished stents (n=6) was compared to non-polished stents (n=5) in a rat A-V shunt model using 125I-fibrinogen and 51Cr-labelled platelets. Total clot weight after 30 min was significantly lower in the polished stents (32.1+/-2.8 vs 18.1+/-4.4: p<0.001). Also, 125I-fibrinogen deposition was significantly lower in the polished stents (1.30+/-0.07 vs 0.66+/-0.04: p<0.001). Platelet deposition was, however, not significantly reduced (12.7+/-3.4 vs 9.87+/-1.9, NS). Subsequently, the effect of electrochemical polishing on neointimal hyperplasia was evaluated in a porcine coronary model. Polished (n=10) and non-polished stents (n=10) were randomly implanted in the right coronary artery of healthy pigs. Neointimal hyperplasia was significantly decreased in the polished stents (0.56+/-0.28 vs 0.94+/-0.34 mm2: p<0.01).</p>","PeriodicalId":79534,"journal":{"name":"Seminars in interventional cardiology : SIIC","volume":"3 3-4","pages":"139-44"},"PeriodicalIF":0.0000,"publicationDate":"1998-09-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Seminars in interventional cardiology : SIIC","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0
Abstract
The potential beneficial effect of metal surface treatment using electrochemical polishing on stent thrombogenicity and neointimal hyperplasia was evaluated in a rat A-V model and a porcine coronary model. Thrombogenicity of polished stents (n=6) was compared to non-polished stents (n=5) in a rat A-V shunt model using 125I-fibrinogen and 51Cr-labelled platelets. Total clot weight after 30 min was significantly lower in the polished stents (32.1+/-2.8 vs 18.1+/-4.4: p<0.001). Also, 125I-fibrinogen deposition was significantly lower in the polished stents (1.30+/-0.07 vs 0.66+/-0.04: p<0.001). Platelet deposition was, however, not significantly reduced (12.7+/-3.4 vs 9.87+/-1.9, NS). Subsequently, the effect of electrochemical polishing on neointimal hyperplasia was evaluated in a porcine coronary model. Polished (n=10) and non-polished stents (n=10) were randomly implanted in the right coronary artery of healthy pigs. Neointimal hyperplasia was significantly decreased in the polished stents (0.56+/-0.28 vs 0.94+/-0.34 mm2: p<0.01).