Validation of reflectance pulse oximetry: an evaluation of a new sensor in piglets.

R Nijland, S Nierlich, H W Jongsma, J G Nijhuis, B Oeseburg, K Springer, P Mannheimer
{"title":"Validation of reflectance pulse oximetry: an evaluation of a new sensor in piglets.","authors":"R Nijland,&nbsp;S Nierlich,&nbsp;H W Jongsma,&nbsp;J G Nijhuis,&nbsp;B Oeseburg,&nbsp;K Springer,&nbsp;P Mannheimer","doi":"10.1023/a:1007332105965","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>A new reflectance pulse oximetry sensor, developed for intrapartum estimation of arterial oxygen saturation (SaO2), was calibrated and evaluated. The sensor contains two light emitting diodes of 735 and 890 nm, and a photodetector at a distance of 14 mm from both light emitting diodes.</p><p><strong>Methods: </strong>In seven Yorkshire/Hampshire piglets, the reflectance sensor (Nellcor Puritan Bennett Inc.) was calibrated using blood sample SaO2 values. The resulting calibration line was evaluated in four Dutch piglets, by comparing pulse oximetry saturation readings (SpO2) with blood sample and intravascular fiberoptic oximetry SaO2 values. Several reflectance sensors were fixed on each animal. Desaturation levels were obtained by changing the gas mixture of oxygen/ nitrous oxide via a tracheal catheter.</p><p><strong>Results: </strong>In the Yorkshire/ Hampshire piglets, the standard deviation of difference (SpO2-SaO2) was 4.7% (n = 364), over an SaO2 range of 17% to 100%. In the Dutch piglets, the mean difference (SpO2-SaO2) was -1.6% and the standard deviation of difference was 5.4%, over the same SaO2 range (n = 254). Comparisons of continuous recordings of reflectance SpO2 and fiberoptic SaO2 revealed variation in individual regression lines.</p><p><strong>Conclusions: </strong>This new 735/890 nm reflectance sensor demonstrates acceptable accuracy in piglets. A further evaluation during labor should assess its feasibility for fetal surveillance.</p>","PeriodicalId":77199,"journal":{"name":"Journal of clinical monitoring","volume":"13 1","pages":"43-9"},"PeriodicalIF":0.0000,"publicationDate":"1997-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.1023/a:1007332105965","citationCount":"10","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of clinical monitoring","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1023/a:1007332105965","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 10

Abstract

Objective: A new reflectance pulse oximetry sensor, developed for intrapartum estimation of arterial oxygen saturation (SaO2), was calibrated and evaluated. The sensor contains two light emitting diodes of 735 and 890 nm, and a photodetector at a distance of 14 mm from both light emitting diodes.

Methods: In seven Yorkshire/Hampshire piglets, the reflectance sensor (Nellcor Puritan Bennett Inc.) was calibrated using blood sample SaO2 values. The resulting calibration line was evaluated in four Dutch piglets, by comparing pulse oximetry saturation readings (SpO2) with blood sample and intravascular fiberoptic oximetry SaO2 values. Several reflectance sensors were fixed on each animal. Desaturation levels were obtained by changing the gas mixture of oxygen/ nitrous oxide via a tracheal catheter.

Results: In the Yorkshire/ Hampshire piglets, the standard deviation of difference (SpO2-SaO2) was 4.7% (n = 364), over an SaO2 range of 17% to 100%. In the Dutch piglets, the mean difference (SpO2-SaO2) was -1.6% and the standard deviation of difference was 5.4%, over the same SaO2 range (n = 254). Comparisons of continuous recordings of reflectance SpO2 and fiberoptic SaO2 revealed variation in individual regression lines.

Conclusions: This new 735/890 nm reflectance sensor demonstrates acceptable accuracy in piglets. A further evaluation during labor should assess its feasibility for fetal surveillance.

反射式脉搏血氧仪的验证:一种新型传感器在仔猪中的评价。
目的:研制一种用于产时动脉血氧饱和度(SaO2)测定的新型反射式脉搏血氧仪,并对其进行标定和评价。该传感器包含两个735 nm和890 nm的发光二极管,并在距离两个发光二极管14mm处安装一个光电探测器。方法:选用7头约克郡/汉普郡仔猪,利用血液样本SaO2值对反射传感器(Nellcor Puritan Bennett Inc.)进行校准。通过将脉搏血氧饱和度(SpO2)读数与血液样本和血管内纤维血氧饱和度(SaO2)值进行比较,对4头荷兰仔猪的校准线进行评估。在每只动物身上都固定了几个反射传感器。通过气管导管改变氧/氧化亚氮的气体混合物来获得去饱和水平。结果:在约克郡/汉普郡仔猪中,SpO2-SaO2差异的标准差为4.7% (n = 364), SaO2范围为17% ~ 100%。在相同的SaO2范围内(n = 254),荷兰仔猪的SpO2-SaO2平均差值为-1.6%,差值标准差为5.4%。连续记录的SpO2反射率和光纤SaO2的比较显示了个别回归线的变化。结论:该新型735/890 nm反射传感器在仔猪中具有良好的准确性。产程进一步评估应评估其胎儿监测的可行性。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信