Quantitative magnetic resonance methods for in vivo investigation of the human liver and spleen. Technical aspects and preliminary clinical results.

Acta radiologica. Supplementum Pub Date : 1996-01-01
C Thomsen
{"title":"Quantitative magnetic resonance methods for in vivo investigation of the human liver and spleen. Technical aspects and preliminary clinical results.","authors":"C Thomsen","doi":"","DOIUrl":null,"url":null,"abstract":"<p><p>This project was initiated with the introduction of magnetic resonance (MR) in Denmark in order to evaluate the possibilities of this technique as a diagnostic aid in non-focal liver and splenic diseases. The signal intensities in the MR image are sensitive to the longitudinal relaxation (T1), the transverse relaxation (T2), flow and chemical shift. All these parameters may be quantified by developing specific pulse sequences sensitive to the parameter in question. Previous studies had indicated that relaxation time measurements might be of value in the diagnosis of liver cirrhosis and haemochromatosis. Measuring relaxation times in these 2 groups of patients posed different challenges. In patients with liver cirrhosis a method had to be developed for simultaneous T1 and T2 relaxation time measurements, which was robust to the respiratory motion of the liver. A combination of multi-echo pulse sequences with different repetition times was chosen, because motion effects were partly refocused. Multi-acquisition was used to improve the signal-to-noise ratio in the heavily saturated experiments with short repetition times, to further reduce the sensitivity to motion. To test the quality of this pulse sequence, phantom experiments were performed, and sensitivity to motion was tested by measuring with and without respiratory synchronization. Respiratory synchronization gave a marked improvement in focal liver diseases, whereas no difference was found in non-focal diseases. Standard imaging sequences with a minimum echo time of 30 ms could not be used to measure the short T2 relaxation times found in patients with increased liver iron. A volume-selective multi-echo spectroscopic pulse was developed with a minimum echo time of 4 ms. Biexponential signal decay could be shown in patients with increased liver iron by using this sequence. Patients with liver cirrhosis, as a group, had increased T1 relaxation times compared to normal volunteers, but an overlap in T1 values was found. No correlation between the degree of fibrosis and the T1 relaxation time was found. Liver iron concentration could be quantified either by using the fast component of the T2 signal decay or by using the decreased signal in spin-echo and gradient echo images. Patients with leukemias and myeloproliferative disorders had prolonged T1 relaxation times in the spleen, but a considerable overlap was found between this group and a group of patients with benign hyperplasia and patients with splenomegaly secondary to portal hypertension. Volume-selective proton spectroscopy was developed and used to quantify the liver fat concentration. The accuracy of the method was about 3 g/100 g. With the implementation of a second generation scanner system it became possible to develop a pulse sequence, using the phase information in the MR signal, to measure portal vein flow during breath-holding. This method made it possible to estimate the portal vein flow during fasting, and the flow increase after eating. Quantitative MR methods may contribute to the diagnosis of non-focal liver diseases by estimation of liver fat and liver iron and by assessment of portal vein blood flow. Increased T1 relaxation time is a sign of a disease process in the liver rather than specific for any liver disease.</p>","PeriodicalId":7159,"journal":{"name":"Acta radiologica. Supplementum","volume":"401 ","pages":"1-34"},"PeriodicalIF":0.0000,"publicationDate":"1996-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Acta radiologica. Supplementum","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

This project was initiated with the introduction of magnetic resonance (MR) in Denmark in order to evaluate the possibilities of this technique as a diagnostic aid in non-focal liver and splenic diseases. The signal intensities in the MR image are sensitive to the longitudinal relaxation (T1), the transverse relaxation (T2), flow and chemical shift. All these parameters may be quantified by developing specific pulse sequences sensitive to the parameter in question. Previous studies had indicated that relaxation time measurements might be of value in the diagnosis of liver cirrhosis and haemochromatosis. Measuring relaxation times in these 2 groups of patients posed different challenges. In patients with liver cirrhosis a method had to be developed for simultaneous T1 and T2 relaxation time measurements, which was robust to the respiratory motion of the liver. A combination of multi-echo pulse sequences with different repetition times was chosen, because motion effects were partly refocused. Multi-acquisition was used to improve the signal-to-noise ratio in the heavily saturated experiments with short repetition times, to further reduce the sensitivity to motion. To test the quality of this pulse sequence, phantom experiments were performed, and sensitivity to motion was tested by measuring with and without respiratory synchronization. Respiratory synchronization gave a marked improvement in focal liver diseases, whereas no difference was found in non-focal diseases. Standard imaging sequences with a minimum echo time of 30 ms could not be used to measure the short T2 relaxation times found in patients with increased liver iron. A volume-selective multi-echo spectroscopic pulse was developed with a minimum echo time of 4 ms. Biexponential signal decay could be shown in patients with increased liver iron by using this sequence. Patients with liver cirrhosis, as a group, had increased T1 relaxation times compared to normal volunteers, but an overlap in T1 values was found. No correlation between the degree of fibrosis and the T1 relaxation time was found. Liver iron concentration could be quantified either by using the fast component of the T2 signal decay or by using the decreased signal in spin-echo and gradient echo images. Patients with leukemias and myeloproliferative disorders had prolonged T1 relaxation times in the spleen, but a considerable overlap was found between this group and a group of patients with benign hyperplasia and patients with splenomegaly secondary to portal hypertension. Volume-selective proton spectroscopy was developed and used to quantify the liver fat concentration. The accuracy of the method was about 3 g/100 g. With the implementation of a second generation scanner system it became possible to develop a pulse sequence, using the phase information in the MR signal, to measure portal vein flow during breath-holding. This method made it possible to estimate the portal vein flow during fasting, and the flow increase after eating. Quantitative MR methods may contribute to the diagnosis of non-focal liver diseases by estimation of liver fat and liver iron and by assessment of portal vein blood flow. Increased T1 relaxation time is a sign of a disease process in the liver rather than specific for any liver disease.

定量磁共振方法在体内研究人肝脏和脾脏。技术方面和初步临床结果。
该项目是随着磁共振(MR)在丹麦的引入而启动的,目的是评估该技术作为非局灶性肝脏和脾脏疾病诊断辅助手段的可能性。磁共振图像中的信号强度对纵向弛豫(T1)、横向弛豫(T2)、流动和化学位移敏感。所有这些参数都可以通过开发对所讨论的参数敏感的特定脉冲序列来量化。以往的研究表明,松弛时间测量可能在肝硬化和血色素沉着病的诊断中有价值。测量这两组患者的放松时间提出了不同的挑战。对于肝硬化患者,必须开发一种同时测量T1和T2松弛时间的方法,该方法对肝脏的呼吸运动具有鲁棒性。选择不同重复次数的多回波脉冲序列组合,因为运动效果部分重新聚焦。在低重复次数的高饱和实验中,采用多采集提高了信噪比,进一步降低了对运动的灵敏度。为了测试该脉冲序列的质量,进行了幻像实验,并通过测量有无呼吸同步来测试对运动的敏感性。呼吸同步对局灶性肝病有明显的改善,而对非局灶性肝病没有发现差异。最低回声时间为30 ms的标准成像序列不能用于测量肝铁增高患者的短T2松弛时间。研制了一种最小回波时间为4 ms的体积选择性多回波光谱脉冲。双指数信号衰减可以在肝铁增加的患者中显示。肝硬化患者作为一个群体,与正常志愿者相比,T1松弛时间增加,但T1值存在重叠。T1弛豫时间与纤维化程度无相关性。肝铁浓度既可以利用T2信号衰减的快速分量,也可以利用自旋回波和梯度回波图像中的衰减信号来定量。白血病和骨髓增生性疾病患者脾脏T1舒张时间延长,但本组与良性增生和门静脉高压症继发脾大患者有相当大的重叠。采用体积选择性质子光谱法定量肝脏脂肪浓度。该方法的准确度约为3 g/100 g。随着第二代扫描仪系统的实施,可以利用MR信号中的相位信息来开发脉冲序列,以测量屏气期间的门静脉流量。这种方法可以估计空腹时门静脉流量,以及进食后门静脉流量的增加。定量MR方法可以通过估计肝脏脂肪和肝铁以及评估门静脉血流来帮助诊断非局灶性肝病。T1松弛时间增加是肝脏疾病进程的标志,而不是任何肝脏疾病所特有的。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信