Accuracy and precision of serum gastrin measurements in commercial laboratories.

V Notar-Francesco, S Samuel, S Kanakamedala, E W Straus
{"title":"Accuracy and precision of serum gastrin measurements in commercial laboratories.","authors":"V Notar-Francesco,&nbsp;S Samuel,&nbsp;S Kanakamedala,&nbsp;E W Straus","doi":"","DOIUrl":null,"url":null,"abstract":"<p><p>Patients referred to us with \"positive\" secretin tests and the diagnosis of Zollinger-Ellison syndrome were found to be achlorhydric. This observation led us to study prospectively the accuracy and precision of serum gastrin determinations from commercial laboratories. Synthetic gastrin (G17) was added to serum to achieve gastrin concentrations of 50, 100, 250, 500, 750, 1000, 3000, and 5000 pg/mL after subtraction of the basal value (24 pg/mL). Three aliquots of each concentration were analyzed by radioimmunoassay in our laboratory (Health Science Center at Brooklyn) and sent to four major commercial laboratories that perform 5000 to 25,000 gastrin assays per year. The reported gastrin concentrations of the triplicate samples demonstrate that many commercial laboratories failed to accurately measure gastrin. Commercial laboratories generally reported higher-than-actual gastrin concentrations in samples containing less than 500 pg/mL and lower-than-actual gastrin concentrations in samples containing more than 500 pg/mL. Of all aliquots containing 100 pg/mL or less, 14 of 24 samples (58%) were reported by commercial laboratories to contain elevated gastrin concentrations. At gastrin concentrations from 250 to 5000 pg/mL, the range of values (highest- to lowest-reported value for each concentration) was greater than 200 pg/mL in 62% of triplicate samples reported by commercial laboratories. These data indicate that determinations by some commercial laboratories lack the precision required to satisfy the current diagnostic criterion (a postsecretin rise from basal gastrin of 200 pg/mL or greater) for Zollinger-Ellison syndrome. Clinicians should be aware of this problem and obtain more basal serum gastrin samples to allow for an analysis of the range of baseline values prior to secretin injection.</p>","PeriodicalId":77227,"journal":{"name":"Journal of the Association for Academic Minority Physicians : the official publication of the Association for Academic Minority Physicians","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"1995-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of the Association for Academic Minority Physicians : the official publication of the Association for Academic Minority Physicians","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Patients referred to us with "positive" secretin tests and the diagnosis of Zollinger-Ellison syndrome were found to be achlorhydric. This observation led us to study prospectively the accuracy and precision of serum gastrin determinations from commercial laboratories. Synthetic gastrin (G17) was added to serum to achieve gastrin concentrations of 50, 100, 250, 500, 750, 1000, 3000, and 5000 pg/mL after subtraction of the basal value (24 pg/mL). Three aliquots of each concentration were analyzed by radioimmunoassay in our laboratory (Health Science Center at Brooklyn) and sent to four major commercial laboratories that perform 5000 to 25,000 gastrin assays per year. The reported gastrin concentrations of the triplicate samples demonstrate that many commercial laboratories failed to accurately measure gastrin. Commercial laboratories generally reported higher-than-actual gastrin concentrations in samples containing less than 500 pg/mL and lower-than-actual gastrin concentrations in samples containing more than 500 pg/mL. Of all aliquots containing 100 pg/mL or less, 14 of 24 samples (58%) were reported by commercial laboratories to contain elevated gastrin concentrations. At gastrin concentrations from 250 to 5000 pg/mL, the range of values (highest- to lowest-reported value for each concentration) was greater than 200 pg/mL in 62% of triplicate samples reported by commercial laboratories. These data indicate that determinations by some commercial laboratories lack the precision required to satisfy the current diagnostic criterion (a postsecretin rise from basal gastrin of 200 pg/mL or greater) for Zollinger-Ellison syndrome. Clinicians should be aware of this problem and obtain more basal serum gastrin samples to allow for an analysis of the range of baseline values prior to secretin injection.

商业实验室血清胃泌素测定的准确性和精密度。
患者提交给我们的“阳性”分泌素试验和佐林格-埃里森综合征的诊断被发现是盐酸。这一观察结果使我们对商业实验室测定血清胃泌素的准确性和精密度进行前瞻性研究。将合成胃泌素(G17)加入血清中,减去基础值(24 pg/mL)后,胃泌素浓度分别为50、100、250、500、750、1000、3000和5000 pg/mL。我们的实验室(布鲁克林健康科学中心)对每种浓度的三等分进行放射免疫分析,并将其送到每年进行5000至25,000次胃泌素检测的四个主要商业实验室。报告的三个重复样品的胃泌素浓度表明,许多商业实验室未能准确测量胃泌素。商业实验室通常报告在低于500 pg/mL的样品中胃泌素浓度高于实际水平,而在高于500 pg/mL的样品中胃泌素浓度低于实际水平。在所有含有100pg /mL或以下的等分液中,商业实验室报告24个样品中有14个(58%)含有升高的胃泌素浓度。在胃泌素浓度为250至5000 pg/mL时,62%的商业实验室报告的三次重复样品的值范围(每种浓度的最高至最低报告值)大于200 pg/mL。这些数据表明,一些商业实验室的测定缺乏满足当前佐林格-埃里森综合征诊断标准(基础胃泌素后分泌素升高200 pg/mL或更高)所需的精度。临床医生应该意识到这个问题,并获得更多的基础血清胃泌素样本,以便在分泌素注射前分析基线值的范围。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信