{"title":"Design and evaluation of a device for measurement of interface pressure","authors":"M.D. Steinberg , E.D. Cooke","doi":"10.1016/0141-5425(93)90059-8","DOIUrl":null,"url":null,"abstract":"<div><p>An electropneumatic device has been designed to measure the interface pressure profile under compression bandages. The device uses commercially available pneumatic sensors (Talley Group Ltd, SJ235/3) and measures interface pressure at up to eight sites simultaneously along the lower limb, with an accuracy of ± 3 mmHg. Measurements can be made in one of two modes: continuously at a rate of up to three samples per second with the results output to a suitable display device, or single measurements of interface pressure can be made and displayed on a digital display incorporated in the device. This enables the monitoring and recording of interface pressure under compression bandages during either ambulation or passive recumbency. The electropneumatic system is described together with its following characteristics: the hysteresis of the Talley pneumatic sensors, the accuracy of pressure measurement and the maximum achievable sample rate. Dynamic measurements in a single normal volunteer are shown.</p></div>","PeriodicalId":75992,"journal":{"name":"Journal of biomedical engineering","volume":"15 6","pages":"Pages 464-468"},"PeriodicalIF":0.0000,"publicationDate":"1993-11-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.1016/0141-5425(93)90059-8","citationCount":"17","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of biomedical engineering","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/0141542593900598","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 17
Abstract
An electropneumatic device has been designed to measure the interface pressure profile under compression bandages. The device uses commercially available pneumatic sensors (Talley Group Ltd, SJ235/3) and measures interface pressure at up to eight sites simultaneously along the lower limb, with an accuracy of ± 3 mmHg. Measurements can be made in one of two modes: continuously at a rate of up to three samples per second with the results output to a suitable display device, or single measurements of interface pressure can be made and displayed on a digital display incorporated in the device. This enables the monitoring and recording of interface pressure under compression bandages during either ambulation or passive recumbency. The electropneumatic system is described together with its following characteristics: the hysteresis of the Talley pneumatic sensors, the accuracy of pressure measurement and the maximum achievable sample rate. Dynamic measurements in a single normal volunteer are shown.
设计了一种电-气装置,用于测量绷带压迫下的界面压力分布。该设备使用市售气动传感器(Talley Group Ltd, SJ235/3),可同时测量下肢多达8个位置的界面压力,精度为±3 mmHg。测量可以在两种模式之一中进行:以每秒最多三个样本的速率连续输出结果到合适的显示设备,或者可以在设备中集成的数字显示器上显示界面压力的单个测量。这使得监测和记录界面压力下的压缩绷带在走动或被动卧卧期间。介绍了电气动系统及其以下特点:Talley气动传感器的滞回性、压力测量的准确性和可实现的最大采样率。显示了单个正常志愿者的动态测量结果。