Quantitative analysis of four EMG amplifiers

E.J. Perreault, I.W. Hunter, R.E. Kearney
{"title":"Quantitative analysis of four EMG amplifiers","authors":"E.J. Perreault,&nbsp;I.W. Hunter,&nbsp;R.E. Kearney","doi":"10.1016/0141-5425(93)90079-E","DOIUrl":null,"url":null,"abstract":"<div><p>Four typical EMG amplifiers were tested quantitatively to observe the diversity and specificity of available equipment. Gain, phase, common mode rejection ratio (CMRR) and noise characteristics were measured for each device. Various gain and phase responses were observed, each best suited to specific application areas. For all amplifiers, the CMRR was shown to decrease dramatically in the presence of input impedance mismatches of more than 10 kΩ between the two electrodes. Because such impedance mismatches are common on the skin surface, these results indicate that proper skin preparation is required to maximize the noise rejection capabilities of the tested amplifiers.</p></div>","PeriodicalId":75992,"journal":{"name":"Journal of biomedical engineering","volume":"15 5","pages":"Pages 413-419"},"PeriodicalIF":0.0000,"publicationDate":"1993-09-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.1016/0141-5425(93)90079-E","citationCount":"13","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of biomedical engineering","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/014154259390079E","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 13

Abstract

Four typical EMG amplifiers were tested quantitatively to observe the diversity and specificity of available equipment. Gain, phase, common mode rejection ratio (CMRR) and noise characteristics were measured for each device. Various gain and phase responses were observed, each best suited to specific application areas. For all amplifiers, the CMRR was shown to decrease dramatically in the presence of input impedance mismatches of more than 10 kΩ between the two electrodes. Because such impedance mismatches are common on the skin surface, these results indicate that proper skin preparation is required to maximize the noise rejection capabilities of the tested amplifiers.

四个肌电放大器的定量分析
对四种典型的肌电放大器进行了定量测试,以观察可用设备的多样性和特异性。测量了每个器件的增益、相位、共模抑制比(CMRR)和噪声特性。观察到各种增益和相位响应,每个都最适合特定的应用领域。对于所有放大器,在两个电极之间输入阻抗失配大于10 kΩ的情况下,CMRR显着降低。由于这种阻抗不匹配在皮肤表面很常见,这些结果表明,需要适当的皮肤准备,以最大限度地提高被测放大器的噪声抑制能力。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信