Low polarization pacing lead for detecting the ventricular-evoked response.

Medical progress through technology Pub Date : 1993-01-01
A Bolz, M Hubmann, R Hardt, J Riedmüller, M Schaldach
{"title":"Low polarization pacing lead for detecting the ventricular-evoked response.","authors":"A Bolz,&nbsp;M Hubmann,&nbsp;R Hardt,&nbsp;J Riedmüller,&nbsp;M Schaldach","doi":"","DOIUrl":null,"url":null,"abstract":"<p><p>The cardiac response to a pacing pulse is potentially useful for rate adaptive pacemakers and threshold tracking systems. However, until now capture recognition of the ventricular-evoked response by the use of a single electrode for stimulation as well as detection was limited by the electrode polarization. Electronic measures against the stimulus polarization artifact have not been successful due to the variability of the after potential or the requirement of additional battery power. Following the idea of Lewin, Myers and Parsonnet, who introduced the idea of a non-polarizable porous electrode for physiological stimulation, titanium nitride (TiN) and iridium (Ir) coatings with fractal surface structure have been developed with high electro-chemical active surface areas and Helmholtz double-layer capacities of up to 50,000 microF/cm2, thus reducing the polarization artifact significantly. Two types of endocardial leads (10 with a fractal TiN coating and 5 with a fractal Ir coating) were implanted in the apex of the right ventricle and the polarization artifact, as well as the evoked response, was measured. Both types of pacing leads show a 90% reduction in the polarization artifact in comparison to conventional leads. If an autoshort of approximately 20 to 50 ms is applied after the pacing pulse, the polarization artifact of these leads is negligible, thus enabling reliable detection of at least the repolarization phase of the ventricular-evoked response, which is fully sufficient for capture recognition. Additionally, due to their low polarization losses, TiN- or Ir-coated electrodes with fractal surface structure have a unique stimulation and detection performance.</p>","PeriodicalId":76136,"journal":{"name":"Medical progress through technology","volume":"19 3","pages":"129-37"},"PeriodicalIF":0.0000,"publicationDate":"1993-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Medical progress through technology","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

The cardiac response to a pacing pulse is potentially useful for rate adaptive pacemakers and threshold tracking systems. However, until now capture recognition of the ventricular-evoked response by the use of a single electrode for stimulation as well as detection was limited by the electrode polarization. Electronic measures against the stimulus polarization artifact have not been successful due to the variability of the after potential or the requirement of additional battery power. Following the idea of Lewin, Myers and Parsonnet, who introduced the idea of a non-polarizable porous electrode for physiological stimulation, titanium nitride (TiN) and iridium (Ir) coatings with fractal surface structure have been developed with high electro-chemical active surface areas and Helmholtz double-layer capacities of up to 50,000 microF/cm2, thus reducing the polarization artifact significantly. Two types of endocardial leads (10 with a fractal TiN coating and 5 with a fractal Ir coating) were implanted in the apex of the right ventricle and the polarization artifact, as well as the evoked response, was measured. Both types of pacing leads show a 90% reduction in the polarization artifact in comparison to conventional leads. If an autoshort of approximately 20 to 50 ms is applied after the pacing pulse, the polarization artifact of these leads is negligible, thus enabling reliable detection of at least the repolarization phase of the ventricular-evoked response, which is fully sufficient for capture recognition. Additionally, due to their low polarization losses, TiN- or Ir-coated electrodes with fractal surface structure have a unique stimulation and detection performance.

检测心室诱发反应的低极化起搏导联。
心脏对起搏脉冲的反应对于自适应心率起搏器和阈值跟踪系统具有潜在的应用价值。然而,到目前为止,通过使用单个电极进行刺激和检测来捕捉心室诱发反应的识别受到电极极化的限制。由于后电位的可变性或需要额外的电池功率,针对刺激极化伪影的电子措施尚未成功。Lewin, Myers和Parsonnet提出了一种用于生理刺激的非极化多孔电极的想法,随后,具有分形表面结构的氮化钛(TiN)和铱(Ir)涂层被开发出来,具有高电化学活性表面积和高达50,000 microof /cm2的亥姆霍兹双层容量,从而显着减少了极化误差。将两种类型的心内膜导线(10个分形TiN涂层和5个分形Ir涂层)植入右心室尖部,测量极化伪影和诱发反应。与传统导联相比,两种起搏导联的极化伪影减少了90%。如果在起搏脉冲后施加大约20至50毫秒的自动短路,则这些导联的极化伪像可以忽略不计,因此可以可靠地检测至少心室诱发反应的复极化阶段,这完全足以用于捕获识别。此外,由于具有分形表面结构的TiN或ir涂层电极具有较低的极化损耗,因此具有独特的激发和检测性能。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书