Real-time digital processing of Doppler ultrasound signals and calculation of flow parameters.

Medical progress through technology Pub Date : 1994-01-01
F S Schlindwein, M H Vieira, C F Vasconcelos, D M Simpson
{"title":"Real-time digital processing of Doppler ultrasound signals and calculation of flow parameters.","authors":"F S Schlindwein,&nbsp;M H Vieira,&nbsp;C F Vasconcelos,&nbsp;D M Simpson","doi":"","DOIUrl":null,"url":null,"abstract":"<p><p>Vascular diseases and their complications are responsible for around 27% of deaths in Brazil. Doppler ultrasound is a non-invasive technique that has been used to study blood flow in intact blood vessels since Satomura first reported the potential of the technique in 1959. Because it is non-invasive it makes sequential studies and those in normals feasible. Whereas in contrast angiography only vessel anatomy is displayed, Doppler ultrasound produces dynamic information on blood-flow. It may be used to estimate flow-rates, to image regions of blood flow (colour Doppler), and to help in locating sites of arterial disease, thus complementing X-ray examinations. This paper describes a system based on a Digital Signal Processor for real-time spectrum analysis of Doppler ultrasound signals, real-time display of sonograms, and calculation and analysis of three parameters of clinical interest derived from the Doppler signal. The system comprises a TMS320C25 development board, which acquires the signal and performs spectrum analysis, and a microcomputer, which reads the spectral estimates, displays them as a sonogram in real-time and calculates a set of spectral parameters proposed in the literature. The system permits a maximum sampling frequency of 40.96 kHz, and in the sonogram, 80 power spectra per second (each with 128 frequency bins) are displayed. In a preliminary study, the stability of the haemodynamic parameters and their dependence on a user-defined threshold value is investigated.</p>","PeriodicalId":76136,"journal":{"name":"Medical progress through technology","volume":"20 1-2","pages":"81-9"},"PeriodicalIF":0.0000,"publicationDate":"1994-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Medical progress through technology","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Vascular diseases and their complications are responsible for around 27% of deaths in Brazil. Doppler ultrasound is a non-invasive technique that has been used to study blood flow in intact blood vessels since Satomura first reported the potential of the technique in 1959. Because it is non-invasive it makes sequential studies and those in normals feasible. Whereas in contrast angiography only vessel anatomy is displayed, Doppler ultrasound produces dynamic information on blood-flow. It may be used to estimate flow-rates, to image regions of blood flow (colour Doppler), and to help in locating sites of arterial disease, thus complementing X-ray examinations. This paper describes a system based on a Digital Signal Processor for real-time spectrum analysis of Doppler ultrasound signals, real-time display of sonograms, and calculation and analysis of three parameters of clinical interest derived from the Doppler signal. The system comprises a TMS320C25 development board, which acquires the signal and performs spectrum analysis, and a microcomputer, which reads the spectral estimates, displays them as a sonogram in real-time and calculates a set of spectral parameters proposed in the literature. The system permits a maximum sampling frequency of 40.96 kHz, and in the sonogram, 80 power spectra per second (each with 128 frequency bins) are displayed. In a preliminary study, the stability of the haemodynamic parameters and their dependence on a user-defined threshold value is investigated.

多普勒超声信号的实时数字处理及流量参数的计算。
在巴西,约27%的死亡是由血管疾病及其并发症造成的。多普勒超声是一种非侵入性技术,自1959年Satomura首次报道该技术的潜力以来,已被用于研究完整血管中的血流。因为它是非侵入性的,它使得连续的研究和那些正常的研究是可行的。相比之下,血管造影只显示血管解剖,多普勒超声产生血流的动态信息。它可用于估计血流速率,对血流区域成像(彩色多普勒),并有助于定位动脉疾病的部位,从而补充x线检查。本文介绍了一个基于数字信号处理器的系统,用于多普勒超声信号的实时频谱分析、超声图的实时显示以及由多普勒信号得出的三个临床感兴趣参数的计算和分析。该系统由TMS320C25开发板组成,TMS320C25开发板采集信号并进行频谱分析,微机读取频谱估计值并实时显示为超声图,并计算出一组文献中提出的频谱参数。系统允许的最大采样频率为40.96 kHz,在超声图中,每秒显示80个功率谱(每个功率谱有128个频箱)。在初步研究中,研究了血流动力学参数的稳定性及其对用户定义阈值的依赖性。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书