[Effects of beam weakening and beam hardening on the measured CT values].

Computertomographie Pub Date : 1981-06-01
E Schultz, K Lackner
{"title":"[Effects of beam weakening and beam hardening on the measured CT values].","authors":"E Schultz,&nbsp;K Lackner","doi":"","DOIUrl":null,"url":null,"abstract":"<p><p>The present article reports on the effects of weakening and hardening of radiation on CT values in the environment of highly absorptive materials, using phantoms for carrying out these measurements. First of all, it was found with the help of geometrically simple phantoms (discs, circular rings), that the measured CT value of each volume element depends on the density distribution in the environment and that the CT values of homogeneous object layer are not uniform, but clearly dependent on the location at which measurement is effected. The standard deviation of the CT values of a homogeneous part of an object layer was also found to depend on the environment. Finally, using anatomically relevant phantoms, it was found that density value changes from max. +20 HU to -65 HU occur in the region of the cranium, whereas in the environment of the lumbar vertebra the density varies between max. +20 HU and -10 HU; in the pleural region, the maximum density variations are +/- 10 HU.</p>","PeriodicalId":79210,"journal":{"name":"Computertomographie","volume":"1 2","pages":"83-8"},"PeriodicalIF":0.0000,"publicationDate":"1981-06-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Computertomographie","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

The present article reports on the effects of weakening and hardening of radiation on CT values in the environment of highly absorptive materials, using phantoms for carrying out these measurements. First of all, it was found with the help of geometrically simple phantoms (discs, circular rings), that the measured CT value of each volume element depends on the density distribution in the environment and that the CT values of homogeneous object layer are not uniform, but clearly dependent on the location at which measurement is effected. The standard deviation of the CT values of a homogeneous part of an object layer was also found to depend on the environment. Finally, using anatomically relevant phantoms, it was found that density value changes from max. +20 HU to -65 HU occur in the region of the cranium, whereas in the environment of the lumbar vertebra the density varies between max. +20 HU and -10 HU; in the pleural region, the maximum density variations are +/- 10 HU.

[梁弱化和梁硬化对CT测量值的影响]。
本文报道了在高吸收材料环境中,辐射减弱和硬化对CT值的影响,并利用模型进行了这些测量。首先,借助几何上简单的幻影(圆盘、圆环)发现,每个体积元的测量CT值取决于环境中的密度分布,均匀物层的CT值并不均匀,而是明显依赖于测量的位置。研究还发现,物体层均匀部分CT值的标准差与环境有关。最后,使用解剖学相关的模型,我们发现密度值从最大值变化到最大值。+20 HU至-65 HU发生在头盖骨区域,而在腰椎环境中,密度在max。+20 HU和-10 HU;胸膜区最大密度变化为+/- 10 HU。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信