{"title":"Regional cerebral blood flow by radioxenon-113 inhalation and dynamic emission tomography.","authors":"N A Lassen,&nbsp;L Henriksen,&nbsp;O B Paulson","doi":"","DOIUrl":null,"url":null,"abstract":"<p><p>A rapidly rotating single photon tomograph is described for atraumatic and three-dimensional measurements of rCBF in ml/100 g/min. The instrument has a high sensitivity permitting the recording of the uptake and washout of xenon-133 from the human brain during and after inhalation of the gas at a concentration of 10 mCi/l for 1 min. The algorithm used to calculate rCBF is described, and clinical results in particular in stroke cases are presented. The rapidly rotating tomograph is better suited for the study of focal cerebral ischemia than the conventional stationary detectors because superposition of tissue layers is avoided.</p>","PeriodicalId":76376,"journal":{"name":"Progress in nuclear medicine","volume":"7 ","pages":"110-7"},"PeriodicalIF":0.0000,"publicationDate":"1981-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Progress in nuclear medicine","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

A rapidly rotating single photon tomograph is described for atraumatic and three-dimensional measurements of rCBF in ml/100 g/min. The instrument has a high sensitivity permitting the recording of the uptake and washout of xenon-133 from the human brain during and after inhalation of the gas at a concentration of 10 mCi/l for 1 min. The algorithm used to calculate rCBF is described, and clinical results in particular in stroke cases are presented. The rapidly rotating tomograph is better suited for the study of focal cerebral ischemia than the conventional stationary detectors because superposition of tissue layers is avoided.

放射性氙-113吸入和动态发射断层扫描的局部脑血流。
描述了一种快速旋转的单光子层析成像仪,用于以ml/ 100g /min测量rCBF的自动和三维测量。该仪器具有高灵敏度,可以记录在吸入浓度为10 mCi/l的气体1分钟期间和之后从人脑中摄取和冲洗氙-133的情况。描述了用于计算rCBF的算法,并介绍了特别是在中风病例中的临床结果。由于避免了组织层的叠加,快速旋转层析成像仪比传统的静止检测器更适合于局灶性脑缺血的研究。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信