Use of patient-specific instrumentation for hip resurfacing arthroplasty: Measurement of accuracy using pre and postoperative CT-scan.

IF 2.7 3区 医学 Q2 ORTHOPEDICS
Mehdi Hormi-Menard, Philippe-Alexandre Faure, Martin Bauters, Ramez Shaker, Julien Girard
{"title":"Use of patient-specific instrumentation for hip resurfacing arthroplasty: Measurement of accuracy using pre and postoperative CT-scan.","authors":"Mehdi Hormi-Menard, Philippe-Alexandre Faure, Martin Bauters, Ramez Shaker, Julien Girard","doi":"10.1016/j.otsr.2026.104934","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Hip resurfacing arthroplasty (HRA) is a valuable option for the treatment of primary osteoarthritis in young patients. However, the learning curve for femoral component implantation remains long. Few data are dedicated to patient-specific instrumentation (PSI) in this indication, and merely applied to limited study groups using X-rays as tool to assess accuracy of the method. Therefore, we conducted a prospective study assessing the results of PSI for HRA using pre- and postoperative CT scans based aiming to: 1) assess the accuracy of the planned axis in the axial and coronal planes, 2) evaluate if 3D planning could reproduce the native head diameter by a maximum deviation of one implant size (2 mm) from the component ultimately implanted.</p><p><strong>Hypothesis: </strong>Our hypothesis was that PSI-assisted femoral component implantation would achieve an average accuracy within 5 ° of the planned axis in the axial and coronal planes.</p><p><strong>Materials and methods: </strong>This prospective study analyzed 49 patients (51 HRA) with complete data from an initial cohort of 50 patients. On pre- and post-operative CT scans, the native caput-collum-diaphyseal (CCD) angle, stem-shaft angle (SSA), femoral anteversion and head diameter were measured. The objectives for the PSI jig were to restore head diameter, femoral anteversion and to implant the femoral component with a planned +5 ° valgus alignment in the coronal plane.</p><p><strong>Results: </strong>During the surgery, the calcar-guided PSI (inferior jig) was used in 39 cases (76.4%), and the posterior neck-guided PSI (posterior jig) was used in 12 cases (23.6%). No jig failures were observed. Native femoral anteversion averaged 20 ° (range, 15-24) compared to 18.5 ° (range, 10-27 °) postoperatively (p = 0,3). The mean preoperative CCD was 132.5 ° (range, 126-137). The planned SSA was 138 ° (range, 132-143) while the postoperative SSA was 136.5 ° (range, 129-143) (p = 0,1). With accurate positioning being defined as placement within 5 ° of the planned anteversion and SSA, 48 cases (94.1%) achieved the desired femoral component positioning. The mean acetabular cup and femoral head diameters were respectively 58 mm (range, 54-64) and 52 mm (range, 48-58). The native head diameter averaged 51.5 mm (range, 47-57) (p = 0,2).</p><p><strong>Discussion: </strong>The accuracy of the HRA position was excellent, with more than 94% of cases achieving desired positioning. The maximum difference between the planned and implanted head sizes was 2 mm (1 size), confirming our hypothesis. These findings support the use of this type of PSI for the HRA procedure.</p><p><strong>Level of evidence: </strong>IV; prospective case series.</p>","PeriodicalId":54664,"journal":{"name":"Orthopaedics & Traumatology-Surgery & Research","volume":" ","pages":"104934"},"PeriodicalIF":2.7000,"publicationDate":"2026-10-20","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Orthopaedics & Traumatology-Surgery & Research","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1016/j.otsr.2026.104934","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"ORTHOPEDICS","Score":null,"Total":0}
引用次数: 0

Abstract

Background: Hip resurfacing arthroplasty (HRA) is a valuable option for the treatment of primary osteoarthritis in young patients. However, the learning curve for femoral component implantation remains long. Few data are dedicated to patient-specific instrumentation (PSI) in this indication, and merely applied to limited study groups using X-rays as tool to assess accuracy of the method. Therefore, we conducted a prospective study assessing the results of PSI for HRA using pre- and postoperative CT scans based aiming to: 1) assess the accuracy of the planned axis in the axial and coronal planes, 2) evaluate if 3D planning could reproduce the native head diameter by a maximum deviation of one implant size (2 mm) from the component ultimately implanted.

Hypothesis: Our hypothesis was that PSI-assisted femoral component implantation would achieve an average accuracy within 5 ° of the planned axis in the axial and coronal planes.

Materials and methods: This prospective study analyzed 49 patients (51 HRA) with complete data from an initial cohort of 50 patients. On pre- and post-operative CT scans, the native caput-collum-diaphyseal (CCD) angle, stem-shaft angle (SSA), femoral anteversion and head diameter were measured. The objectives for the PSI jig were to restore head diameter, femoral anteversion and to implant the femoral component with a planned +5 ° valgus alignment in the coronal plane.

Results: During the surgery, the calcar-guided PSI (inferior jig) was used in 39 cases (76.4%), and the posterior neck-guided PSI (posterior jig) was used in 12 cases (23.6%). No jig failures were observed. Native femoral anteversion averaged 20 ° (range, 15-24) compared to 18.5 ° (range, 10-27 °) postoperatively (p = 0,3). The mean preoperative CCD was 132.5 ° (range, 126-137). The planned SSA was 138 ° (range, 132-143) while the postoperative SSA was 136.5 ° (range, 129-143) (p = 0,1). With accurate positioning being defined as placement within 5 ° of the planned anteversion and SSA, 48 cases (94.1%) achieved the desired femoral component positioning. The mean acetabular cup and femoral head diameters were respectively 58 mm (range, 54-64) and 52 mm (range, 48-58). The native head diameter averaged 51.5 mm (range, 47-57) (p = 0,2).

Discussion: The accuracy of the HRA position was excellent, with more than 94% of cases achieving desired positioning. The maximum difference between the planned and implanted head sizes was 2 mm (1 size), confirming our hypothesis. These findings support the use of this type of PSI for the HRA procedure.

Level of evidence: IV; prospective case series.

髋关节置换术中患者专用器械的使用:使用术前和术后ct扫描测量准确性。
背景:髋关节置换术(HRA)是治疗年轻患者原发性骨关节炎的一个有价值的选择。然而,股骨假体植入术的学习曲线仍然很长。很少有数据专门用于该适应症的患者特异性仪器(PSI),并且仅应用于有限的研究组,使用x射线作为评估方法准确性的工具。因此,我们进行了一项前瞻性研究,利用术前和术后CT扫描评估PSI治疗HRA的结果,旨在:1)评估计划轴在轴向和冠状面上的准确性,2)评估3D规划是否可以通过与最终植入的组件最大偏差1个植入物尺寸(2mm)来复制原始头部直径。假设:我们的假设是psi辅助的股骨假体植入术在轴位和冠状面上的计划轴5°内的平均精度。材料和方法:这项前瞻性研究分析了49例患者(51例HRA),初始队列50例患者的完整数据。在术前和术后的CT扫描中,测量了原始的头-柱-骨干(CCD)角、茎-干角(SSA)、股骨前倾角和股骨头直径。PSI夹具的目的是恢复股骨头直径,股骨前倾,并在冠状面计划+5°外翻对准植入股骨假体。结果:术中39例(76.4%)采用骨钙引导下卡索(下卡索),12例(23.6%)采用后颈引导下卡索(后卡索)。没有观察到夹具失效。术后股骨前倾平均为20°(范围15-24),而术后平均为18.5°(范围10-27°)(p = 0,3)。术前平均CCD为132.5°(范围126-137)。计划SSA为138°(范围,132-143),而术后SSA为136.5°(范围,129-143)(p = 0,1)。准确定位被定义为放置在计划前倾角和SSA的5°范围内,48例(94.1%)实现了所需的股骨假体定位。髋臼杯和股骨头的平均直径分别为58 mm(范围54-64)和52 mm(范围48-58)。原生头直径平均为51.5 mm(范围47-57)(p = 0,2)。讨论:HRA定位的准确性非常好,超过94%的病例实现了期望的定位。计划头尺寸与植入头尺寸的最大差异为2mm(1个尺寸),证实了我们的假设。这些发现支持在HRA手术中使用这种类型的PSI。证据等级:四级;前瞻性病例系列。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
5.10
自引率
26.10%
发文量
329
审稿时长
12.5 weeks
期刊介绍: Orthopaedics & Traumatology: Surgery & Research (OTSR) publishes original scientific work in English related to all domains of orthopaedics. Original articles, Reviews, Technical notes and Concise follow-up of a former OTSR study are published in English in electronic form only and indexed in the main international databases.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书