{"title":"Use of patient-specific instrumentation for hip resurfacing arthroplasty: Measurement of accuracy using pre and postoperative CT-scan.","authors":"Mehdi Hormi-Menard, Philippe-Alexandre Faure, Martin Bauters, Ramez Shaker, Julien Girard","doi":"10.1016/j.otsr.2026.104934","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Hip resurfacing arthroplasty (HRA) is a valuable option for the treatment of primary osteoarthritis in young patients. However, the learning curve for femoral component implantation remains long. Few data are dedicated to patient-specific instrumentation (PSI) in this indication, and merely applied to limited study groups using X-rays as tool to assess accuracy of the method. Therefore, we conducted a prospective study assessing the results of PSI for HRA using pre- and postoperative CT scans based aiming to: 1) assess the accuracy of the planned axis in the axial and coronal planes, 2) evaluate if 3D planning could reproduce the native head diameter by a maximum deviation of one implant size (2 mm) from the component ultimately implanted.</p><p><strong>Hypothesis: </strong>Our hypothesis was that PSI-assisted femoral component implantation would achieve an average accuracy within 5 ° of the planned axis in the axial and coronal planes.</p><p><strong>Materials and methods: </strong>This prospective study analyzed 49 patients (51 HRA) with complete data from an initial cohort of 50 patients. On pre- and post-operative CT scans, the native caput-collum-diaphyseal (CCD) angle, stem-shaft angle (SSA), femoral anteversion and head diameter were measured. The objectives for the PSI jig were to restore head diameter, femoral anteversion and to implant the femoral component with a planned +5 ° valgus alignment in the coronal plane.</p><p><strong>Results: </strong>During the surgery, the calcar-guided PSI (inferior jig) was used in 39 cases (76.4%), and the posterior neck-guided PSI (posterior jig) was used in 12 cases (23.6%). No jig failures were observed. Native femoral anteversion averaged 20 ° (range, 15-24) compared to 18.5 ° (range, 10-27 °) postoperatively (p = 0,3). The mean preoperative CCD was 132.5 ° (range, 126-137). The planned SSA was 138 ° (range, 132-143) while the postoperative SSA was 136.5 ° (range, 129-143) (p = 0,1). With accurate positioning being defined as placement within 5 ° of the planned anteversion and SSA, 48 cases (94.1%) achieved the desired femoral component positioning. The mean acetabular cup and femoral head diameters were respectively 58 mm (range, 54-64) and 52 mm (range, 48-58). The native head diameter averaged 51.5 mm (range, 47-57) (p = 0,2).</p><p><strong>Discussion: </strong>The accuracy of the HRA position was excellent, with more than 94% of cases achieving desired positioning. The maximum difference between the planned and implanted head sizes was 2 mm (1 size), confirming our hypothesis. These findings support the use of this type of PSI for the HRA procedure.</p><p><strong>Level of evidence: </strong>IV; prospective case series.</p>","PeriodicalId":54664,"journal":{"name":"Orthopaedics & Traumatology-Surgery & Research","volume":" ","pages":"104934"},"PeriodicalIF":2.7000,"publicationDate":"2026-10-20","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Orthopaedics & Traumatology-Surgery & Research","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1016/j.otsr.2026.104934","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"ORTHOPEDICS","Score":null,"Total":0}
引用次数: 0
Abstract
Background: Hip resurfacing arthroplasty (HRA) is a valuable option for the treatment of primary osteoarthritis in young patients. However, the learning curve for femoral component implantation remains long. Few data are dedicated to patient-specific instrumentation (PSI) in this indication, and merely applied to limited study groups using X-rays as tool to assess accuracy of the method. Therefore, we conducted a prospective study assessing the results of PSI for HRA using pre- and postoperative CT scans based aiming to: 1) assess the accuracy of the planned axis in the axial and coronal planes, 2) evaluate if 3D planning could reproduce the native head diameter by a maximum deviation of one implant size (2 mm) from the component ultimately implanted.
Hypothesis: Our hypothesis was that PSI-assisted femoral component implantation would achieve an average accuracy within 5 ° of the planned axis in the axial and coronal planes.
Materials and methods: This prospective study analyzed 49 patients (51 HRA) with complete data from an initial cohort of 50 patients. On pre- and post-operative CT scans, the native caput-collum-diaphyseal (CCD) angle, stem-shaft angle (SSA), femoral anteversion and head diameter were measured. The objectives for the PSI jig were to restore head diameter, femoral anteversion and to implant the femoral component with a planned +5 ° valgus alignment in the coronal plane.
Results: During the surgery, the calcar-guided PSI (inferior jig) was used in 39 cases (76.4%), and the posterior neck-guided PSI (posterior jig) was used in 12 cases (23.6%). No jig failures were observed. Native femoral anteversion averaged 20 ° (range, 15-24) compared to 18.5 ° (range, 10-27 °) postoperatively (p = 0,3). The mean preoperative CCD was 132.5 ° (range, 126-137). The planned SSA was 138 ° (range, 132-143) while the postoperative SSA was 136.5 ° (range, 129-143) (p = 0,1). With accurate positioning being defined as placement within 5 ° of the planned anteversion and SSA, 48 cases (94.1%) achieved the desired femoral component positioning. The mean acetabular cup and femoral head diameters were respectively 58 mm (range, 54-64) and 52 mm (range, 48-58). The native head diameter averaged 51.5 mm (range, 47-57) (p = 0,2).
Discussion: The accuracy of the HRA position was excellent, with more than 94% of cases achieving desired positioning. The maximum difference between the planned and implanted head sizes was 2 mm (1 size), confirming our hypothesis. These findings support the use of this type of PSI for the HRA procedure.
期刊介绍:
Orthopaedics & Traumatology: Surgery & Research (OTSR) publishes original scientific work in English related to all domains of orthopaedics. Original articles, Reviews, Technical notes and Concise follow-up of a former OTSR study are published in English in electronic form only and indexed in the main international databases.