Diagnostic image quality of shoulder magnetic resonance imaging at 0.55T in healthy subjects and patients with rotator cuff injury: Prospective comparison with 1.5T
IF 2.5 4区 医学Q3 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING
{"title":"Diagnostic image quality of shoulder magnetic resonance imaging at 0.55T in healthy subjects and patients with rotator cuff injury: Prospective comparison with 1.5T","authors":"Jing Liu, Yue Gao, Ge Gao, Wei Li, Xiang Wang, Jinxia Zhu, Xiaoming Liu, Xiaoye Wang, Kun Zhou, Heng Liu, Jianxing Qiu","doi":"10.1002/acm2.70825","DOIUrl":null,"url":null,"abstract":"<div>\n \n \n <section>\n \n <h3> Background</h3>\n \n <p>Lower‑field 0.55 T MRI scanners with large‑bore design could offer patients more comfortable feeling and flexible position. With the advances in deep‑learning reconstruction and hardware optimization, 0.55T have greatly improved their image quality and cost‑effectiveness for musculoskeletal applications. However, its feasibility for shoulder evaluation remains unstudied.</p>\n </section>\n \n <section>\n \n <h3> Purpose</h3>\n \n <p>To evaluate the image quality of shoulder using 0.55T MRI in volunteers and patients with rotator cuff injury with comparison of conventional 1.5T.</p>\n </section>\n \n <section>\n \n <h3> Methods</h3>\n \n <p>Fifty participants (25 volunteers, 25 patients) were prospectively enrolled. All participants underwent shoulder MRI both at open 0.55T with contour coils and deep learning (DL) and 1.5T MR without DL. The nonparametric method was utilized to compare image quality scores and quantitative metrics (signal-to-noise ratio (SNR) and contrast-to-noise ratio (|CNR|)). Arthroscopy have been conducted in 16 patients. Kappa statistics was used to assess interobserver and interstudy agreement for lesion evaluation. Comfort feeling was also compared.</p>\n </section>\n \n <section>\n \n <h3> Results</h3>\n \n <p>The scan time was 11′25″ for 0.55T MRI and 11′42″ for 1.5T MRI. The interobserver analyses showed high consistency (κ = 0.771–1.000, <i>P: 0.008 ∼ </i>< 0.001) among 3 readers. The scores for coronal fat suppressed T2 weighted imaging (fsT2WI) and coronal T1 weighted imaging (T1WI) were significantly higher for 0.55T MRI. Most sequences at 0.55T demonstrated significantly higher SNRs (<i>P:0.04 ∼ </i>< 0.001). Additionally, |CNR| values for lesions were significantly higher in fs proton weighted imaging (PDWI) and fsT2WI at 0.55T. Perfect interstudy agreement between 0.55T and 1.5T was observed between 2 readers (Kappa = 1, <i>P </i>< 0.001) for lesion assessment. Agreement between MRI findings and arthroscopic results was perfect or good. The 0.55T system received higher scores in comfort ratings.</p>\n </section>\n \n <section>\n \n <h3> Conclusion</h3>\n \n <p>The 0.55T MRI with DL, showed comparable or better image quality in healthy subjects and patients with rotator cuff injury to 1.5T MRI without DL.</p>\n </section>\n </div>","PeriodicalId":14989,"journal":{"name":"Journal of Applied Clinical Medical Physics","volume":"27 11","pages":""},"PeriodicalIF":2.5000,"publicationDate":"2026-10-20","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://onlinelibrary.wiley.com/doi/epdf/10.1002/acm2.70825","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Applied Clinical Medical Physics","FirstCategoryId":"3","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1002/acm2.70825","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0
Abstract
Background
Lower‑field 0.55 T MRI scanners with large‑bore design could offer patients more comfortable feeling and flexible position. With the advances in deep‑learning reconstruction and hardware optimization, 0.55T have greatly improved their image quality and cost‑effectiveness for musculoskeletal applications. However, its feasibility for shoulder evaluation remains unstudied.
Purpose
To evaluate the image quality of shoulder using 0.55T MRI in volunteers and patients with rotator cuff injury with comparison of conventional 1.5T.
Methods
Fifty participants (25 volunteers, 25 patients) were prospectively enrolled. All participants underwent shoulder MRI both at open 0.55T with contour coils and deep learning (DL) and 1.5T MR without DL. The nonparametric method was utilized to compare image quality scores and quantitative metrics (signal-to-noise ratio (SNR) and contrast-to-noise ratio (|CNR|)). Arthroscopy have been conducted in 16 patients. Kappa statistics was used to assess interobserver and interstudy agreement for lesion evaluation. Comfort feeling was also compared.
Results
The scan time was 11′25″ for 0.55T MRI and 11′42″ for 1.5T MRI. The interobserver analyses showed high consistency (κ = 0.771–1.000, P: 0.008 ∼ < 0.001) among 3 readers. The scores for coronal fat suppressed T2 weighted imaging (fsT2WI) and coronal T1 weighted imaging (T1WI) were significantly higher for 0.55T MRI. Most sequences at 0.55T demonstrated significantly higher SNRs (P:0.04 ∼ < 0.001). Additionally, |CNR| values for lesions were significantly higher in fs proton weighted imaging (PDWI) and fsT2WI at 0.55T. Perfect interstudy agreement between 0.55T and 1.5T was observed between 2 readers (Kappa = 1, P < 0.001) for lesion assessment. Agreement between MRI findings and arthroscopic results was perfect or good. The 0.55T system received higher scores in comfort ratings.
Conclusion
The 0.55T MRI with DL, showed comparable or better image quality in healthy subjects and patients with rotator cuff injury to 1.5T MRI without DL.
期刊介绍:
Journal of Applied Clinical Medical Physics is an international Open Access publication dedicated to clinical medical physics. JACMP welcomes original contributions dealing with all aspects of medical physics from scientists working in the clinical medical physics around the world. JACMP accepts only online submission.
JACMP will publish:
-Original Contributions: Peer-reviewed, investigations that represent new and significant contributions to the field. Recommended word count: up to 7500.
-Review Articles: Reviews of major areas or sub-areas in the field of clinical medical physics. These articles may be of any length and are peer reviewed.
-Technical Notes: These should be no longer than 3000 words, including key references.
-Letters to the Editor: Comments on papers published in JACMP or on any other matters of interest to clinical medical physics. These should not be more than 1250 (including the literature) and their publication is only based on the decision of the editor, who occasionally asks experts on the merit of the contents.
-Book Reviews: The editorial office solicits Book Reviews.
-Announcements of Forthcoming Meetings: The Editor may provide notice of forthcoming meetings, course offerings, and other events relevant to clinical medical physics.
-Parallel Opposed Editorial: We welcome topics relevant to clinical practice and medical physics profession. The contents can be controversial debate or opposed aspects of an issue. One author argues for the position and the other against. Each side of the debate contains an opening statement up to 800 words, followed by a rebuttal up to 500 words. Readers interested in participating in this series should contact the moderator with a proposed title and a short description of the topic