Nicholas Nelson, Sorour Hosseini, Geoff Nelson, Thomas Martin, Bradley Nordin, Jessica Huang, Hui Zhao
{"title":"Image quality robustness of HyperSight-enabled versus conventional iterative cone-beam CT across patient size and imaging geometry","authors":"Nicholas Nelson, Sorour Hosseini, Geoff Nelson, Thomas Martin, Bradley Nordin, Jessica Huang, Hui Zhao","doi":"10.1002/acm2.70874","DOIUrl":null,"url":null,"abstract":"<div>\n \n \n <section>\n \n <h3> Background</h3>\n \n <p>Recent advances in cone-beam computed tomography (CBCT) have led to the development of next-generation CBCT systems with enhanced image quality and quantitative imaging capabilities. These advances are particularly relevant in the context adaptive radiation therapy, where improved soft tissue visualization and Hounsfield unit (HU) stability are increasingly important for target localization and CBCT-based dose calculation.</p>\n </section>\n \n <section>\n \n <h3> Purpose</h3>\n \n <p>To determine how image quality and HU stability change with increasing phantom size, non-cylindrical geometries, and exposure level, and compare the robustness of Acuros-based HyperSight-enabled iterative CBCT (HS-iCBCT) and conventional iterative CBCT on a TrueBeam (TB-iCBCT) under these conditions.</p>\n </section>\n \n <section>\n \n <h3> Methods</h3>\n \n <p>The CatPhan® 604 phantom, combined with custom annular extensions, was used to simulate increasing patient size and non-cylindrical imaging geometries. Three configurations (no annulus, small annulus, and large annulus) were imaged across a range of exposure levels (mAs) using HS-iCBCT and TB-iCBCT. Image evaluation metrics included noise, contrast-to-noise ratio (CNR), low-contrast visibility (LCV), image uniformity, spatial resolution, and Hounsfield unit (HU) constancy.</p>\n </section>\n \n <section>\n \n <h3> Results</h3>\n \n <p>For the no-annulus configuration, image quality was comparable between systems, with slightly lower noise observed for TB-iCBCT. With increasing phantom size through the added annuli, TB-iCBCT exhibited substantial degradation in noise, contrast, and uniformity, whereas HS-iCBCT maintained more consistent performance. Spatial resolution was primarily dependent on field of view and was similar between systems, with smaller fields of view provided improved spatial resolution. HU constancy degraded with increasing phantom size and material density for both systems but remained substantially more stable for HS-iCBCT.</p>\n </section>\n \n <section>\n \n <h3> Conclusions</h3>\n \n <p>HS-iCBCT demonstrated improved robustness to patient size and imaging geometry, providing superior image quality and HU stability compared to TB-iCBCT. These improvements may support potential lower-dose imaging and more reliable performance in larger patients and non-cylindrical imaging geometries commonly encountered in clinical radiotherapy.</p>\n </section>\n </div>","PeriodicalId":14989,"journal":{"name":"Journal of Applied Clinical Medical Physics","volume":"27 11","pages":""},"PeriodicalIF":2.5000,"publicationDate":"2026-10-20","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://onlinelibrary.wiley.com/doi/epdf/10.1002/acm2.70874","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Applied Clinical Medical Physics","FirstCategoryId":"3","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1002/acm2.70874","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0
Abstract
Background
Recent advances in cone-beam computed tomography (CBCT) have led to the development of next-generation CBCT systems with enhanced image quality and quantitative imaging capabilities. These advances are particularly relevant in the context adaptive radiation therapy, where improved soft tissue visualization and Hounsfield unit (HU) stability are increasingly important for target localization and CBCT-based dose calculation.
Purpose
To determine how image quality and HU stability change with increasing phantom size, non-cylindrical geometries, and exposure level, and compare the robustness of Acuros-based HyperSight-enabled iterative CBCT (HS-iCBCT) and conventional iterative CBCT on a TrueBeam (TB-iCBCT) under these conditions.
Methods
The CatPhan® 604 phantom, combined with custom annular extensions, was used to simulate increasing patient size and non-cylindrical imaging geometries. Three configurations (no annulus, small annulus, and large annulus) were imaged across a range of exposure levels (mAs) using HS-iCBCT and TB-iCBCT. Image evaluation metrics included noise, contrast-to-noise ratio (CNR), low-contrast visibility (LCV), image uniformity, spatial resolution, and Hounsfield unit (HU) constancy.
Results
For the no-annulus configuration, image quality was comparable between systems, with slightly lower noise observed for TB-iCBCT. With increasing phantom size through the added annuli, TB-iCBCT exhibited substantial degradation in noise, contrast, and uniformity, whereas HS-iCBCT maintained more consistent performance. Spatial resolution was primarily dependent on field of view and was similar between systems, with smaller fields of view provided improved spatial resolution. HU constancy degraded with increasing phantom size and material density for both systems but remained substantially more stable for HS-iCBCT.
Conclusions
HS-iCBCT demonstrated improved robustness to patient size and imaging geometry, providing superior image quality and HU stability compared to TB-iCBCT. These improvements may support potential lower-dose imaging and more reliable performance in larger patients and non-cylindrical imaging geometries commonly encountered in clinical radiotherapy.
期刊介绍:
Journal of Applied Clinical Medical Physics is an international Open Access publication dedicated to clinical medical physics. JACMP welcomes original contributions dealing with all aspects of medical physics from scientists working in the clinical medical physics around the world. JACMP accepts only online submission.
JACMP will publish:
-Original Contributions: Peer-reviewed, investigations that represent new and significant contributions to the field. Recommended word count: up to 7500.
-Review Articles: Reviews of major areas or sub-areas in the field of clinical medical physics. These articles may be of any length and are peer reviewed.
-Technical Notes: These should be no longer than 3000 words, including key references.
-Letters to the Editor: Comments on papers published in JACMP or on any other matters of interest to clinical medical physics. These should not be more than 1250 (including the literature) and their publication is only based on the decision of the editor, who occasionally asks experts on the merit of the contents.
-Book Reviews: The editorial office solicits Book Reviews.
-Announcements of Forthcoming Meetings: The Editor may provide notice of forthcoming meetings, course offerings, and other events relevant to clinical medical physics.
-Parallel Opposed Editorial: We welcome topics relevant to clinical practice and medical physics profession. The contents can be controversial debate or opposed aspects of an issue. One author argues for the position and the other against. Each side of the debate contains an opening statement up to 800 words, followed by a rebuttal up to 500 words. Readers interested in participating in this series should contact the moderator with a proposed title and a short description of the topic