Image quality robustness of HyperSight-enabled versus conventional iterative cone-beam CT across patient size and imaging geometry

IF 2.5 4区 医学 Q3 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING
Nicholas Nelson, Sorour Hosseini, Geoff Nelson, Thomas Martin, Bradley Nordin, Jessica Huang, Hui Zhao
{"title":"Image quality robustness of HyperSight-enabled versus conventional iterative cone-beam CT across patient size and imaging geometry","authors":"Nicholas Nelson,&nbsp;Sorour Hosseini,&nbsp;Geoff Nelson,&nbsp;Thomas Martin,&nbsp;Bradley Nordin,&nbsp;Jessica Huang,&nbsp;Hui Zhao","doi":"10.1002/acm2.70874","DOIUrl":null,"url":null,"abstract":"<div>\n \n \n <section>\n \n <h3> Background</h3>\n \n <p>Recent advances in cone-beam computed tomography (CBCT) have led to the development of next-generation CBCT systems with enhanced image quality and quantitative imaging capabilities. These advances are particularly relevant in the context adaptive radiation therapy, where improved soft tissue visualization and Hounsfield unit (HU) stability are increasingly important for target localization and CBCT-based dose calculation.</p>\n </section>\n \n <section>\n \n <h3> Purpose</h3>\n \n <p>To determine how image quality and HU stability change with increasing phantom size, non-cylindrical geometries, and exposure level, and compare the robustness of Acuros-based HyperSight-enabled iterative CBCT (HS-iCBCT) and conventional iterative CBCT on a TrueBeam (TB-iCBCT) under these conditions.</p>\n </section>\n \n <section>\n \n <h3> Methods</h3>\n \n <p>The CatPhan® 604 phantom, combined with custom annular extensions, was used to simulate increasing patient size and non-cylindrical imaging geometries. Three configurations (no annulus, small annulus, and large annulus) were imaged across a range of exposure levels (mAs) using HS-iCBCT and TB-iCBCT. Image evaluation metrics included noise, contrast-to-noise ratio (CNR), low-contrast visibility (LCV), image uniformity, spatial resolution, and Hounsfield unit (HU) constancy.</p>\n </section>\n \n <section>\n \n <h3> Results</h3>\n \n <p>For the no-annulus configuration, image quality was comparable between systems, with slightly lower noise observed for TB-iCBCT. With increasing phantom size through the added annuli, TB-iCBCT exhibited substantial degradation in noise, contrast, and uniformity, whereas HS-iCBCT maintained more consistent performance. Spatial resolution was primarily dependent on field of view and was similar between systems, with smaller fields of view provided improved spatial resolution. HU constancy degraded with increasing phantom size and material density for both systems but remained substantially more stable for HS-iCBCT.</p>\n </section>\n \n <section>\n \n <h3> Conclusions</h3>\n \n <p>HS-iCBCT demonstrated improved robustness to patient size and imaging geometry, providing superior image quality and HU stability compared to TB-iCBCT. These improvements may support potential lower-dose imaging and more reliable performance in larger patients and non-cylindrical imaging geometries commonly encountered in clinical radiotherapy.</p>\n </section>\n </div>","PeriodicalId":14989,"journal":{"name":"Journal of Applied Clinical Medical Physics","volume":"27 11","pages":""},"PeriodicalIF":2.5000,"publicationDate":"2026-10-20","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://onlinelibrary.wiley.com/doi/epdf/10.1002/acm2.70874","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Applied Clinical Medical Physics","FirstCategoryId":"3","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1002/acm2.70874","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

Abstract

Background

Recent advances in cone-beam computed tomography (CBCT) have led to the development of next-generation CBCT systems with enhanced image quality and quantitative imaging capabilities. These advances are particularly relevant in the context adaptive radiation therapy, where improved soft tissue visualization and Hounsfield unit (HU) stability are increasingly important for target localization and CBCT-based dose calculation.

Purpose

To determine how image quality and HU stability change with increasing phantom size, non-cylindrical geometries, and exposure level, and compare the robustness of Acuros-based HyperSight-enabled iterative CBCT (HS-iCBCT) and conventional iterative CBCT on a TrueBeam (TB-iCBCT) under these conditions.

Methods

The CatPhan® 604 phantom, combined with custom annular extensions, was used to simulate increasing patient size and non-cylindrical imaging geometries. Three configurations (no annulus, small annulus, and large annulus) were imaged across a range of exposure levels (mAs) using HS-iCBCT and TB-iCBCT. Image evaluation metrics included noise, contrast-to-noise ratio (CNR), low-contrast visibility (LCV), image uniformity, spatial resolution, and Hounsfield unit (HU) constancy.

Results

For the no-annulus configuration, image quality was comparable between systems, with slightly lower noise observed for TB-iCBCT. With increasing phantom size through the added annuli, TB-iCBCT exhibited substantial degradation in noise, contrast, and uniformity, whereas HS-iCBCT maintained more consistent performance. Spatial resolution was primarily dependent on field of view and was similar between systems, with smaller fields of view provided improved spatial resolution. HU constancy degraded with increasing phantom size and material density for both systems but remained substantially more stable for HS-iCBCT.

Conclusions

HS-iCBCT demonstrated improved robustness to patient size and imaging geometry, providing superior image quality and HU stability compared to TB-iCBCT. These improvements may support potential lower-dose imaging and more reliable performance in larger patients and non-cylindrical imaging geometries commonly encountered in clinical radiotherapy.

与传统的迭代锥束CT相比,hypersight支持的图像质量鲁棒性跨越了患者尺寸和成像几何形状
锥形束计算机断层扫描(CBCT)的最新进展导致了下一代CBCT系统的发展,该系统具有增强的图像质量和定量成像能力。这些进展与适应性放射治疗尤其相关,其中改善的软组织可视化和Hounsfield单位(HU)稳定性对于目标定位和基于cbct的剂量计算越来越重要。目的确定图像质量和HU稳定性如何随着幻影尺寸、非圆柱形几何形状和曝光水平的增加而变化,并比较基于acuros的HyperSight-enabled迭代CBCT (HS-iCBCT)和基于TrueBeam的常规迭代CBCT (TB-iCBCT)在这些条件下的鲁棒性。方法使用CatPhan®604假体,结合定制的环形扩展,模拟增加的患者尺寸和非圆柱形成像几何形状。使用HS-iCBCT和TB-iCBCT在暴露水平(mAs)范围内对三种构型(无环空、小环空和大环空)进行成像。图像评价指标包括噪声、噪声对比比(CNR)、低对比度可见度(LCV)、图像均匀性、空间分辨率和霍斯菲尔德单位(HU)稳定性。结果对于无环隙配置,两种系统的图像质量相当,TB-iCBCT观察到的噪声略低。通过增加环空增加幻相尺寸,TB-iCBCT在噪声、对比度和均匀性方面表现出明显的下降,而HS-iCBCT则保持了更稳定的性能。空间分辨率主要依赖于视场,系统之间相似,较小的视场提供更好的空间分辨率。在这两种系统中,HU常数都随着模体尺寸和材料密度的增加而下降,但HS-iCBCT的HU常数基本保持稳定。与TB-iCBCT相比,HS-iCBCT对患者尺寸和成像几何形状的稳健性更好,提供了更好的图像质量和HU稳定性。这些改进可能支持潜在的低剂量成像和更可靠的表现,在较大的患者和临床放疗中常见的非圆柱形成像几何。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
3.60
自引率
19.00%
发文量
331
审稿时长
3 months
期刊介绍: Journal of Applied Clinical Medical Physics is an international Open Access publication dedicated to clinical medical physics. JACMP welcomes original contributions dealing with all aspects of medical physics from scientists working in the clinical medical physics around the world. JACMP accepts only online submission. JACMP will publish: -Original Contributions: Peer-reviewed, investigations that represent new and significant contributions to the field. Recommended word count: up to 7500. -Review Articles: Reviews of major areas or sub-areas in the field of clinical medical physics. These articles may be of any length and are peer reviewed. -Technical Notes: These should be no longer than 3000 words, including key references. -Letters to the Editor: Comments on papers published in JACMP or on any other matters of interest to clinical medical physics. These should not be more than 1250 (including the literature) and their publication is only based on the decision of the editor, who occasionally asks experts on the merit of the contents. -Book Reviews: The editorial office solicits Book Reviews. -Announcements of Forthcoming Meetings: The Editor may provide notice of forthcoming meetings, course offerings, and other events relevant to clinical medical physics. -Parallel Opposed Editorial: We welcome topics relevant to clinical practice and medical physics profession. The contents can be controversial debate or opposed aspects of an issue. One author argues for the position and the other against. Each side of the debate contains an opening statement up to 800 words, followed by a rebuttal up to 500 words. Readers interested in participating in this series should contact the moderator with a proposed title and a short description of the topic
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书