Geometric and dosimetric effects of an automated clustering and plan optimization pipeline for multiple brain metastasis radiosurgery

IF 2.5 4区 医学 Q3 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING
Adam D. Yock, Guozhen Luo, Sean All, Austin N. Kirschner, Cornelis Kamerling
{"title":"Geometric and dosimetric effects of an automated clustering and plan optimization pipeline for multiple brain metastasis radiosurgery","authors":"Adam D. Yock,&nbsp;Guozhen Luo,&nbsp;Sean All,&nbsp;Austin N. Kirschner,&nbsp;Cornelis Kamerling","doi":"10.1002/acm2.70869","DOIUrl":null,"url":null,"abstract":"<div>\n \n \n <section>\n \n <h3> Background</h3>\n \n <p>Treating multiple brain metastases simultaneously with single-isocenter multi-target (SIMT) stereotactic radiosurgery (SRS) can streamline treatment planning and delivery. However, it can also result in targets at large distances from the isocenter, increasing sensitivity to rotational setup errors and requiring wider collimator leaves. Grouping the targets and treating each group with its own plan retains the practical benefits of SIMT SRS while mitigating the concerns of targets at large distances off-isocenter.</p>\n </section>\n \n <section>\n \n <h3> Purpose</h3>\n \n <p>This work presents a prototype of a newly integrated treatment planning pipeline that automatically clusters targets into groups and subsequently optimizes a plan for each group simultaneously. Treatment scenarios featuring varying numbers of isocenters were compared.</p>\n </section>\n \n <section>\n \n <h3> Methods</h3>\n \n <p>Twenty-nine treatment courses for patients receiving SRS for 3–15 brain metastases were retrospectively analyzed. Multiple treatment scenarios were prepared for each patient-course by grouping targets into 1–5 isocenters using k-means clustering. PTV margins were added using two methods, one with a magnitude that varied with distance to the isocenter, and a second with a constant 1-mm magnitude. A non-clinical, command-line version of the Brainlab Elements Multiple Brain Mets SRS software was used to optimize plans within each scenario simultaneously. Cumulative doses were analyzed in terms of geometric values (target distances to isocenter, total PTV volume), dosimetric values (volume of normal brain (brain-GTVs) receiving 5, 8, 10, 12, and 18 Gy (V<sub>5Gy</sub>, V<sub>8Gy</sub>, V<sub>10Gy</sub>, V<sub>12Gy</sub>, and V<sub>18Gy</sub>)), and logistical values (number of Monitor Units (MUs)). A representative sample of 69% of plans with variable margins was reviewed by clinical SRS experts.</p>\n </section>\n \n <section>\n \n <h3> Results</h3>\n \n <p>As the number of isocenters increased, distances from the targets to the isocenters decreased, and the proportion of targets within a given distance increased. For example, 36%, 77%, 94%, 99%, and 100% of targets were within 4 cm of the isocenter for 1–5 isocenters, respectively. For the variable margin method, the total PTV volume relative to the single-isocenter scenario had a median decrease of 17.5, 24.3, 31.6, and 37.8% for 2–5 isocenters, respectively. Similarly, the relative change in the normal brain dose metrics decreased with increasing number of isocenters (e.g. V<sub>12Gy</sub> decreased 20.1, 26.2, 36.2, and 40.9%, respectively). The relative change in dose metrics at each level was found to have a linear relationship with that of the total PTV volume (e.g. V<sub>12Gy</sub>, <i>R</i><sup>2</sup> = 0.81). For the constant margin method, total PTV volume remained constant, but normal brain dose metrics still decreased due to the grouping and simultaneous plan optimization alone (e.g. V<sub>12Gy</sub> decreased 6.4, 10.5, 15.1, and 17.1%). Regarding logistical values, the number of MUs increased with each additional isocenter for both margin methods, but at a rate slower than proportional to the number of isocenters. Of the evaluated plans, nearly all (97.7%) were clinically acceptable without modification.</p>\n </section>\n \n <section>\n \n <h3> Conclusion</h3>\n \n <p>The SIMT SRS treatment planning pipeline presented here features automatic grouping of targets followed by automatic dose optimization of up to five isocenters. By quantifying the geometric, dosimetric, and logistical effects of adding isocenters, this pipeline can empower clinicians to make resource-conscious treatment decisions optimal for each patient.</p>\n </section>\n </div>","PeriodicalId":14989,"journal":{"name":"Journal of Applied Clinical Medical Physics","volume":"27 11","pages":""},"PeriodicalIF":2.5000,"publicationDate":"2026-10-20","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://onlinelibrary.wiley.com/doi/epdf/10.1002/acm2.70869","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Applied Clinical Medical Physics","FirstCategoryId":"3","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1002/acm2.70869","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

Abstract

Background

Treating multiple brain metastases simultaneously with single-isocenter multi-target (SIMT) stereotactic radiosurgery (SRS) can streamline treatment planning and delivery. However, it can also result in targets at large distances from the isocenter, increasing sensitivity to rotational setup errors and requiring wider collimator leaves. Grouping the targets and treating each group with its own plan retains the practical benefits of SIMT SRS while mitigating the concerns of targets at large distances off-isocenter.

Purpose

This work presents a prototype of a newly integrated treatment planning pipeline that automatically clusters targets into groups and subsequently optimizes a plan for each group simultaneously. Treatment scenarios featuring varying numbers of isocenters were compared.

Methods

Twenty-nine treatment courses for patients receiving SRS for 3–15 brain metastases were retrospectively analyzed. Multiple treatment scenarios were prepared for each patient-course by grouping targets into 1–5 isocenters using k-means clustering. PTV margins were added using two methods, one with a magnitude that varied with distance to the isocenter, and a second with a constant 1-mm magnitude. A non-clinical, command-line version of the Brainlab Elements Multiple Brain Mets SRS software was used to optimize plans within each scenario simultaneously. Cumulative doses were analyzed in terms of geometric values (target distances to isocenter, total PTV volume), dosimetric values (volume of normal brain (brain-GTVs) receiving 5, 8, 10, 12, and 18 Gy (V5Gy, V8Gy, V10Gy, V12Gy, and V18Gy)), and logistical values (number of Monitor Units (MUs)). A representative sample of 69% of plans with variable margins was reviewed by clinical SRS experts.

Results

As the number of isocenters increased, distances from the targets to the isocenters decreased, and the proportion of targets within a given distance increased. For example, 36%, 77%, 94%, 99%, and 100% of targets were within 4 cm of the isocenter for 1–5 isocenters, respectively. For the variable margin method, the total PTV volume relative to the single-isocenter scenario had a median decrease of 17.5, 24.3, 31.6, and 37.8% for 2–5 isocenters, respectively. Similarly, the relative change in the normal brain dose metrics decreased with increasing number of isocenters (e.g. V12Gy decreased 20.1, 26.2, 36.2, and 40.9%, respectively). The relative change in dose metrics at each level was found to have a linear relationship with that of the total PTV volume (e.g. V12Gy, R2 = 0.81). For the constant margin method, total PTV volume remained constant, but normal brain dose metrics still decreased due to the grouping and simultaneous plan optimization alone (e.g. V12Gy decreased 6.4, 10.5, 15.1, and 17.1%). Regarding logistical values, the number of MUs increased with each additional isocenter for both margin methods, but at a rate slower than proportional to the number of isocenters. Of the evaluated plans, nearly all (97.7%) were clinically acceptable without modification.

Conclusion

The SIMT SRS treatment planning pipeline presented here features automatic grouping of targets followed by automatic dose optimization of up to five isocenters. By quantifying the geometric, dosimetric, and logistical effects of adding isocenters, this pipeline can empower clinicians to make resource-conscious treatment decisions optimal for each patient.

多脑转移放射手术中自动聚类和计划优化管道的几何和剂量效应
背景单等中心多靶点(SIMT)立体定向放射外科(SRS)同时治疗多发性脑转移瘤可以简化治疗计划和递送。然而,它也可能导致目标距离等心很远,增加对旋转设置误差的敏感性,并且需要更宽的准直器叶片。将目标分组并按照自己的计划处理每组目标,既保留了SIMT SRS的实际优势,又减轻了远距离非等心目标的担忧。本工作提出了一种新型集成治疗计划管道的原型,该管道可以自动将目标聚类成组,并随后同时优化每个组的计划。比较了具有不同数量等中心的治疗方案。方法回顾性分析3 ~ 15例脑转移瘤患者接受SRS治疗的29个疗程。采用k-means聚类方法将靶点分为1-5个等中心,为每个患者病程准备多个治疗方案。使用两种方法添加PTV边缘,一种方法的幅度随距离等心的远近而变化,另一种方法的幅度恒定为1 mm。Brainlab Elements Multiple Brain Mets SRS软件的非临床命令行版本用于同时优化每种情况下的计划。根据几何值(目标到等中心的距离,总PTV体积),剂量学值(接受5、8、10、12和18Gy (V5Gy、V8Gy、V10Gy、V12Gy和V18Gy)的正常脑(脑- gtv)体积)和后勤值(监测单位数(MUs))分析累积剂量。临床SRS专家对69%可变边际计划的代表性样本进行了审查。结果随着等距中心数目的增加,目标到等距中心的距离减小,目标在一定距离内的比例增大。例如,对于1-5个等中心,分别有36%、77%、94%、99%和100%的目标在距等中心4 cm范围内。对于可变边际方法,相对于单等中心方案,2-5个等中心方案的总PTV体积的中位数分别减少了17.5%、24.3、31.6和37.8%。同样,正常脑剂量指标的相对变化随着等中心数量的增加而下降(例如V12Gy分别下降20.1%、26.2%、36.2%和40.9%)。各水平剂量指标的相对变化与PTV总容积的相对变化呈线性关系(如V12Gy, R2 = 0.81)。对于恒裕度法,总PTV体积保持不变,但正常脑剂量指标仍因分组和同时计划优化而下降(例如V12Gy下降6.4、10.5、15.1%和17.1%)。关于后勤价值,对于两种边际方法,每增加一个等中心,最小值的数量就会增加,但速度低于与等中心数量成比例的速度。在评估的方案中,几乎所有(97.7%)都是临床可接受的,无需修改。结论本文提出的SIMT SRS治疗计划管道具有自动分组靶点以及多达5个等中心的自动剂量优化的特点。通过量化几何、剂量学和增加等中心的后勤影响,该管道可以使临床医生能够为每个患者做出具有资源意识的最佳治疗决策。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
3.60
自引率
19.00%
发文量
331
审稿时长
3 months
期刊介绍: Journal of Applied Clinical Medical Physics is an international Open Access publication dedicated to clinical medical physics. JACMP welcomes original contributions dealing with all aspects of medical physics from scientists working in the clinical medical physics around the world. JACMP accepts only online submission. JACMP will publish: -Original Contributions: Peer-reviewed, investigations that represent new and significant contributions to the field. Recommended word count: up to 7500. -Review Articles: Reviews of major areas or sub-areas in the field of clinical medical physics. These articles may be of any length and are peer reviewed. -Technical Notes: These should be no longer than 3000 words, including key references. -Letters to the Editor: Comments on papers published in JACMP or on any other matters of interest to clinical medical physics. These should not be more than 1250 (including the literature) and their publication is only based on the decision of the editor, who occasionally asks experts on the merit of the contents. -Book Reviews: The editorial office solicits Book Reviews. -Announcements of Forthcoming Meetings: The Editor may provide notice of forthcoming meetings, course offerings, and other events relevant to clinical medical physics. -Parallel Opposed Editorial: We welcome topics relevant to clinical practice and medical physics profession. The contents can be controversial debate or opposed aspects of an issue. One author argues for the position and the other against. Each side of the debate contains an opening statement up to 800 words, followed by a rebuttal up to 500 words. Readers interested in participating in this series should contact the moderator with a proposed title and a short description of the topic
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书