Development and validation of a pregnant phantom with adjustable fetal presentation for region-specific fetal radiation dose assessment during pelvic radiography

IF 2.5 4区 医学 Q3 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING
Yuta Matsunaga, Yumi Kataoka, Koichi Chida, Yasuki Asada
{"title":"Development and validation of a pregnant phantom with adjustable fetal presentation for region-specific fetal radiation dose assessment during pelvic radiography","authors":"Yuta Matsunaga,&nbsp;Yumi Kataoka,&nbsp;Koichi Chida,&nbsp;Yasuki Asada","doi":"10.1002/acm2.70859","DOIUrl":null,"url":null,"abstract":"<div>\n \n \n <section>\n \n <h3> Background</h3>\n \n <p>Accurate assessment of fetal radiation dose during diagnostic X-ray examinations remains challenging because commercially available pregnancy phantoms do not adequately reproduce fetal anatomy or permit organ-specific dose measurements.</p>\n </section>\n \n <section>\n \n <h3> Purpose</h3>\n \n <p>The purpose of this study was to develop a pregnant phantom with adjustable fetal presentation and to evaluate its anatomical realism and applicability for fetal dose assessment during pelvic radiography.</p>\n </section>\n \n <section>\n \n <h3> Methods</h3>\n \n <p>The developed phantom comprised an anthropomorphic phantom, a whole-body fetal phantom, acrylic components, and water-filled plastic bags. The fetal phantom was fabricated using three-dimensional printing technology and incorporated bone-equivalent and soft-tissue-equivalent materials. Fetal presentations could be adjusted to cephalic or breech presentations by repositioning the fetal phantom within the abdominal phantom. Computed tomography (CT) images were acquired to measure CT numbers of the fetal brain, thorax, abdomen, pelvic bone, and surrounding water-filled regions. Absorbed doses were measured using real-time dosimeters placed at five locations in the fetal head and three locations in the fetal abdomen. Dose measurements were performed during pelvic radiography and compared with those obtained using a conventional homogeneous pregnancy phantom.</p>\n </section>\n \n <section>\n \n <h3> Results</h3>\n \n <p>The mean CT numbers of the fetal brain, thorax, abdomen, pelvic bone, and water-filled regions were 2.1, -0.8, -0.4, 323.8, and 1.6 HU, respectively. These values were within the expected ranges for soft tissue and bone and were comparable to those of a conventional homogeneous phantom. During pelvic radiography, the mean absorbed doses measured using the adjustable fetal-presentation phantom were 1.34 mGy in the fetal head and 0.07 mGy in the fetal abdomen for the cephalic presentation, and 0.05 mGy and 1.67 mGy, respectively, for the breech presentation. In comparison, the mean absorbed dose measured in the fetal region of the conventional homogeneous phantom was 2.05 mGy. The findings demonstrated substantial differences in regional fetal dose distribution according to fetal presentation. In particular, the fetal head dose in the cephalic presentation was lower than that measured using the conventional phantom. This difference may be partly attributable to attenuation by the fetal skull; however, the comparison should be interpreted with caution given differences in phantom geometry, internal materials, and dosimeter locations between the two phantoms.</p>\n </section>\n \n <section>\n \n <h3> Conclusion</h3>\n \n <p>A novel pregnant phantom with adjustable fetal presentation was successfully developed. The phantom reproduced anatomically relevant structures using realistic bone-equivalent and soft-tissue-equivalent materials and enabled direct dose measurements in specific fetal anatomical regions. Compared with a conventional homogeneous phantom, the developed phantom provided a more detailed assessment of fetal dose distributions and allowed evaluation of the effects of fetal presentation on radiation exposure during pelvic radiography.</p>\n </section>\n </div>","PeriodicalId":14989,"journal":{"name":"Journal of Applied Clinical Medical Physics","volume":"27 10","pages":""},"PeriodicalIF":2.5000,"publicationDate":"2026-10-14","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://onlinelibrary.wiley.com/doi/epdf/10.1002/acm2.70859","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Applied Clinical Medical Physics","FirstCategoryId":"3","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1002/acm2.70859","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

Abstract

Background

Accurate assessment of fetal radiation dose during diagnostic X-ray examinations remains challenging because commercially available pregnancy phantoms do not adequately reproduce fetal anatomy or permit organ-specific dose measurements.

Purpose

The purpose of this study was to develop a pregnant phantom with adjustable fetal presentation and to evaluate its anatomical realism and applicability for fetal dose assessment during pelvic radiography.

Methods

The developed phantom comprised an anthropomorphic phantom, a whole-body fetal phantom, acrylic components, and water-filled plastic bags. The fetal phantom was fabricated using three-dimensional printing technology and incorporated bone-equivalent and soft-tissue-equivalent materials. Fetal presentations could be adjusted to cephalic or breech presentations by repositioning the fetal phantom within the abdominal phantom. Computed tomography (CT) images were acquired to measure CT numbers of the fetal brain, thorax, abdomen, pelvic bone, and surrounding water-filled regions. Absorbed doses were measured using real-time dosimeters placed at five locations in the fetal head and three locations in the fetal abdomen. Dose measurements were performed during pelvic radiography and compared with those obtained using a conventional homogeneous pregnancy phantom.

Results

The mean CT numbers of the fetal brain, thorax, abdomen, pelvic bone, and water-filled regions were 2.1, -0.8, -0.4, 323.8, and 1.6 HU, respectively. These values were within the expected ranges for soft tissue and bone and were comparable to those of a conventional homogeneous phantom. During pelvic radiography, the mean absorbed doses measured using the adjustable fetal-presentation phantom were 1.34 mGy in the fetal head and 0.07 mGy in the fetal abdomen for the cephalic presentation, and 0.05 mGy and 1.67 mGy, respectively, for the breech presentation. In comparison, the mean absorbed dose measured in the fetal region of the conventional homogeneous phantom was 2.05 mGy. The findings demonstrated substantial differences in regional fetal dose distribution according to fetal presentation. In particular, the fetal head dose in the cephalic presentation was lower than that measured using the conventional phantom. This difference may be partly attributable to attenuation by the fetal skull; however, the comparison should be interpreted with caution given differences in phantom geometry, internal materials, and dosimeter locations between the two phantoms.

Conclusion

A novel pregnant phantom with adjustable fetal presentation was successfully developed. The phantom reproduced anatomically relevant structures using realistic bone-equivalent and soft-tissue-equivalent materials and enabled direct dose measurements in specific fetal anatomical regions. Compared with a conventional homogeneous phantom, the developed phantom provided a more detailed assessment of fetal dose distributions and allowed evaluation of the effects of fetal presentation on radiation exposure during pelvic radiography.

开发和验证妊娠幻影与可调节胎儿表现胎儿放射剂量评估在骨盆造影期间。
背景:在诊断x线检查中准确评估胎儿辐射剂量仍然具有挑战性,因为市售的妊娠幻象不能充分再现胎儿解剖结构或允许器官特异性剂量测量。目的:本研究的目的是开发一个可调节胎儿体位的妊娠假体,并评估其解剖学真实性和在骨盆x线摄影中胎儿剂量评估的适用性。方法:采用拟人型假体、胎儿全身假体、亚克力构件和充水塑料袋组成假体。采用三维打印技术,结合骨等效材料和软组织等效材料制备胎儿体模。通过将胎儿幻像重新定位在腹部幻像内,胎儿的呈现可以调整为头位或臀位。计算机断层扫描(CT)图像测量胎儿的大脑,胸部,腹部,骨盆骨和周围充满水的区域的CT数。在胎儿头部的五个位置和腹部的三个位置使用实时剂量计测量吸收剂量。剂量测量是在盆腔造影期间进行的,并与使用传统均匀妊娠假体获得的剂量进行比较。结果:胎儿脑、胸、腹、骨盆骨、积水区CT平均值分别为2.1、-0.8、-0.4、323.8、1.6 HU。这些值在软组织和骨骼的预期范围内,与传统的均质假体相当。在骨盆x线摄影中,使用可调节胎儿呈现幻象测量的平均吸收剂量在胎儿头位为1.34 mGy,在胎儿腹部为0.07 mGy,在胎儿臀位为0.05 mGy和1.67 mGy。相比之下,在胎区测量的平均吸收剂量为2.05 mGy。研究结果表明,根据胎儿的表现,胎儿剂量的区域分布存在实质性差异。特别是,在头位表现中,胎头剂量低于使用传统幻象测量的剂量。这种差异可能部分归因于胎儿颅骨的衰减;然而,考虑到两种幽灵的几何形状、内部材料和剂量计位置的差异,应该谨慎解释比较。结论:成功研制了一种可调节胎位的新型妊娠假体。该模型使用真实的骨等效和软组织等效材料复制了解剖学上相关的结构,并在特定的胎儿解剖区域进行了直接剂量测量。与传统的均质假体相比,开发的假体提供了更详细的胎儿剂量分布评估,并允许评估胎儿在骨盆造影期间对辐射暴露的影响。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
3.60
自引率
19.00%
发文量
331
审稿时长
3 months
期刊介绍: Journal of Applied Clinical Medical Physics is an international Open Access publication dedicated to clinical medical physics. JACMP welcomes original contributions dealing with all aspects of medical physics from scientists working in the clinical medical physics around the world. JACMP accepts only online submission. JACMP will publish: -Original Contributions: Peer-reviewed, investigations that represent new and significant contributions to the field. Recommended word count: up to 7500. -Review Articles: Reviews of major areas or sub-areas in the field of clinical medical physics. These articles may be of any length and are peer reviewed. -Technical Notes: These should be no longer than 3000 words, including key references. -Letters to the Editor: Comments on papers published in JACMP or on any other matters of interest to clinical medical physics. These should not be more than 1250 (including the literature) and their publication is only based on the decision of the editor, who occasionally asks experts on the merit of the contents. -Book Reviews: The editorial office solicits Book Reviews. -Announcements of Forthcoming Meetings: The Editor may provide notice of forthcoming meetings, course offerings, and other events relevant to clinical medical physics. -Parallel Opposed Editorial: We welcome topics relevant to clinical practice and medical physics profession. The contents can be controversial debate or opposed aspects of an issue. One author argues for the position and the other against. Each side of the debate contains an opening statement up to 800 words, followed by a rebuttal up to 500 words. Readers interested in participating in this series should contact the moderator with a proposed title and a short description of the topic
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书