Automated linac downtime tracking across 4 radiation oncology clinics within a single hospital system utilizing statistical process control

IF 2.5 4区 医学 Q3 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING
Caroline E. Serafini, Matthew Thomas, Xiaofeng Yang, Pretesh Patel, Justin Roper, Maxine Washington, Anees Dhabaan, Hania Al-Hallaq
{"title":"Automated linac downtime tracking across 4 radiation oncology clinics within a single hospital system utilizing statistical process control","authors":"Caroline E. Serafini,&nbsp;Matthew Thomas,&nbsp;Xiaofeng Yang,&nbsp;Pretesh Patel,&nbsp;Justin Roper,&nbsp;Maxine Washington,&nbsp;Anees Dhabaan,&nbsp;Hania Al-Hallaq","doi":"10.1002/acm2.70863","DOIUrl":null,"url":null,"abstract":"<div>\n \n <section>\n \n <h3> Background</h3>\n \n <p>Linear accelerator (linac) downtime disrupts clinical workflow and may delay patient treatment; however, downtime events are frequently undocumented or recorded in informal reporting systems. Data-driven approaches using routinely collected operational metrics may improve downtime surveillance and provide insight into system-level adaptation.</p>\n </section>\n \n <section>\n \n <h3> Purpose</h3>\n \n <p>This study aims to evaluate if statistical process control (SPC) methods can be applied to routine treatment and scheduling data from an oncology information system (OIS) to identify linac downtime events and characterize operational response across multiple radiation oncology clinics within a single hospital system.</p>\n </section>\n \n <section>\n \n <h3> Methods</h3>\n \n <p>Daily machine-level operational data from 10 linacs across 4 clinics were analyzed over 29 months (January 2024–May 2026). Machine metrics included daily patient volume, cancellations, and operating hours. For each machine, rolling 45-day baseline means and standard deviations were computed, and SPC rules (± 3σ control limits and &gt; 10 cancellations / day) were applied to detect downtime days. SPC-flagged events were compared with email-based downtime event logs to calculate sensitivity, specificity, positive predictive value (PPV), and negative predictive value (NPV). Downtime events were categorized by cause, and intra-hospital patient redistribution was assessed by identifying concurrent &gt; 4σ volume increases on alternate machines.</p>\n </section>\n \n <section>\n \n <h3> Results</h3>\n \n <p>Over the 29-month study period, 138 downtime events were documented in downtime event logs. SPC identified 23 additional events exhibiting operational signatures consistent with downtime. Using reported events as the reference standard, SPC achieved sensitivity 0.70, specificity 0.99, PPV 0.79, and NPV 0.99. Downtime frequently resulted in compensatory increases in patient volume on alternate machines, with 41 transfer pairs identified across three hospitals. When the reported downtime cause was specified, machine mechanical causes accounted for the largest proportion (48%) of downtime events.</p>\n </section>\n \n <section>\n \n <h3> Conclusions</h3>\n \n <p>SPC applied to routinely collected operational data provides a scalable framework for automated linac downtime detection. Integration of automated monitoring with structured reporting may improve downtime surveillance and inform proactive operational planning in radiation oncology</p>\n </section>\n </div>","PeriodicalId":14989,"journal":{"name":"Journal of Applied Clinical Medical Physics","volume":"27 10","pages":""},"PeriodicalIF":2.5000,"publicationDate":"2026-10-14","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://onlinelibrary.wiley.com/doi/epdf/10.1002/acm2.70863","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Applied Clinical Medical Physics","FirstCategoryId":"3","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1002/acm2.70863","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

Abstract

Background

Linear accelerator (linac) downtime disrupts clinical workflow and may delay patient treatment; however, downtime events are frequently undocumented or recorded in informal reporting systems. Data-driven approaches using routinely collected operational metrics may improve downtime surveillance and provide insight into system-level adaptation.

Purpose

This study aims to evaluate if statistical process control (SPC) methods can be applied to routine treatment and scheduling data from an oncology information system (OIS) to identify linac downtime events and characterize operational response across multiple radiation oncology clinics within a single hospital system.

Methods

Daily machine-level operational data from 10 linacs across 4 clinics were analyzed over 29 months (January 2024–May 2026). Machine metrics included daily patient volume, cancellations, and operating hours. For each machine, rolling 45-day baseline means and standard deviations were computed, and SPC rules (± 3σ control limits and > 10 cancellations / day) were applied to detect downtime days. SPC-flagged events were compared with email-based downtime event logs to calculate sensitivity, specificity, positive predictive value (PPV), and negative predictive value (NPV). Downtime events were categorized by cause, and intra-hospital patient redistribution was assessed by identifying concurrent > 4σ volume increases on alternate machines.

Results

Over the 29-month study period, 138 downtime events were documented in downtime event logs. SPC identified 23 additional events exhibiting operational signatures consistent with downtime. Using reported events as the reference standard, SPC achieved sensitivity 0.70, specificity 0.99, PPV 0.79, and NPV 0.99. Downtime frequently resulted in compensatory increases in patient volume on alternate machines, with 41 transfer pairs identified across three hospitals. When the reported downtime cause was specified, machine mechanical causes accounted for the largest proportion (48%) of downtime events.

Conclusions

SPC applied to routinely collected operational data provides a scalable framework for automated linac downtime detection. Integration of automated monitoring with structured reporting may improve downtime surveillance and inform proactive operational planning in radiation oncology

利用统计过程控制,在单个医院系统内跨4个放射肿瘤学诊所进行自动直线机停机跟踪。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
3.60
自引率
19.00%
发文量
331
审稿时长
3 months
期刊介绍: Journal of Applied Clinical Medical Physics is an international Open Access publication dedicated to clinical medical physics. JACMP welcomes original contributions dealing with all aspects of medical physics from scientists working in the clinical medical physics around the world. JACMP accepts only online submission. JACMP will publish: -Original Contributions: Peer-reviewed, investigations that represent new and significant contributions to the field. Recommended word count: up to 7500. -Review Articles: Reviews of major areas or sub-areas in the field of clinical medical physics. These articles may be of any length and are peer reviewed. -Technical Notes: These should be no longer than 3000 words, including key references. -Letters to the Editor: Comments on papers published in JACMP or on any other matters of interest to clinical medical physics. These should not be more than 1250 (including the literature) and their publication is only based on the decision of the editor, who occasionally asks experts on the merit of the contents. -Book Reviews: The editorial office solicits Book Reviews. -Announcements of Forthcoming Meetings: The Editor may provide notice of forthcoming meetings, course offerings, and other events relevant to clinical medical physics. -Parallel Opposed Editorial: We welcome topics relevant to clinical practice and medical physics profession. The contents can be controversial debate or opposed aspects of an issue. One author argues for the position and the other against. Each side of the debate contains an opening statement up to 800 words, followed by a rebuttal up to 500 words. Readers interested in participating in this series should contact the moderator with a proposed title and a short description of the topic
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书